Background: A disturbed synchronization of the ventricular contraction can cause a highly developed systolic heart failure in affected patients, which can often be explained by a diseased left bundle branch block (LBBB). If medication remains unresponsive, the concerned patients will be treated with a cardiac resynchronization therapy (CRT) system. The aim of this study was to integrate His bundle pacing into the Offenburg heart rhythm model in order to visualize the electrical pacing field generated by His bundle pacing.
Methods: Modelling and electrical field simulation activities were performed with the software CST (Computer Simulation Technology) from Dessault Systèms. CRT with biventricular pacing is to be achieved by an apical right ventricular electrode and an additional left ventricular electrode, which is floated into the coronary vein sinus. This conventional type of biventricular pacing leads to a reduction of the left ventricular ejection fraction. Furthermore, the non-responder rate of the CRT therapy is about one third of the CRT patients.
Results: His bundle pacing represents a physiological alternative to conventional cardiac pacing and cardiac resynchronization. An electrode implanted in the His bundle emits a stronger electrical pacing field than the electrical pacing field of conventional cardiac pacemakers. The pacing of the His bundle was performed by the Medtronic Select Secure 3830 electrode with pacing voltage amplitudes of 3 V, 2 V and 1.5 V in combination with a pacing pulse duration of 1 ms.
Conclusions: Compared to conventional cardiac pacemaker pacing, His bundle pacing is capable of bridging LBBB conduction disorders in the left ventricle. The His bundle pacing electrical field is able to spread via the physiological pathway in the right and left ventricles for CRT with a narrow QRS-complex in the surface ECG.
A disturbed synchronization of the ventricular contraction can cause a highly developed systolic heart failure in affected patients with reduction of the left ventricular ejection fraction, which can often be explained by a diseased left bundle branch block (LBBB). If medication remains unresponsive, the concerned patients will be treated with a cardiac resynchronization therapy (CRT) system. The aim of this study was to integrate His-bundle pacing into the Offenburg heart rhythm model in order to visualize the electrical pacing field generated by His-Bundle-Pacing. Modelling and electrical field simulation activities were performed with the software CST (Computer Simulation Technology) from Dessault Systèms. CRT with biventricular pacing is to be achieved by an apical right ventricular electrode and an additional left ventricular electrode, which is floated into the coronary vein sinus. The non-responder rate of the CRT therapy is about one third of the CRT patients. His- Bundle-Pacing represents a physiological alternative to conventional cardiac pacing and cardiac resynchronization. An electrode implanted in the His-bundle emits a stronger electrical pacing field than the electrical pacing field of conventional cardiac pacemakers. The pacing of the Hisbundle was performed by the Medtronic Select Secure 3830 electrode with pacing voltage amplitudes of 3 V, 2 V and 1,5 V in combination with a pacing pulse duration of 1 ms. Compared to conventional pacemaker pacing, His-bundle pacing is capable of bridging LBBB conduction disorders in the left ventricle. The His-bundle pacing electrical field is able to spread via the physiological pathway in the right and left ventricles for CRT with a narrow QRS-complex in the surface ECG.
Oesophageal Electrode Probe and Device for Cardiological Treatment and/or Diagnosis (US20200261024)
(2020)
An oesophageal electrode probe for bioimpedance measurement and/or for neurostimulation is provided; a device for transoesophageal cardiological treatment and/or cardiological diagnosis is also provided; a method for the open-loop or closed-loop control of a cardiological catheter ablation device and/or a cardiological, circulatory and/or respiratory support device is also provided. The oesophageal electrode probe comprises a bioimpedance measuring device for measuring the bioimpedance of at least one part of tissue surrounding the oesophageal electrode probe. The bioimpedance device comprises at least one first and one second electrode. The at least one first electrode is arranged on a side of the oesophageal electrode probe facing towards the heart. The at least one second electrode is arranged on a side of the oesophageal electrode probe facing away from the heart. The device comprises the oesophageal electrode probe and a control and/or evaluation device.
Oesophageal Electrode Probe and Device for Cardiological Treatment and/or Diagnosis (EP3706626A1)
(2020)
The invention relates to an oesophageal electrode probe (10) for bioimpedance measurement and/or for neurostimulation; a device (100) for transoesophageal cardiological treatment and/or cardiological diagnosis; and a method for the open-loop or closed-loop control of a cardiac catheter ablation device and/or a cardiac, circulatory and/or respiratory support device. The oesophageal electrode probe comprises a bioimpedance measuring device for measuring the bioimpedance of at least one part of the tissue surrounding the oesophageal electrode probe. The bioimpedance device comprises at least one first and one second electrode, wherein the at least one first electrode (12A) is arranged on a side (14) of the oesophageal electrode probe facing towards the heart and the at least one second electrode (12B) is arranged on a side (16) of the oesophageal electrode probe facing away from the heart. The device (100) comprises the oesophageal electrode probe (10) and a control and/or evaluation device (30), which is configured for receiving a first bioimpedance measurement signal from the at least one first electrode (12A) and a second bioimpedance measurement signal from the at least one second electrode (12B), and comparing same, and generating a control signal on the basis of the comparison. The control signal can be a signal for the open-loop or closed-loop control of a cardiac catheter ablation device and/or a cardiac, circulatory and/or respiratory support device.