Background: The electrical field (E-field) of the biventricular (BV) stimulation is important for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex. The 3D modeling allows the simulation of CRT and high frequency (HF) ablation.
Purpose: The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of atrial and BV stimulation and high frequency (HF) ablation in atrial fibrillation (AF).
Methods: The modeling and simulation was carried out using the electromagnetic simulation software CST (CST Darmstadt). Five multipolar left ventricular (LV) electrodes, one epicardial LV electrode, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modeled. Selected electrodes were integrated into the Offenburg heart rhythm model for the electrical field simulation. The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
Results: The right atrial stimulation was performed with an amplitude of 1.5 V with a pulse width of 0.5. The far-field potentials generated by the atrial stimulation were perceived by the right and left ventricular electrode. The far-field potential at a distance of 1 mm from the right ventricular electrode tip was 36.1 mV. The far-field potential at a distance of 1 mm from the left ventricular electrode tip was measured with 37.1 mV. The RV and LV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode with a pulse width of 0.5 ms each. The far-field potentials generated by the BV stimulations could be perceived by the RA electrode. The far-field potential at the RA electrode tip was 32.86 mV. AV node ablation was simulated with an applied power of 5 W at 420 kHz and 10 W at 500 kHz at the distal 8 mm ablation electrode.
Conclusions: Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF. The 3D simulation of the electrical field and temperature profile may be used to optimize the CRT and AF ablation.
Cardiac resynchronization therapy with atrioventricular and interventricular delay optimized biventricular pacing is an established therapy for symptomatic heart failure patients with prolongation of QRS duration, left bundle branch block and reduced left ventricular ejection fraction. The aim of the investigation was to evaluate right atrial, right ventricular and left ventricular electrical signals of implantable electronic cardiac devices with and without signal averaging technique with novel LabVIEW software. Electrical interatrial conduction delay and inter-ventricular conduction delay may be useful parameters to evaluate electrical atrial and ventricular desynchronization in heart failure patients.
Abstract: Electric field of biventricular (BV) pacing, left ventricular (LV) electrode position and electrical interventricular desynchronization are important parameters for successful cardiac resynchronization therapy (CRT) in patients with heart failure, sinus rhythm and reduced LV ejection fraction. The aim of the study was to evaluate electric pacing field of transesophageal left atrial (LA) pacing and BV pacing with 3D heart rhythm simulation. Bipolar right atrial (RA), right ventricular (RV), LV electrodes and multipolar hemispherical esophageal LA electrodes were modeled with CST (Computer Simulation Technology, Darmstadt). Electric pacing field were simulated with bipolar RA and RV pacing with Solid S (Biotronik) electrode, bipolar LV pacing with Attain 4194 (Medtronic) electrode and bipolar LA pacing with TO8 (Osypka) esophageal electrode. 3D heart rhythm model with esophagus allowed electric pacing field simulation of 4-chamber pacing with bipolar intracardiac RA, RV, LV pacing and bipolar transesophageal LA pacing. The pacing amplitudes were 3V RA pacing amplitude, 50V LA pacing amplitude, 1.5V RV pacing amplitude and 3V LV pacing amplitude with 0.5ms pacing pulse duration. The atrioventricular delay between RA pacing and BV pacing was 140ms atrioventricular pacing delay and simultaneous RV and LV pacing. Electric pacing fields were simulated during the different pacing modes AAI, VVI, DDD and DDD0V. The intracardiac far-field pacing potentials were evaluated with intracardiac electrodes and a distance of 1mm from the electrodes with RA electrode 1.104V, RV electrode 0.703V and LV electrode 1.32V. The transesophageal far-field pacing potential was evaluated with transesophageal electrode and a distance of 10mm from the elelctrode with LA electrode 6.076V. Heart rhythm model simulation with esophagus allows evaluation of electric pacing fields in AAI, VVI, DDD, DDD0V and DDD0D pacing modes. Electric pacing field of RA, RV and LV pacing in combination with LA pacing may additional useful pacing mode in CRT non-responders.
The high frequency (HF) catheter ablation is the gold standard for the therapy of many cardiac tachyarrhythmias, such as atrioventricular node re-entry tachycardia (AVNRT), atrioventricular re-entry tachycardia (AVRT) or atrial flutter (AFL). The aim of the study was to simulate the HF ablation of AVNRT, AVRT, AFL and its heat propagation in reference to the supplied power with different electrode material and electrode size. The modeling and simulation were performed with the thermal and electromagnetic simulation software CST® (Computer Simulation Technology, Darmstadt). The modeling and simulation were carried out using ablation catheters with 4 mm tip electrode and 8 mm tip electrode with different electrode materials. Both electrode types were made of platinum and gold respectively. For the measurement of the heat propagation in the heart tissue, the catheters were integrated in the Offenburg heart rhythm model. The HF ablation procedures were performed with the 4 mm platinum tip electrode, with an application duration of 45 seconds and a power output of 40 watts. The HF ablation of the atrioventricular node slow pathway produced a maximum temperature of 66.33 °C. The Kent bundle HF ablation in the left atrium achieved a maximum temperature of 67.14 °C. The HF ablation of the right atrial isthmus resulted 65.96 °C. The 8 mm distal platinum tip electrode and a power output of 60 watts reached 72.85 °C. The 8 mm distal gold tip electrode and a power output of 60 watt reached 64.66 °C, due to the improved thermal conductivity of gold. Virtual heart and ablation electrode models allow the static and dynamic simulation of HF ablation with different electrode material and electrode size. The 3D simulation of the temperature profile may be used to optimize the AVNRT, AVRT and AFL HF ablation.