Introduction: Patient selection for cardiac resynchronization therapy (CRT) requires quantification of left ventricular conduction delay (LVCD). After implantation of biventricular pacing systems, individual AV delay (AVD) programming is essential to ensure hemodynamic response. To exclude adverse effects, AVD should exceed individual implant-related interatrial conduction times (IACT). As result of a pilot study, we proposed the development of a programmer-based transoesophageal left heart electrogram (LHE) recording to simplify both, LVCD and IACT measurement. This feature was implemented into the Biotronik ICS3000 programmer simultaneously with 3-channel surface ECG.
Methods: A 5F oesophageal electrode was perorally applied in 44 heart failure CRT-D patients (34m, 10f, 65±8 yrs., QRS=162±21ms). In position of maximum left ventricular deflection, oesophageal LVCD was measured between onsets of QRS in surface ECG and oesophageal left ventricular deflection. Then, in position of maximum left atrial deflection (LA), IACT in VDD operation (As-LA) was calculated by difference between programmed AV delay and the measured interval from onset of left atrial deflection to ventricular stimulus in the oesophageal electrogram. IACT in DDD operation (Ap-LA) was measured between atrial stimulus and LA..
Results: LVCD of the CRT patients was characterized by a minimum of 47ms with mean of 69±23ms. As-LA and Ap-LA were found to be 41±23ms and 125±25ms, resp., at mean. In 7 patients (15,9%), IACT measurement in DDD operation uncovered adverse AVD if left in factory settings. In this cases, Ap-LA exceeded the factory AVD. In 6 patients (13,6%), IACT in VDD operation was less than or equal 10ms indicating the need for short AVD.
Conclusion: Response to CRT requires distinct LVCD and AVD optimization. The ICS3000 oesophageal LHE feature can be utilized to measure LVCD in order to justify selection for CRT. IACT measurement simplifies AV delay optimization in patients with CRT systems irrespective of their make and model.
AV delay (AVD) optimization is mandatory in cardiac resynchronization (CRT) for heart failure. Several time consuming methods exist. We initialized development of left-atrial electrogram (LAE) feature for Biotronik ICS3000 programmer. It can be utilized to approximate optimal AV delay in CRT patients with pacing systems irrespective of make and model. Using this feature, we studied the share of interatrial conduction intervals (IACT) on individual echo AVD in 45 CRT patients (34m, 11f, mean age 69±6yrs.). The percentage of IACT on optimal echo AVD resulted in44.5±22.1% for VDD and 70.7±10.9% for DDD operation. In all patients, optimal echo AVDs exceeded the individual IACT by a duration of 52.5±33.3ms (p<0.001), at mean. Therefore, if AV delay optimization is not possible or not practicable in CRT patients, AVD should be approximated by individually measuring IACT and adding about 50ms.