Background: R-wave synchronised atrial pacing is an effective temporary pacing
therapy in infants with postoperative junctional ectopic tachycardia. In the technique
currently used, adverse short or long intervals between atrial pacing and ventricular
sensing (AP–VS) may be observed during routine clinical practice.
Objectives: The aim of the study was to analyse outcomes of R-wave synchronised
atrial pacing and the relationship between maximum tracking rates and AP–VS
intervals.
Methods: Calculated AP–VS intervals were compared with those predicted by experienced
pediatric cardiologist.
Results: A maximum tracking rate (MTR) set 10 bpm higher than the heart rate (HR)
may result in undesirable short AP–VS intervals (minimum 83 ms). A MTR set 20 bpm
above the HR is the hemodynamically better choice (minimum 96 ms). Effects of either
setting on the AP–VS interval could not be predicted by experienced observers. In our
newly proposed technique the AP–VS interval approaches 95 ms for HR > 210 bpm
and 130 ms for HR < 130 bpm. The progression is linear and decreases strictly
(− 0.4 ms/bpm) between the two extreme levels.
Conclusions: Adjusting the AP–VS interval in the currently used technique is complex
and may imply unfavorable pacemaker settings. A new pacemaker design is advisable
to allow direct control of the AP–VS interval.