Termination of atrial flutter (AFL) is not possible in all AFL patients (P) with transesophageal left atrial pacing (TLAP) with undirected electrical pacing field (EPF) and high atrial pacing threshold. Purpose of the study was to evaluate bipo-lar transesophageal left atrial electrocardiography (TLAE) and TLAP with directed EPF for evaluation and termination of AFL with and without simultaneous transesophageal echocardiography (TEE).
Methods: AFL P were analysed using either a TO electrode with one cylindrical (CE) and three or seven hemispherical electrodes (HE) or TEE electrode with four HE (Osypka, Rheinfelden, Germany). Burst TLAP cycle length was between 200msand 50ms.
Results: AFL cycle length was 233±30 ms with mean ventricular cycle length of 540±149 ms. AFL could be terminated by rapid bipolar TLAP with directed EPF using HE-HE and CE-HE with induction of atrial fibrillation (AF), induction of AF and spontaneous conversion to sinus rhythm and direct conversion to sinus rhythm. Directed EPF was simulated with finite element method.
Conclusions: AFL can be evaluated by bipolar TLAE. AFL can be terminated with rapid TLAP with directed EPF with and without simultaneous TEE. Bipolar TLAE with rapid TLAP is a safe, simple and useful method for evaluation and termination of AFL.
Cardiac resynchronisation therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients with interventricular conduction delay (IVCD). The aim of the study was to evaluate transesophageal IVCD and left ventricular (LV) pacing with directed electrical pacing field (EPF) in HF patients.
Methods: HF patients were analysed with bipolar transesophageal LV electrocardiogram recording and LV pacing with constant voltage stimulus output, 4 ms stimulus duration, distal cylindrical electrode (CE) and seven 6 mm hemispherical electrodes (HE) with 15 mm electrode distance (TO, Dr. Osypka, Rheinfelden, Germany).
Results: LV electrocardiogram recording with HE-HE and CE-HE evaluated a mean IVCD of 79.9 ± 36.7 ms. Directed EPF with CE-HE and HE-HE allowed LV VAT (n=12) and LV D00 pacing (n=5) with a mean effective capture output of 97.35 ± 6.64 V. In 15 responders with IVCD of 87 ± 33 ms arterial pulse pressure (PP) increased from 65 ± 24 mmHg to 79 ± 27 mmHg (p < 0.001). EPF was simulated with finite element method.
Conclusions: Transesophageal LV electrocardiography and directed EPF pacing with CE and HE allowed the evaluation of IVCD and PP to select patients for BV pacing.
Introduction: Cardiac resynchronization therapy (CRT) with left ventricular (LV) pacing is an established therapy for heart failure (HF) patients (P) with ventricular desynchronisation and reduced LV ejection fraction (EF). The aim of this study was to test the utilization of the transesophageal approach to measure arterial pulse pressure (PP) during LV pacing and electrical interventricular conduction delay (IVCD), to better select patients for CRT.
Methods: 32 HF patients (age 64 ± 10 years; 5 females, 27 males) with New York Heart Association (NYHA) class 2.8 ± 0.6, 27 ± 11 % LV EF and 155 ± 35 ms QRS duration were analysed with semi-invasive left cardiac pacing and electrocardiography. Esophageal TO8 Osypka catheter of 10.5 F diameter was perorally applied to the esophagus and placed in the position of maximum left atrial (LA) deflection and maximum LV deflection to measure PP with VAT or D00 pacing modes.
Results: Temporary transesophageal LV pacing was possible with VAT mode (n=16) and D00 mode (n=16) in all patients. In 15 Δ-PP-responders, PP was higher during LV pacing on than LV pacing off (78.3 ± 26.6 versus 65.9 ± 23.7 mmHg, P < 0.001) and NYHA class improved from 3.1 ± 0.35 to 2.1 ± 0.35 (P < 0.001) during 29 ± 26 month biventricular (BV) pacing follow-up (6 Medtronic and 9 Boston BV pacing devices). In 17 Δ-PP-non-responders, PP was not higher during LV pacing on than LV pacing off (61.5 ± 23.9 versus 60.9 ± 23.5 mmHg, P = 0.066). IVCD was significant longer in Δ-PP-responders than in Δ-PP-non-responders (87 ± 33 ms versus 37± 29 ms, P < 0.001).
Conclusion: Semi-invasive transesophageale LA and LV pacing with D00 and VAT mode and LV electrogram recording may be useful techniques to predict CRT improvement.
Capture threshold (CT) for transesophageal left atrial (LA) pacing (TLAP) and transesophageal left ventricular (LV) pacing (TLVP) with conventional cylindrical electrodes (CE) are higher than TLAP feeling threshold (FT). Purpose of the study was to evaluate focused TLAP CT and FT for supraventricular tachycardia (SVT) initiation and focused TLVP CT for cardiac resynchronisation therapy (CRT) simulation.
Methods: SVT initiation in patients (P) with palpitations (n=49, age 47 ± 17 years) was analysed during spontaneous rhythm and during focused bipolar TLAP with atrial constant current stimulus output, distal CE and three or seven 6 mm hemispherical electrodes (HE) (TO, Osypka AG, Rheinfelden, Germany). CRT simulation in heart failure P (n=75, age 62 ± 11 years) was evaluated by focused bipolar TLAP and/or TLVP with ventricular constant voltage stimulus output and different pacing mode.
Results: Focused electrical pacing field between CE and HE (n=28) allowed low threshold TLAP with 8.0 ± 2.6 mA CT at 9.9 ms stimulus duration (SD) which was lower than 9.2 ± 4.5 mA FT at 9.9 ms SD. Focused electrical pacing field between HE and HE (n=21) allowed low threshold TLAP with 8.1 ± 2.2 mA CT at 9.9 ms SD which was lower than 9.8 ± 5.0 mA FT at 9.9 ms SD. SVT initiation by programmed AAI TLAP was possible in 23 P and not possible in 26 P. CRT simulation was evaluated with TLAP and TLVP with VAT, D00 and V00 pacing mode and 95.5 ± 10.9 V TLVP CT at 4.0 ms SD.
Conclusions: Programmed focused AAI TLAP allowed initiation of SVT with very low CT and high FT and focused electrical pacing field between CE-HE and HE-HE.CRT simulation with focused TLAP and/or TLVP with VAT, D00 and V00 pacing mode may be a useful technique to detect responders to CRT.
Das Ausmaß der elektrischen ventrikulären Desynchronisation bei reduzierter linksventrikulärer Funktion ist von Bedeutung für den Erfolg der Resynchronisationstherapie der Herzinsuffizienz mit biventrikulärer Stimulation. Das Ziel der Untersuchung besteht in der nichtinvasiven Messung der elektrischen inter-ventrikulären Desynchronisation mit und ohne ischämische Herzerkrankung bei kardialen Resynchronisationstherapie Respondern. Bei Patienten mit 25,3 ± 7,3 % reduzierter linksventrikulärer Ejektionsfraktion und 166,9 ± 38,5 ms QRS-Dauer wurde das transösophageale linksventrikuläre EKG abgeleitet. Die QRS-Dauer korrelierte mit dem interventrikulären und links-ventrikulären Delay bei Resynchronisationstherapie Respondern mit nicht-ischämischer Herzerkrankung.
Transösophageales interventrikuläres Delay bei Vorhofflimmern und kardialer Resynchronisation
(2013)
Die transösophageale linksventrikuläre Elektrokardiographie ermöglicht die Evaluierung der elektrischen ventrikulären Desynchronisation im Rahmen der kardialen Resynchronisationstherapie der Herzinsuffizienz. Das Ziel der Untersuchung besteht in der präoperativen Abschätzung des transösophagealen interventrikulären Delays bei Vorhofflimmern und kardialer Resynchronisationstherapie. Bei Patienten mit Vorhofflimmern, Herzinsuffizienz New York Heart Association Klasse 3,0 ± 0,2 und QRS-Dauer 159,6 ± 23,9 ms wurde das fokusierte transösophageale linksventrikuläre EKG abgeleitet. Die kardiale Resynchronisationstherapie Responder QRS-Dauer korrelierte mit dem transösophagealen interventrikulären Delay bei Vorhofflimmern.
Hintergrund: Das elektrische interventrikuläre Delay (IVD) und die Lage der linksventrikulären (LV) Elektrode zum Ort der spätesten LV Erregung sind bei Patienten (P) mit Herzinsuffizienz (HF), reduzierter LV Funktion und breiter QRS Dauer (QRSD) von Bedeutung für den Erfolg der kardialen Resynchronisationstherapie (CRT). Die LV Elektrokardiographie ermöglicht eine Abschätzung des elektrischen IVD. Ziel der Studie besteht in der nicht-invasiven Evaluierung des elektrischen IVD bei Patienten (P) mit Vorhofflimmern (AFib) mit und ohne CRT mit biventrikulärer (BV) Stimulation.
Methoden: Bei 49 HF P mit AFib (Alter 63,9 ± 10,8 Jahre; 43 Männer und 6 Frauen) mit New York Heart Association (NYHA) Klasse 2,9 ± 0,4, LV Ejektionsfraktion 26,03 ± 7,99 % und QRS-Dauer (QRSD) 143,69 ± 35,62 ms wurde das elektrische IVD als Intervall zwischen Beginn des QRS-Komplexes im Oberflächen EKG und Beginn des LV Signals im transösophagealen LV EKG bei 31 HF P mit AFib und bei 18 HF P mit AFib und CRT präoperativ bestimmt. Das fokussierte bipolare LV EKG wurde mittels Osypka TO Sonde mit halbkugelförmigen Elektroden in Höhe des maximalen LV Signals registriert.
Ergebnisse: Bei 31 HF P mit AFib betrugen QRSD 135,48 ± 38,78 ms, IVD 49,55 ± 26,38 ms, QRSD-IVD-Verhältnis 3,12 ± 1,11 und das IVD korrelierte mit der QRSD (r=0,75, P<0,001) und dem QRSD-IVD-Verhältnis (r=-0,67, P<0,001) (Fig.). Bei 18 HF P mit AFib und CRT Defibrillator betrugen QRSD 157,83 ± 24,38 ms, IVD 61,94 ± 26,88 ms, QRSD-IVD-Verhältnis 3,12 ± 1,89 und das IVD korrelierte mit der QRSD (r=0,47, P=0,049) und dem QRSD-IVD-Verhältnis (r=-0,73, P<0,001). Bei 72,2 % CRT Responder (R) (n=13) betrugen QRSD 158,15 ± 22,4 ms, IVD 64,23 ± 24,62 ms, QRSD-IVD-Verhältnis 2,82 ± 1,32 und das IVD korrelierte mit der QRSD (r=0,57, P=0,043) und dem QRSD-IVD-Verhältnis (r=-0,76, P=0,0024). Bei 27,8 % CRT Non-Responder (NR) (n=5) betrugen QRSD 157 ± 31,94 ms, IVD 56 ± 34,52 ms, QRSD-IVD-Verhältnis 3,88 ± 2,98 und das IVD korrelierte nicht mit der QRSD (r=0,33, P=0,591) und dem QRSD-IVD-Verhältnis (r=-0,732, P=0,159). Die CRT R verbesserten sich in der NYHA Klasse von 3 ± 0,2 auf 2,2 ± 0,3 (P<0,001) während 15,3 ± 13,1 Monaten BV Stimulation. Bei 15 CRT NR kam es zu keiner Verbesserung der NYHA Klasse von 3 auf 3,3 ± 0,97 (P=0,529) während 18,8 ± 20,7 Monaten BV Stimulation.
Schlussfolgerungen: Das transösophageale LV EKG ermöglicht bei HF-P mit AFib die nichtinvasive Messung des elektrischen IVD präoperativ vor CRT. IVD und QRSD-IVD-Verhältnis sind möglicherweise einfach anwendbare Parameter zur Vorhersage von CRT R und CRT NR bei P mit AFib.