Background: Transesophageal left atrial (LA) pacing and transesophageal LA ECG recording are semi-invasive techniques for diagnostic and therapy of supraventricular rhythm disturbance. Cardiac resynchronization therapy (CRT) with right atrial (RA) sensed biventricular pacing is an established therapy for heart failure patients with reduced left ventricular (LV) ejection fraction, sinus rhythm and interventricular electrical desynchronization.
Purpose: The aim of the study was to evaluate electromagnetic and voltage pacing fields of the combination of RA pacing, LA pacing and biventricular pacing in patients with long interatrial and interventricular electrical desynchronization.
Methods: The modelling and electromagnetic simulations of transesophageal LA pacing in combination with RA pacing and biventricular pacing would be staged and analyzed with the CST (Computer Simulation Technology) software. Different electrodes were modelled in order to simulate different types of bipolar pacing in the 3D-CAD Offenburg heart rhythm model: The bipolar Solid S (Biotronik) electrode where modelled for RA pacing and right ventricular (RV) pacing, Attain 4194 (Medtronic) for LV pacing and TO8 (Osypka) multipolar esophageal electrode with hemispheric electrodes for LA pacing.
Results: The pacemaker amplitudes for the electromagnetic pacing simulations were performed with 3 V for RA pacing, 1.5 V for RV pacing, 50 V for LA pacing and 3V for LV pacing with pacing impulse duration of 0.5 ms for RA, RV and LV pacing and 10 ms for LA pacing. The atrioventricular pacing delay after RA pacing was 140 ms. The different pacing modes AAI, VVI, DDD, DDD0V and DDD0D were evaluated for the analysis of the electric pacing field propagation of pacemaker, CRT and LA pacing. The pacing results were compared at minimum (LOW) and maximum (HIGH) parameter settings. While the LOW setting produced fewer tetrahedral and more inaccurate results, the HIGH setting produced many tetrahedral and therefore more accurate results.
Conclusions: The simulation of the combination of transesophageal LA pacing with RA sensed biventricular pacing is possible with the Offenburg heart rhythm model. The new temporary 4-chamber pacing method may be additional useful method in CRT non-responders with long interatrial electrical delay.
Die transösophageale Neurostimulation ist eine neue Therapieform und könnte unter anderem zur Schmerzlinderung während einer transösophagealen Linksherzstimulation angewendet werden. Sie ist in die Kategorie der Rückenmarksstimulation (SCS) einzuordnen, die die meist verwendete Technik der Neurostimulation ist. Die derzeit auf dem Markt vorhandenen Ösophaguskatheter werden bei einer elektrophysiologischen Untersuchung mit Ablation und transösophagealer Echokardiographie zur Temperaturüberwachung eingesetzt. Das Ziel dieser Arbeit war, das vorhandene Offenburger Herzrhythmusmodell, um die Wirbelsäule zu erweitern, einen neuen Ösophagus-Elektroden- Katheter für die transösophageale elektrische Stimulation des Rückenmarks zu modellieren und mittels 3D-Computer-Simulationen auf Ihre Wirksamkeit zu untersuchen.
Spinal cord stimulation (SCS) is the most commonly used technique of neurostimulation. It involves the stimulation of the spinal cord and is therefore used to treat chronic pain. The existing esophageal catheters are used for temperature monitoring during an electrophysiology study with ablation and transesophageal echocardiography. The aim of the study was to model the spine and new esophageal electrodes for the transesophageal electrical pacing of the spinal cord, and to integrate them in the Offenburg heart rhythm model for the static and dynamic simulation of transesophageal neurostimulation. The modeling and simulation were both performed with the electromagnetic and thermal simulation software CST (Computer Simulation Technology, Darmstadt). Two new esophageal catheters were modelled as well as a thoracic spine based on the dimensions of a human skeleton. The simulation of directed transesophageal neurostimulation is performed using the esophageal balloon catheter with an electric pacing potential of 5 V and a trapezoidal signal. A potential of 4.33 V can be measured directly at the electrode, 3.71 V in the myocardium at a depth of 2 mm, 2.68 V in the thoracic vertebra at a depth of 10 mm, 2.1 V in the thoracic vertebra at a depth of 50 mm and 2.09 V in the spinal cord at a depth of 70 mm. The relation between the voltage delivered to the electrodes and the voltage applied to the spinal cord is linear. Virtual heart rhythm and catheter models as well as the simulation of electrical pacing fields and electrical sensing fields allow the static and dynamic simulation of directed transesophageal electrical pacing of the spinal cord. The 3D simulation of the electrical sensing and pacing fields may be used to optimize transesophageal neurostimulation.
Spinal cord stimulation (SCS) is the most commonly used technique of neurostimulation. It involves the stimulation of the spinal cord and is therefore used to treat chronic pain. The existing esophageal catheters are used for temperature monitoring during an electrophysiology study with ablation and transesophageal echocardiography. The aim of the study was to model the spine and new esophageal electrodes for the transesophageal electrical pacing of the spinal cord, and to integrate them in the Offenburg heart rhythm model for the static and dynamic simulation of transesophageal neurostimulation. The modeling and simulation were both performed with the electromagnetic and thermal simulation software CST (Computer Simulation Technology, Darmstadt). Two new esophageal catheters were modelled as well as a thoracic spine based on the dimensions of a human skeleton. The simulation of directed transesophageal neurostimulation is performed using the esophageal balloon catheter with an electric pacing potential of 5 V and a trapezoidal signal. A potential of 4.33 V can be measured directly at the electrode, 3.71 V in the myocardium at a depth of 2 mm, 2.68 V in the thoracic vertebra at a depth of 10 mm, 2.1 V in the thoracic vertebra at a depth of 50 mm and 2.09 V in the spinal cord at a depth of 70 mm. The relation between the voltage delivered to the electrodes and the voltage applied to the spinal cord is linear. Virtual heart rhythm and catheter models as well as the simulation of electrical pacing fields and electrical sensing fields allow the static and dynamic simulation of directed transesophageal electrical pacing of the spinal cord. The 3D simulation of the electrical sensing and pacing fields may be used to optimize transesophageal neurostimulation.