Background: The application of high-frequency ablation is used for the treatment of tachycardia arrhythmias and is a respected method. Ablation with high frequency current leads to the targeted heat destruction of myocardial tissue at specific sites and thus prevents the pathological propagation of excitation through these structures.
Purpose: The aim of this study was to simulate heat propagation during RF ablation with modeled electrodes in different sizes and materials. The simulation was performed on atrioventricular node re-entry tachycardia (AVNRT), atrioventricular re-entry tachycardia (AVRT) and atrial flutter (AFL).
Methods: Using the modeling and simulation software CST, ablation catheters with 4 mm and 8 mm tip electrodes were modeled from gold and platinum for each. The designed catheters correspond to the manufacturer"s specifications of Medtronic, Biotronik and Osypka. The catheters were integrated into the Offenburg heart rhythm model to simulate and compare the heat propagation during an ablation application, which also takes into account the blood flow in the four heart chambers. A power of 5 W - 40 W was simulated for the 4 mm electrodes and a power of 50 W - 80 W for the 8 mm electrodes.
Results: During the simulated HF ablation application, the temperature at the ablation electrode was measured at different powers. This is 40.67°C at 5 W, 44.34°C at 10 W, 51.76°C at 20 W, 59.0°C at 30 W, and 66.33°C at 40 W. The measured temperature during 40 W application is 39.5°C at 0,5 mm depth in the myocardium and 37.5°C at 2 mm depth.
In the simulation, the 8 mm platinum electrode reached an ablation temperature of 72.85°C at its tip during an applied power of 60 W. In contrast, the 8 mm platinum electrode reached a depth of 5 mm at 39.5 C° and at a depth of 2 mm at 37.5 °C. In contrast, the 8 mm gold electrode reached a temperature of 64.66°C with the same performance. This is due to the thermal properties of gold, which has a better thermal conductivity than platinum.
Conclusions: CST offers the possibility to carry out a static and dynamic simulation of a heart model and the ablation electrodes integrated in it during an HF ablation. In variation with different electrode sizes and materials, therapy methods for the treatment of AVNRT, AVRT and AFL can be optimized
Electrode modelling and simulation of diagnostic and pulmonary vein isolation in atrial fibrillation
(2022)
Background: Transesophageal left atrial (LA) pacing and transesophageal LA ECG recording are semi-invasive techniques for diagnostic and therapy of supraventricular rhythm disturbance. Cardiac resynchronization therapy (CRT) with right atrial (RA) sensed biventricular pacing is an established therapy for heart failure patients with reduced left ventricular (LV) ejection fraction, sinus rhythm and interventricular electrical desynchronization.
Purpose: The aim of the study was to evaluate electromagnetic and voltage pacing fields of the combination of RA pacing, LA pacing and biventricular pacing in patients with long interatrial and interventricular electrical desynchronization.
Methods: The modelling and electromagnetic simulations of transesophageal LA pacing in combination with RA pacing and biventricular pacing would be staged and analyzed with the CST (Computer Simulation Technology) software. Different electrodes were modelled in order to simulate different types of bipolar pacing in the 3D-CAD Offenburg heart rhythm model: The bipolar Solid S (Biotronik) electrode where modelled for RA pacing and right ventricular (RV) pacing, Attain 4194 (Medtronic) for LV pacing and TO8 (Osypka) multipolar esophageal electrode with hemispheric electrodes for LA pacing.
Results: The pacemaker amplitudes for the electromagnetic pacing simulations were performed with 3 V for RA pacing, 1.5 V for RV pacing, 50 V for LA pacing and 3V for LV pacing with pacing impulse duration of 0.5 ms for RA, RV and LV pacing and 10 ms for LA pacing. The atrioventricular pacing delay after RA pacing was 140 ms. The different pacing modes AAI, VVI, DDD, DDD0V and DDD0D were evaluated for the analysis of the electric pacing field propagation of pacemaker, CRT and LA pacing. The pacing results were compared at minimum (LOW) and maximum (HIGH) parameter settings. While the LOW setting produced fewer tetrahedral and more inaccurate results, the HIGH setting produced many tetrahedral and therefore more accurate results.
Conclusions: The simulation of the combination of transesophageal LA pacing with RA sensed biventricular pacing is possible with the Offenburg heart rhythm model. The new temporary 4-chamber pacing method may be additional useful method in CRT non-responders with long interatrial electrical delay.
The present invention relates to open-loop and closed-loop control units for extracorporeal circulatory support, to systems comprising such an open-loop and closed-loop control unit, and to corresponding methods. An open-loop and closed-loop control unit (10) for extracorporeal circulatory support is proposed, which is configured to receive a measurement of an ECG signal (12) of a supported patient over a predefined period of time, wherein the ECG signal (12) comprises multiple data points for each time point within a heart cycle. The open-loop and closed-loop control unit (10) comprises an evaluation unit (100) which is configured to evaluate the data points for at least one time point in a spatial and/or temporal manner and to determine at least one amplitude change (14) within the heart cycle based on the evaluated data points. The open-loop and closed-loop control unit (10) is further configured to output an open-loop and/or closed-loop signal (16) for extracorporeal circulatory support at a predefined point in time after the at least one amplitude change (14).
The present invention relates to open-loop and closed-loop control units for extracorporeal circulatory support, to systems comprising such an open-loop and closed-loop control unit, and to corresponding methods. An open-loop and closed-loop control unit (10) for extracorporeal circulatory support is proposed, which is configured to receive a measurement of an ECG signal (12) of a supported patient over a predefined period of time, wherein the ECG signal (12) comprises multiple data points for each time point within a heart cycle. The open-loop and closed-loop control unit (10) comprises an evaluation unit (100) which is configured to evaluate the data points for at least one time point in a spatial and/or temporal manner and to determine at least one amplitude change (14) within the heart cycle based on the evaluated data points. The open-loop and closed-loop control unit (10) is further configured to output an open-loop and/or closed-loop signal (16) for extracorporeal circulatory support at a predefined point in time after the at least one amplitude change (14).
Device and method for monitoring and optimising a temporal trigger stability (WO2023094554A1)
(2023)
The present invention relates to devices for monitoring and optimising a temporal trigger stability of an extracorporeal circulatory support means, and to open-loop and closed-loop control units for the extracorporeal circulatory support means comprising such a device, and to corresponding methods. A device (10) for monitoring a temporal trigger stability of an extracorporeal circulatory support means is accordingly proposed, which device is designed to receive a first dataset (14) of a measurement of an ECG signal of a supported patient over a predefined period of time. The device (10) comprises an evaluation unit (16), which is designed to determine or identify a plurality of R triggers (26) from the first dataset (14), wherein the evaluation unit (16) is also designed to receive or provide a second dataset (20) having evaluated ECG signals and a plurality of R triggers (28) and to selectively map the second dataset (20) on the first dataset (14). The device is also designed to emit a signal (22) that characterises a temporal gap between successive R triggers (26) from the first dataset (14) and successive R triggers (28) from the second dataset (20) which are mapped on the first dataset.
Oesophageal Electrode Probe and Device for Cardiological Treatment and/or Diagnosis (US20200261024)
(2020)
An oesophageal electrode probe for bioimpedance measurement and/or for neurostimulation is provided; a device for transoesophageal cardiological treatment and/or cardiological diagnosis is also provided; a method for the open-loop or closed-loop control of a cardiological catheter ablation device and/or a cardiological, circulatory and/or respiratory support device is also provided. The oesophageal electrode probe comprises a bioimpedance measuring device for measuring the bioimpedance of at least one part of tissue surrounding the oesophageal electrode probe. The bioimpedance device comprises at least one first and one second electrode. The at least one first electrode is arranged on a side of the oesophageal electrode probe facing towards the heart. The at least one second electrode is arranged on a side of the oesophageal electrode probe facing away from the heart. The device comprises the oesophageal electrode probe and a control and/or evaluation device.
Disturbances of the cardiac conduction system causing reentry mechanisms above the atrioventricular (AV) node are induced by at least one accessory pathway with different conducting properties and refractory periods. This work aims to further develop the already existing and continuously expanding Offenburg heart rhythm model to visualise the most common supraventricular reentry tachycardias to provide a better understanding of the cause of the respective reentry mechanism.
Background: Pulmonary vein isolation (PVI) using cryoballoon catheters are a recognized method for the treatment of atrial fibrillation (AF). This method offers shorter treatment duration in contrast to the classical therapy with high-frequency (HF) ablation.
Purpose: The aim of this study was to integrate different cryoballoon catheters and a HF catheter into a heart rhythm model and to compare them by means of static and dynamic electromagnetic and thermal simulation in use under AF.
Methods: The cryoballoon catheters from Medtronic and the HF ablation catheter from Osypka were modelled virtually with the aid of manufacturer specifications and the CST (Computer Simulation Technology, Darmstadt) simulation program. The cryoballoon catheter was located in the lower left pulmonary vein of the virtual heart rhythm model for the realization of pulmonary vein isolation (PVI) by cryoenergy. The simulated temperature at the balloon surface was -50°C during the simulation.
Results: During a simulated 20 second application of a cryoballoon catheter at -50°C, a temperature of -24°C was measured at a depth of 0.5 mm in the myocardium. At a depth of 1 mm the temperature was -3°C, at 2 mm depth 18°C and at 3 mm depth 29°C. Under the 15 second application of a RF catheter with a 8 mm electrode and a power of 5 W at 420 kHz, the temperature at the tip of the electrode was 110°C. At a depth of 0.5 mm in the myocardium, the temperature was 75°C, at a depth of 1 mm 58°C, at 2 mm depth 45°C and at 3 mm depth 38°C.
Conclusions: The simulation of temperature profiles during the virtual application of several catheter models in the heart rhythm model allows the static and dynamic simulation of PVI by cryoballoon ablation and RF ablation. The three-dimensional simulation can be used to improve ablation applications by creating a model in personalized cardiac rhythm therapy from MRI or CT data of a heart and finding a favourable position for ablation of AF.