TY - CHAP U1 - Konferenzveröffentlichung A1 - Eisenträger, Daniela A1 - Haltenberger, Anna A1 - Ismer, Bruno A1 - Lorenz, Martin A1 - Straube, Kerstin A1 - Töpfer, Anja A1 - Allmann, Jakob A1 - Figulla, Hans Reiner T1 - Analysing-software for electrocardiographic desynchronization – which parameters needs the physician? T2 - Biomedical Engineering / Biomedizinische Technik N2 - Introduction: Cardiac resynchronization therapy (CRT) with biventricular (BV) pacing is an established therapy for heart failure (HF) patients with ventricular desynchronization and reduced left ventricular (LV) ejection fraction. The aim of this study was to evaluate electrical ventricular desynchronization with transthoracic and transesophageal signal averaging electrocardiography in HF, to better select patients for CRT. Methods: 13 HF patients (age 68 ± 10 years; 2 females, 11 males) with New York Heart Association (NYHA) class 2.8 ± 0.5, 28.6 ± 12.6 % LV ejection fraction and 155 ± 24 ms QRS duration (QRSD) were analysed with transthoracic and transesophageal electrocardiogram recording and novel National Intruments LabView 2009 signal averaging software. Esophageal TO Osypka catheter was perorally applied to the esophagus and placed in the position of maximum LV de-flection. The 0.05-Hz high-pass filtered surface electrocardiogram and the 10-Hz high-pass filtered bipolar transesophageal electrocardiogram were recorded with Bard EP-System and 1000-Hz sampling rate. Results: Transesophageal LV electrogram recording was possible in all HF patients (n=13). Transesophageal interventricular conduction delay (IVCD) was 51 ± 19 ms and measured between the earliest onset of QRS in the 12-channel surface electrocardiogram and the onset of the LV deflection in the transesophageal electrocardiogram. Transesophageal intra-left ventricular delay (LVCD) was 90 ± 16 ms and measured between the onset and offset of the LV deflection in the transesophageal electrocardiogram. QRSD to transesophageal IVCD ratio was 3.43 ± 1.31 ms, QRSD to transesophageal LVCD ratio was 1.75 ± 0.28 ms and QRSD was evaluated between onset and offset of QRS signal in the 12-channel surface electrocardiogram. Conclusion: Determination of IVCD, LVCD, QRSD-to-IVCD-ratio and QRSD-to-LVCD-ratio by transesophageal LV electrogram recording with LabView 2009 signal averaging technique may be useful parameters of ventricular desynchronisation to improve patient selection for CRT. Y1 - 2011 SN - 0013-5585 (Print) SS - 0013-5585 (Print) SN - 1862-278X (Online) SS - 1862-278X (Online) U6 - https://doi.org/10.1515/bmt.2011.850 DO - https://doi.org/10.1515/bmt.2011.850 VL - 56 IS - S1 SP - 6 S1 - 1 PB - Walter de Gruyter CY - Berlin, Boston ER -