TY - CHAP U1 - Konferenzveröffentlichung A1 - Härtig, Johannes A1 - Hörth, Johannes A1 - Melichercik, Juraj A1 - Ismer, Bruno T1 - Exclusion of Adverse Hemodynamic Programming in Cardiac Resynchronization Therapy T2 - Biomedical engineering = Biomedizinische Technik N2 - Decrease of non-responder rate is the main chal-lenge in cardiac resynchronization therapy. The problem could be solved, partly, in the follow-up by consequent indi-vidualization of hemodynamic pacing parameters. The eso-phageal electrogram feature of the Biotronik ICS 3000 programmer was used in the follow-up of 20 heart failure patients carrying implants for cardiac resynchronization therapy. Adverse hemodynamic programming of the sensed and paced AV delay could be easily observed and replaced by the individual optimal duration in 3 patients (15%) VDD and DDD operation.This result proves the value of esophageal electrogram recording CRT follow-up. KW - Herzkrankheit KW - Synchronisierung Y1 - 2013 SN - 0013-5585 SS - 0013-5585 U6 - https://doi.org/10.1515/bmt-2013-4216 DO - https://doi.org/10.1515/bmt-2013-4216 VL - 58 IS - SI-1-Track-I SP - 2 S1 - 2 PB - DeGruyter ER -