TY - THES U1 - Master Thesis A1 - Hohmann, Siegfried T1 - Development of a parallelizable QCM-D array for the mass spectrometric analysis of proteins N2 - Quarz crystal microbalances allow the monitoring of the adsorption process of mass from a liquid to their surface. The adsorbed mass can be analysed regarding to its protein content using mass spectromety. To ensure the protein identification the results of several measurements can be combined. A high content QCM-D array was developed to allow up to ten measurements parallel. The samples can be routed inside the array distributing one sample to several chips. The fluidic parts were prototyped using 3D printing. The assembled array was tight and the sample routing function could be demonstrated. A temperature controller was developed and implemented. The parameters for the PID controller were determined and the controller was shown to be able to keep the temperature constant over long time with high accuracy. KW - QCM-D Array KW - Mass spectrometry KW - High content measurement KW - 3D printing KW - Temperature control KW - Proteine KW - Massenspektrometrie KW - 3D-Druck KW - Temperaturregelung Y2 - 2015 U6 - https://nbn-resolving.org/urn:nbn:de:bsz:ofb1-opus4-9751 UN - https://nbn-resolving.org/urn:nbn:de:bsz:ofb1-opus4-9751 SP - II, 43, xlii S1 - II, 43, xlii CY - Offenburg ER -