Hintergrund: Das elektrische interventrikuläre Delay (IVD) ist bei Patienten (P) mit Herzinsuffizienz (HF), reduzierter linksventrikulärer (LV) Funktion und verbreitertem QRS Komplex von Bedeutung für den Erfolg der kardialen Resynchronisationstherapie (CRT). Die transösophageale LV Elektrokardiographie (EKG) ermöglicht die Bestimmung des elektrischen IVD und linksventrikulären Delays (LVD). Das Ziel der Studie besteht in der Untersuchung des transösophagealen elektrischen IVD, LVD und deren Verhältnis zur QRS Dauer bei rechtsventrikulärer (RV) Stimulation vor Aufrüstung auf eine biventrikuläre (BV) Stimulation.
Methoden: Bei 11 HF P (Alter 69,0 ± 7,9 Jahre; 10 Männer und 1 Frau) mit DDD Schrittmacher (n=10), DDD Defibrillator (n=1) und RV Stimulation, New York Heart Association (NYHA) Klasse 3,0 ± 0,2, LV Ejektionsfraktion 24,5 ± 4,9 % und QRS-Dauer 228,2 ± 44,8 ms wurden das elektrische IVD als Intervall zwischen Beginn des QRS-Komplexes im Oberflächen EKG und Beginn des LV Signals im transösophagealen LV EKG und das elektrische LVD als Intervall zwischen Beginn und Ende des LV Signals im transösophagealen LV EKG präoperativ vor Aufrüstung auf CRT Defibrillator (n=8) und CRT Schrittmacher (n=3) bestimmt. Der Anstieg des arteriellen Pulse Pressure (PP) wurde zwischen RV Stimulation und transösophagealer LV Stimulation mit unterschiedlichem AV-Delay (n=5) vor Aufrüstung von RV auf BV Stimulation getestet.
Ergebnisse: Bei RV Stimulation betrugen IVD 86,54 ± 32,80 ms, LVD 94,45 ± 23,80 ms, QRS-IVD-Verhältnis 2,63 ± 0,81 mit negativer Korrelation zwischen IVD und QRS-IVD-Verhältnis (r=-0,668 P=0,0248) (Fig.) und QRS-LVD-Verhältnis 2,33 ± 0,73. Vorhofsynchrone ventrikuläre Stimulation führte zu 63,6 ± 27,7 mmHg PP bei RV Stimulation und 80,6 ± 38,5 mmHg PP bei LV Stimulation und der PP erhöhte sich bei LV Stimulation mit optimalem AV Delay um 17 ± 11,2 mmHg gegenüber RV Stimulation (P<0,001). Nach Aufrüstung von RV Stimulation auf BV Stimulation verbesserten sich die NYHA Klasse von 3,1 ± 0,2 auf 2,2 ± 0,3 während 30,4 ± 29,6 Monaten CRT.
Schlussfolgerungen: Das transösophageale LV EKG ermöglicht die Bestimmung des elektrischen IVD und LVD bei RV Stimulation zur Evaluierung der interventrikulären und linksventrikulären elektrischen Desynchronisation. IVD, LVD und deren Verhältnis zur QRS Dauer können möglicherweise zur Vorhersage einer CRT Response vor Aufrüstung von RV auf BV Stimulation genutzt werden.
Heart rhythm model and simulation of electrophysiological studies and high-frequency ablations
(2017)
Background: The simulation of complex cardiologic structures has the potential to replace clinical studies due to its high efficiency regarding time and costs. Furthermore, the method is more careful for the patients’ health than the conventional ways. The aim of the study was to create an anatomic CAD heart rhythm model (HRM) as accurate as possible, and to show its usefulness for cardiac electrophysiological studies (EPS) and high-frequency (HF) ablations.
Methods: All natural heart components of the new HRM were based on MRI records, which guaranteed electronic functionality. The software CST (Computer Simulation Technology, Darmstadt) was used for the construction, while CST’s material library assured genuine tissue properties. It should be applicable to simulate different heart rhythm diseases as well as various diffusions of electromagnetic fields, caused by electrophysiological conduction, inside the heart tissue.
Results: It was achievable to simulate normal sinus rhythm and fourteen different heart rhythm disturbance with different atrial and ventricular conduction delays. The simulated biological excitation of healthy and sick HRM were plotted by simulated electrodes of four polar right atrial catheter, six polar His bundle catheter, ten polar coronary sinus catheter, four polar ablation catheter and eight polar transesophageal left cardiac catheter (Fig.). Accordingly, six variables were rebuilt and inserted into the anatomic HRM in order to establish heart catheters for ECG monitoring and HF ablation. The HF ablation catheters made it possible to simulate various types of heart rhythm disturbance ablations with different HF ablation catheters and also showed a functional visualisation of tissue heating. The use of tetrahedral meshing HRM made it attainable to store the results faster accompanied by a higher degree of space saving. The smart meshing function reduced unnecessary high resolutions for coarse structures.
Conclusions: The new HRM for EPS simulation may be additional useful for simulation of heart rhythm disturbance, cardiac pacing, HF ablation and for locating and identification of complex fractioned signals within the atrium during atrial fibrillation HF ablation.
Background: Cardiac resynchronization therapy (CRT) is an established therapy for heart failure (HF) patients (P) with reduced left ventricular (LV) ejection fraction and electrical interventricular desynchronization, but not all P improved clinically. The aim of the study was to evaluate electrical interventricular delay (IVD) to LV delay (LVD) ratio in atrial fibrillation (AF) CRT responder (R) and non-responder (NR).
Methods: AF P (n = 18, age 60.6 ± 11.4 years, 1 female, 17 males) with HF New York Heart Association (NYHA) class 3.0 ± 0.2, 25.3 ± 5.9 % LV ejection fraction and 157.8 ± 24.4 ms QRS duration (QRSD) were measured by surface ECG and focused transesophageal bipolar LV ECG before implantation of CRT pacemaker (n = 2) or CRT defibrillator (n = 16). IVD was measured between onset of QRS in the surface ECG and onset of LV signal in the LV ECG. LVD was measured between onset and offset of LV signal in the LV ECG.
Results: Electrical ventricular desynchronization in AF CRT P were 61.9 ± 26.9ms IVD, 80.6 ± 24.3ms LVD, 0.85 ± 0.41 IVD-LVD-ratio (Figure), 3.12 ± 1.89 QRSD-IVD-ratio and 2.07 ± 0.47 QRSD-LVD-ratio. There were 72.2 % AF CRT R (n = 13) with 64.2 ± 24.6ms IVD and 77.8 ± 21.6ms LVD with Pearson correlation to 0.89 ± 0.39 IVD-LVD-ratio (r = 0.87, P < 0.01; r = -0.69, P < 0.01), 2.82 ± 1.32 QRSD-IVD-ratio (r = -0.76, P < 0.01; r = 0.67, P = 0.011) and 2.13 ± 0.46 QRSD-LVD-ratio (r = 0.57, P = 0.041; r = -0.85, P < 0.01). There were 27.8% AF CRT NR (n = 5) with 56.0 ± 34.5ms IVD and 87.8 ± 31.9ms LVD without correlation to 0.74 ± 0.48 IVD-LVD-ratio, 3.88 ± 2.98 QRSD-IVD-ratio and 1.90 ± 0.48 QRSD-LVD-ratio. During 15.3 ± 13.1 month CRT follow-up, the AF CRT R NYHA class improved from 3.0 ± 0.2 to 2.2 ± 0.3 (P < 0.001). During 18.8 ± 20.7 month CRT follow-up, the AF CRT NR NYHA class not improved from 3 to 3.3 ± 0.97.
In contrast to conventional aortic valve replacement, the Transcatheter Aortic Valve Implantation (TAVI) is a new highly specialist alternative to surgical valve replacement for patients with symptomatic severe aortic stenosis and high operative risk. The procedure was performed in a minimally invasive way and was introduced at the university heart centre, Freiburg – Bad Krozingen in 2008. The results have been getting better and better over the years. The aim of the investigation is the analysis of electrocardiogram conduction time and the electrocardiography changes recorded hours and days after the procedure depending on artificial heart valve models, which may lead to pacemaker implantation, even the analysis of the effectiveness of treatment.
Transthoracic impedance cardiography (ICG) is a non-invasive method for determination of hemodynamic parameters. The basic principle of transthoracic ICG is the measurement of electrical conductivity of the thorax over the time. The aim of the study was the analysis of hemodynamic parameters from healthy individuals and the evaluation of various hemodynamic monitoring devices. Fourteen men (mean age 25 ± 4.59 years) and twelve women (mean age 24 ± 3.5 years) were measured during the cardiovascular engineering laboratory at Offenburg University of Applied Sciences, Offenburg, Germany. The ICG recordings were measured with the devices CardioScreen 1000, CardioScreen 2000 and TensoScreen with the corresponding Software Cardiovascular Lab 2.5 (Medis Medizinische Messtechnik GmbH, Illmenau, Germany). In order to create identical frame conditions, all measurements were recorded in the same position and for the same duration. Various positions were simulated from horizontal lying position to vertical standing position. Altogether, more than 30 hemodynamic parameters were measured.
Cardiac resynchronization therapy (CRT) is an established class I level A biventricular pacing therapy in chronic heart failure patients with left bundle branch block and reduced left ventricular ejection fraction, but not all patients improved clinically. Purpose of the study was to evaluate electrical interatrial conduction delay (IACD) to interventricular conduction delay (IVCD) ratio with focused transesophageal left atrial and left ventricular electrocardiography.
Methods: Thirty eight chronic heart failure patients (age 63.4±10.2 years; 3 females, 35 males) with New York Heart Association (NYHA) functional class 3.0±0.2 and 171.71±36.17ms QRS duration were analysed using posterior left atrial and left ventricular transesophageal electrocardiography with hemispherical electrodes before CRT. Electrical IACD was measured between onset of P-wave in the surface ECG and onset of left atrial signal. Electrical IVCD was measured between onset of QRS complex in the surface ECG and onset of left ventricular signal.
Results: Electrical IACD and IVCD could be evaluated by transesophageal left atrial and left ventricular electrocardiography in all heart failure patients with correlation to 1.18±0.92 IACD-IVCD-ratio (r=-0.57, P<0.001; r=0.66, P<0.001). There were 32 CRT responder with reduction of NYHA class from 3.0±0.22 to 1.97±0.31 (P<0.001) during 16.5±18.9 month CRT with 75.19±33.49ms IACD, 78.91±24.73ms IVCD, 1.04±0.66 IACD-IVCD-ratio and correlation between IACD and IACDIVCD- ratio (r=0.84, P<0.001). There were 6 CRT nonresponder with no reduction of NYHA class from 3.0±0.3 to 2.9±0.5 during 14.3±13.7 month biventricular pacing, 50.0±28.26ms IVCD (P=0.014), 1.92±1.65 IACD-IVCD-ratio (P=0,029) and correlation between 67.0±24.9ms IACD and IACD-IVCD-ratio (r=0.85, P=0.031).
Conclusions: Focused transesophageal left atrial and left ventricular electrocardiography can be utilized to analyse electrical IACD and IVCD in heart failure patients. IACDIVDC- ratio may be a useful parameter to evaluate electrical left cardiac desynchronization in heart failure patients.
Cardiac resynchronization therapy (CRT) is an established biventricular pacing therapy in heart failure patients with left bundle branch block and reduced left ventricular ejection fraction, but not all patients improved clinically as CRT responder. Purpose of the study was to evaluate electrical left atrial conduction delay (LACD) with focused transesophageal electrocardiography in CRT responder and CRT non-responder.
Methods: Twenty heart failure patients (age 66.6±8.2 years; 2 females, 18 males) with New York Heart Association functional class 3.0±0.3 and 174.2±40.2ms QRS duration were analysed using posterior left atrial transesophageal electrocardiography with hemispherical electrodes. Electrical LACD was measured between onset and offset of transesophageal left atrial signal before implantation of CRT devices.
Results: Electrical LACD could be evaluated by bipolar transesophageal left atrial electrocardiography using TO Osypka electrode in all heart failure patients with negative correlation between 54.7±18.1ms LACD and 24.9±6.4% left ventricular ejection fraction (r=-0.65, P=0.002). There were 16 CRT responders with reduction of New York Heart Association functional class from 3.0±0.29 to 2.1±0.2 (r=0.522, P=0.038) during 9.41±10.96 month biventricular pacing and negative correlation between 49.6±14.2ms LACD and 26.0±6.2% left ventricular ejection fraction (r=-0.533, P=0.034). There were 4 CRT non-responders with no reduction of New York Heart Association functional class from 3.0±0.4 to 2.8±0.5 (r=0.816, P=0.184) during with 13.88±16.39 month biventricular pacing and no correlation between 75.25±19.17ms LACD and 20.75±6.4% left ventricular ejection fraction (r=-0.831, P=0.169).
Conclusions: Focused transesophageal left atrial electrocardiography can be utilized to analyse electrical LACD in heart failure patients. LACD correlated negative with left ventricular ejection fraction in CRT responders. LACD may be a useful parameter to evaluate electrical left atrial desynchronization in heart failure patients.
Hintergrund: Das elektrische interventrikuläre Delay (IVD) und die Lage der linksventrikulären (LV) Elektrode zum Ort der spätesten LV Erregung sind bei Patienten (P) mit Herzinsuffizienz (HF), reduzierter LV Funktion und breiter QRS Dauer (QRSD) von Bedeutung für den Erfolg der kardialen Resynchronisationstherapie (CRT). Die LV Elektrokardiographie ermöglicht eine Abschätzung des elektrischen IVD. Ziel der Studie besteht in der nicht-invasiven Evaluierung des elektrischen IVD bei Patienten (P) mit Vorhofflimmern (AFib) mit und ohne CRT mit biventrikulärer (BV) Stimulation.
Methoden: Bei 49 HF P mit AFib (Alter 63,9 ± 10,8 Jahre; 43 Männer und 6 Frauen) mit New York Heart Association (NYHA) Klasse 2,9 ± 0,4, LV Ejektionsfraktion 26,03 ± 7,99 % und QRS-Dauer (QRSD) 143,69 ± 35,62 ms wurde das elektrische IVD als Intervall zwischen Beginn des QRS-Komplexes im Oberflächen EKG und Beginn des LV Signals im transösophagealen LV EKG bei 31 HF P mit AFib und bei 18 HF P mit AFib und CRT präoperativ bestimmt. Das fokussierte bipolare LV EKG wurde mittels Osypka TO Sonde mit halbkugelförmigen Elektroden in Höhe des maximalen LV Signals registriert.
Ergebnisse: Bei 31 HF P mit AFib betrugen QRSD 135,48 ± 38,78 ms, IVD 49,55 ± 26,38 ms, QRSD-IVD-Verhältnis 3,12 ± 1,11 und das IVD korrelierte mit der QRSD (r=0,75, P<0,001) und dem QRSD-IVD-Verhältnis (r=-0,67, P<0,001) (Fig.). Bei 18 HF P mit AFib und CRT Defibrillator betrugen QRSD 157,83 ± 24,38 ms, IVD 61,94 ± 26,88 ms, QRSD-IVD-Verhältnis 3,12 ± 1,89 und das IVD korrelierte mit der QRSD (r=0,47, P=0,049) und dem QRSD-IVD-Verhältnis (r=-0,73, P<0,001). Bei 72,2 % CRT Responder (R) (n=13) betrugen QRSD 158,15 ± 22,4 ms, IVD 64,23 ± 24,62 ms, QRSD-IVD-Verhältnis 2,82 ± 1,32 und das IVD korrelierte mit der QRSD (r=0,57, P=0,043) und dem QRSD-IVD-Verhältnis (r=-0,76, P=0,0024). Bei 27,8 % CRT Non-Responder (NR) (n=5) betrugen QRSD 157 ± 31,94 ms, IVD 56 ± 34,52 ms, QRSD-IVD-Verhältnis 3,88 ± 2,98 und das IVD korrelierte nicht mit der QRSD (r=0,33, P=0,591) und dem QRSD-IVD-Verhältnis (r=-0,732, P=0,159). Die CRT R verbesserten sich in der NYHA Klasse von 3 ± 0,2 auf 2,2 ± 0,3 (P<0,001) während 15,3 ± 13,1 Monaten BV Stimulation. Bei 15 CRT NR kam es zu keiner Verbesserung der NYHA Klasse von 3 auf 3,3 ± 0,97 (P=0,529) während 18,8 ± 20,7 Monaten BV Stimulation.
Schlussfolgerungen: Das transösophageale LV EKG ermöglicht bei HF-P mit AFib die nichtinvasive Messung des elektrischen IVD präoperativ vor CRT. IVD und QRSD-IVD-Verhältnis sind möglicherweise einfach anwendbare Parameter zur Vorhersage von CRT R und CRT NR bei P mit AFib.