Refine
Year of publication
Document Type
- Conference Proceeding (69)
- Article (reviewed) (14)
- Patent (11)
- Contribution to a Periodical (6)
- Article (unreviewed) (2)
Conference Type
- Konferenz-Abstract (53)
- Konferenzartikel (12)
- Konferenz-Poster (4)
Language
- English (76)
- German (24)
- Other language (1)
- Multiple languages (1)
Has Fulltext
- no (102) (remove)
Is part of the Bibliography
- yes (102) (remove)
Keywords
- CST (7)
- HF-Ablation (7)
- Herzrhythmusmodell (6)
- CRT (5)
- Heart rhythm model (5)
- Herzkrankheit (5)
- Modeling and simulation (5)
- Synchronisierung (4)
- Cardiac resynchronization therapy (3)
- Herzschrittmacher (3)
Institute
Open Access
- Open Access (65)
- Closed Access (29)
- Bronze (4)
- Closed (2)
Introduction: Cardiac resynchronisation therapy (CRT) with atrioventricular (AV) and interventricular (VV) optimized biventricular pacing (BV) is an established therapy for heart failure (HF) patients. The aim of the study was to compare AV and VV delay optimization with cardiac output (CO), cardiac index (CI), contractility index (IC) and acceleration index (ACI) impedance cardiographic (ICG) methods in CRT.
Methods: 15 HF patients (age 66 ± 10 years; 2 females, 13 males) in New York Heart Association (NYHA) class 3.1 ± 0.4, left ventricular (LV) ejection fraction 21.3 ± 7.8 % and QRS duration 176.1 ± 31.7 ms underwent AV and VV delay optimization with CO, CI, IC and ACI (Cardioscreen ®, Medis GmbH, Ilmenau, Germany) at different AV and VV delay BV pacing settings versus right ventricular (RV) pacing one day after implantation of a CRT device.
Results: Optimal AV delay after atrial sensing was 108.6 ± 20.3 ms (n=14) and optimal AV delay after atrial pacing 190 ± 14.1 ms (n=2) with AV delay range from 80 ms to 200 ms. Optimal VV delay was -12.3 ± 25.9 ms left ventricular before RV pacing. RV versus BV pacing mode resulted in improvement of CO from 3.4 ± 1.2 l/min to 4.4 ± 1.4 l/min (p<0.001), CI from 1.8 ± 0.64 l/min/m² to 2.4 ± 0.78 l/min/m² (p<0.001), IC from 0.028 ± 0.011 1/s to 0.036 ± 0.013 1/s (p<0.001) and ACI from 0.667 ± 0.227 1/s² to 0.834 ± 0.282 1/s² (p<0.002). During 34 ± 26 month BV pacing, the NYHA class improved from 3.1 ± 0.4 to 2.1 ± 0.4 (p<0.001).
Conclusion: AV and VV delay optimized BV pacing acutely improve hemodynamic parameters of transthoracic ICG and their NYHA class during long-term follow-up. ICG may be a simple and useful technique to optimize AV and VV delay in CRT.
Introduction: Cardiac resynchronisation therapy (CRT) with atrioventricular (AV) and interventricular (VV) optimized biventricular pacing (BV) is an established therapy for heart failure (HF) patients with electrical interventricular conduction delay (IVCD). The aim of the study was to compare AV and VV delay optimization with cardiac output (CO) and acceleration index (ACI) impedance cardiographic (ICG) methods.
Methods: HF patients with IVCD 86.8 ± 33 ms (n=15, age 66 ± 10 years; 2 females, 13 males), New York Heart Association (NYHA) functional class 3.1 ± 0.4, left ventricular (LV) ejection fraction 21.3 ± 7.8 % and QRS duration 176.1 ± 31.7 ms underwent AV and VV delay optimization with CO and ACI methods (Cardioscreen, Medis GmbH, Ilmenau, Germany). After evaluation of optimal AV delay, we evaluated optimal VV delay during simultaneous LV and right ventricular (RV) pacing (LV=RV), LV before RV pacing (LV-RV) and RV before LV pacing (RV-LV).
Results: Optimal VV delay was -12.3 ± 25.9 ms LV-RV pacing with VV delay range from -80 ms LV-RV pacing to +20 ms RV-LV pacing and RV=LV pacing. Optimal AV delay after atrial sensing was 108.6 ± 20.3 ms (n=14) and optimal AV delay after atrial pacing 190 ± 14.1 ms (n=2) with AV delay range from 80 ms to 200 ms. RV versus BV pacing mode resulted in improvement of CO from 3.4 ± 1.2 l/min to 4.4 ± 1.4 l/min (p<0.001) and ACI from 0.667 ± 0.227 1/s² to 0.834 ± 0.282 1/s² (p<0.002). During 34 ± 26 month BV pacing, the NYHA class improved from 3.1 ± 0.4 to 2.1 ± 0.4 (p<0.001).
Conclusion: AV and VV delay optimized BV pacing acutely improve ICG CO and ACI and their NYHA class during long-term follow-up. ICG may be a simple and useful technique to optimize AV and VV delay in CRT.
Capture threshold (CT) for transesophageal left atrial (LA) pacing (TLAP) and transesophageal left ventricular (LV) pacing (TLVP) with conventional cylindrical electrodes (CE) are higher than TLAP feeling threshold (FT). Purpose of the study was to evaluate focused TLAP CT and FT for supraventricular tachycardia (SVT) initiation and focused TLVP CT for cardiac resynchronisation therapy (CRT) simulation.
Methods: SVT initiation in patients (P) with palpitations (n=49, age 47 ± 17 years) was analysed during spontaneous rhythm and during focused bipolar TLAP with atrial constant current stimulus output, distal CE and three or seven 6 mm hemispherical electrodes (HE) (TO, Osypka AG, Rheinfelden, Germany). CRT simulation in heart failure P (n=75, age 62 ± 11 years) was evaluated by focused bipolar TLAP and/or TLVP with ventricular constant voltage stimulus output and different pacing mode.
Results: Focused electrical pacing field between CE and HE (n=28) allowed low threshold TLAP with 8.0 ± 2.6 mA CT at 9.9 ms stimulus duration (SD) which was lower than 9.2 ± 4.5 mA FT at 9.9 ms SD. Focused electrical pacing field between HE and HE (n=21) allowed low threshold TLAP with 8.1 ± 2.2 mA CT at 9.9 ms SD which was lower than 9.8 ± 5.0 mA FT at 9.9 ms SD. SVT initiation by programmed AAI TLAP was possible in 23 P and not possible in 26 P. CRT simulation was evaluated with TLAP and TLVP with VAT, D00 and V00 pacing mode and 95.5 ± 10.9 V TLVP CT at 4.0 ms SD.
Conclusions: Programmed focused AAI TLAP allowed initiation of SVT with very low CT and high FT and focused electrical pacing field between CE-HE and HE-HE.CRT simulation with focused TLAP and/or TLVP with VAT, D00 and V00 pacing mode may be a useful technique to detect responders to CRT.
Cardiac resynchronization therapy (CRT) with biventricular (BV) pacing is an established therapy in approximately two-thirds of symptomatic heart failure (HF) patients (P) with left bundle branch block (LBBB). The aim of this study was to evaluate left atrial (LA) conduction delay (LACD) and left ventricular (LV) conduction delay (LVCD) using pre-implantational transesophageal electrocardiography (ECG) in sinus rhythm (SR) CRT responder (R) and non-responder (NR).
Methods: SR HF P (n=52, age 63.6±10.4 years; 6 females, 46 males) with New York Heart Association (NYHA) class 3.0±0.2, 24.4±7.1 % LV ejection fraction and 171.2±37.6 ms QRS duration (QRSD) were measured by bipolar filtered transesophageal LA and LV ECG recording with hemispherical electrodes (HE) TO catheter (Osypka AG, Rheinfelden, Germany). LACD was measured between onset of P-wave in the surface ECG and onset of LA deflection in the LA ECG. LVCD was measured between onset of QRS in the surface ECG and onset of LV deflection in the LV ECG.
Results: There were 78.8 % SR CRT R (n=41) with 171.2±36.9 ms QRSD, 73.3±25.7 ms LACD, 80.0±24.0 ms LVCD and 2.3±0.5 QRSD-LVCD-ratio. SR CRT R QRSD correlated with LACD (r=0.688, P<0.001) and LVCD (r=0.699, P<0.001). There were 21.2 % SR CRT NR (n=11) with 153.4±22.4 ms QRSD (P=0.133), 69.8±24.8 ms LACD (n=6, P=0.767), 54.2±31.0 ms LVCD (P<0.0046) and 3.9±2.5 QRSD-LVCD-ratio (P<0.001). SR CRT NR QRSD not corre-lated with IACD (r=-0.218, P=0.678) and IVCD (r=0.042, P=0.903). During a 22.8±21.3 month CRT follow-up, the CRT R NYHA class improved from 3.1±0.3 to 1.9±0.3 (P<0.001). In CRT NR, NYHA class not improved (2.9±0.4 to 2.9±0.2, P=1) during 11.2±9.8 months BV pacing.
Conclusions: Transesophageal LA and LV ECG with HE can be utilized to analyse LACD and LVCD in HF P. Pre-implantational LVCD and QRSD-LVCD-ratio may be additional useful parameters to improve P selection for SR CRT.
Introduction: Cardiac resynchronization therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients (P) with ventricular desynchronisation, but not all patients improved clinically. Aim of this study was to evaluate electrical intra-left ventricular conduction delay (LVCD) and interventricular conduction delay (IVCD), to better select patients for CRT.
Methods: 65 HF patients (age 63.4 ± 10.6 years; 7 females, 58 males) with New York Heart Association (NYHA) class 3 ± 0.2, 24.4 ± 6.7 % left ventricular (LV) ejection fraction and 167.4 ± 35.6 ms QRSD were included. Esophageal TO Osypka focused hemispherical electrodes catheter was perorally applied in position of maximum LV deflection to measure LVCD between onset and offset of LV deflection and IVCD between earliest onset of QRS in the 12-channel surface ECG and onset of LV deflection in the focused bipolar transesophageal LV electrogram.
Results: There were 50 responders with LVCD of 76.5 ± 20.4 ms, IVCD of 80.5 ± 26.1 ms (P=0.34) and QRSD of 171 ± 37.7 ms. 15 non-responders had longer LVCD of 90 ± 28.5 ms (P = 0.045), shorter IVCD of 50.1 ± 29.1 ms (P < 0.001) and QRSD of 155.3 ± 25 ms (P=0.14). During 21.3 ± 20.3 month BV pacing follow-up, the responder`s NYHA classes improved from 3 ± 0.2 to 2. ± 0.3 (P < 0.001) whereas the non-responders NYHA classes did not improve from 3 ± 0.2 to 2.9 ± 0.3 (P = 0.43) during 15.7 ± 13.9 month BV pacing follow-up (53 Boston, 10 Medtronic and 2 St. Jude CRT devices).
Conclusion: Determination of electrical LVCD and IVCD by focused bipolar transesophageal LV electrogram recording may be an additional useful technique to improve patient selection for CRT.
Transösophageales interventrikuläres Delay bei Vorhofflimmern und kardialer Resynchronisation
(2013)
Die transösophageale linksventrikuläre Elektrokardiographie ermöglicht die Evaluierung der elektrischen ventrikulären Desynchronisation im Rahmen der kardialen Resynchronisationstherapie der Herzinsuffizienz. Das Ziel der Untersuchung besteht in der präoperativen Abschätzung des transösophagealen interventrikulären Delays bei Vorhofflimmern und kardialer Resynchronisationstherapie. Bei Patienten mit Vorhofflimmern, Herzinsuffizienz New York Heart Association Klasse 3,0 ± 0,2 und QRS-Dauer 159,6 ± 23,9 ms wurde das fokusierte transösophageale linksventrikuläre EKG abgeleitet. Die kardiale Resynchronisationstherapie Responder QRS-Dauer korrelierte mit dem transösophagealen interventrikulären Delay bei Vorhofflimmern.