Introduction: To simplify AV delay (AVD) optimization in cardiac resynchronization therapy (CRT), we reported that the hemodynamically optimal AVD for VDD and DDD mode CRT pacing can be approximated by individually measuring implant-related interatrial conduction intervals (IACT) in oesophageal electrogram (LAE) and adding about 50ms. The programmer-based St Jude QuickOpt algorithm is utilizing this finding. By automatically measuring IACT in VDD operation, it predicts the sensed AVD by adding either 30ms or 60ms. Paced AVD is strictly 50ms longer than sensed AVD. As consequence of those variations, several studies identified distinct inaccuracies of QuickOpt. Therefore, we aimed to seek for better approaches to automate AVD optimization.
Methods: In a study of 35 heart failure patients (27m, 8f, age: 67±8y) with Insync III Marquis CRT-D systems we recorded telemetric electrograms between left ventricular electrode and superior vena cava shock coil (LVtip/SVC = LVCE) simultaneously with LAE. By LVCE we measured intervals As-Pe in VDD and Ap-Pe in DDD operation between right atrial sense-event (As) or atrial stimulus (Ap), resp., and end of the atrial activity (Pe). As-Pe and Ap-Pe were compared with As-LA an Ap-LA in LAE, respectively.
Results: End of the left atrial activity in LVCE could clearly be recognized in 35/35 patients in VDD and 29/35 patients in DDD operation. We found mean intervals As-LA of 40.2±24.5ms and Ap-LA of 124.3±20.6ms. As-Pe was 94.8±24.1ms and Ap-Pe was 181.1±17.8ms. Analyzing the sums of As-LA + 50ms with duration of As-Pe and Ap-LA + 50ms with duration of Ap-Pe, the differences were 4.7±9.2ms and 4.2±8.6ms, resp., only. Thus, hemodynamically optimal timing of the ventricular stimulus can be triggered by automatically detecting Pe in LVCE.
Conclusion: Based on minimal deviations between LAE and LVCE approach, we proposed companies to utilize the LVCE in order to automate individual AVD optimization in CRT pacing.
Introduction: Patient selection for cardiac resynchronization therapy (CRT) requires quantification of left ventricular conduction delay (LVCD). After implantation of biventricular pacing systems, individual AV delay (AVD) programming is essential to ensure hemodynamic response. To exclude adverse effects, AVD should exceed individual implant-related interatrial conduction times (IACT). As result of a pilot study, we proposed the development of a programmer-based transoesophageal left heart electrogram (LHE) recording to simplify both, LVCD and IACT measurement. This feature was implemented into the Biotronik ICS3000 programmer simultaneously with 3-channel surface ECG.
Methods: A 5F oesophageal electrode was perorally applied in 44 heart failure CRT-D patients (34m, 10f, 65±8 yrs., QRS=162±21ms). In position of maximum left ventricular deflection, oesophageal LVCD was measured between onsets of QRS in surface ECG and oesophageal left ventricular deflection. Then, in position of maximum left atrial deflection (LA), IACT in VDD operation (As-LA) was calculated by difference between programmed AV delay and the measured interval from onset of left atrial deflection to ventricular stimulus in the oesophageal electrogram. IACT in DDD operation (Ap-LA) was measured between atrial stimulus and LA..
Results: LVCD of the CRT patients was characterized by a minimum of 47ms with mean of 69±23ms. As-LA and Ap-LA were found to be 41±23ms and 125±25ms, resp., at mean. In 7 patients (15,9%), IACT measurement in DDD operation uncovered adverse AVD if left in factory settings. In this cases, Ap-LA exceeded the factory AVD. In 6 patients (13,6%), IACT in VDD operation was less than or equal 10ms indicating the need for short AVD.
Conclusion: Response to CRT requires distinct LVCD and AVD optimization. The ICS3000 oesophageal LHE feature can be utilized to measure LVCD in order to justify selection for CRT. IACT measurement simplifies AV delay optimization in patients with CRT systems irrespective of their make and model.
In-vivo and in-vitro comparison of implant-based CRT optimization - What provide new algorithms?
(2011)
Introduction: In cardiac resynchronization therapy (CRT), individual AV delay (AVD) optimization can effectively increase hemodynamics and reduce non-responder rate. Accurate, automatic and easily comprehensible algorithms for the follow-up are desirable. QuickOpt is the first attempt of a semi-automatic intracardiac electrogram (IEGM) based AVD algorithm. We aimed to compare its accuracy and usefulness by in-vitro and in-vivo studies.
Methods: Using the programmable ARSI-4 four-chamber heart rhythm and IEGM simulator (HKP, Germany), the QuickOpt feature of an Epic HF system (St. Jude, USA) was tested in-vitro by simulated atrial IEGM amplitudes between 0.3 and 3.5mV during both, manual and automatic atrial sensing between 0.2 and 1.0mV. Subsequently, in 21 heart failure patients with implanted biventricular defibrillators, QuickOpt was performed in-vivo. Results of the algorithm for VDD and DDD stimulation were compared with echo AV delay optimization.
Results: In-vitro simulations demonstrated a QuickOpt measuring accuracy of ± 8ms. Depending on atrial IEGM amplitude, the algorithm proposed optimal AVD between 90 and 150ms for VDD and between 140 and 200ms for DDD operation, respectively. In-vivo, QuickOpt difference between individual AVD in DDD and VDD mode was either 50ms (20pts) or 40ms (1pt). QuickOpt and echo AVD differed by 41 ± 25ms (7 – 90ms) in VDD and by 18 ± 24ms (17-50ms) in DDD operation. Individual echo AVD difference between both modes was 73 ± 20ms (30-100ms).
Conclusion: The study demonstrates the value of in-vitro studies. It predicted QuickOpt deficiencies regarding IEGM amplitude dependent AVD proposals constrained to fixed individual differences between DDD and VDD mode. Consequently, in-vivo, the algorithm provided AVD of predominantly longer duration than echo in both modes. Accepting echo individualization as gold standard, QuickOpt should not be used alone to optimize AVD in CRT patients.
Patients with focal ventricular tachycardia are at risk of hemodynamic failure and if no treatment is provided the mortality rate can exceed 30%. Therefore, medical professionals must be adequately trained in the management of these conditions. To achieve the best treatment, the origin of the abnormality should be known, as well as the course of the disease. This study provides an opportunity to visualize various focal ventricular tachycardias using the Offenburg cardiac rhythm model.
Background: A disturbed synchronization of the ventricular contraction can cause a highly developed systolic heart failure in affected patients, which can often be explained by a diseased left bundle branch block (LBBB). If medication remains unresponsive, the concerned patients will be treated with a cardiac resynchronization therapy (CRT) system. The aim of this study was to integrate His bundle pacing into the Offenburg heart rhythm model in order to visualize the electrical pacing field generated by His bundle pacing.
Methods: Modelling and electrical field simulation activities were performed with the software CST (Computer Simulation Technology) from Dessault Systèms. CRT with biventricular pacing is to be achieved by an apical right ventricular electrode and an additional left ventricular electrode, which is floated into the coronary vein sinus. This conventional type of biventricular pacing leads to a reduction of the left ventricular ejection fraction. Furthermore, the non-responder rate of the CRT therapy is about one third of the CRT patients.
Results: His bundle pacing represents a physiological alternative to conventional cardiac pacing and cardiac resynchronization. An electrode implanted in the His bundle emits a stronger electrical pacing field than the electrical pacing field of conventional cardiac pacemakers. The pacing of the His bundle was performed by the Medtronic Select Secure 3830 electrode with pacing voltage amplitudes of 3 V, 2 V and 1.5 V in combination with a pacing pulse duration of 1 ms.
Conclusions: Compared to conventional cardiac pacemaker pacing, His bundle pacing is capable of bridging LBBB conduction disorders in the left ventricle. The His bundle pacing electrical field is able to spread via the physiological pathway in the right and left ventricles for CRT with a narrow QRS-complex in the surface ECG.
Background: Pulmonary vein isolation (PVI) using cryoballoon catheters are a recognized method for the treatment of atrial fibrillation (AF). This method offers shorter treatment duration in contrast to the classical therapy with high-frequency (HF) ablation.
Purpose: The aim of this study was to integrate different cryoballoon catheters and a HF catheter into a heart rhythm model and to compare them by means of static and dynamic electromagnetic and thermal simulation in use under AF.
Methods: The cryoballoon catheters from Medtronic and the HF ablation catheter from Osypka were modelled virtually with the aid of manufacturer specifications and the CST (Computer Simulation Technology, Darmstadt) simulation program. The cryoballoon catheter was located in the lower left pulmonary vein of the virtual heart rhythm model for the realization of pulmonary vein isolation (PVI) by cryoenergy. The simulated temperature at the balloon surface was -50°C during the simulation.
Results: During a simulated 20 second application of a cryoballoon catheter at -50°C, a temperature of -24°C was measured at a depth of 0.5 mm in the myocardium. At a depth of 1 mm the temperature was -3°C, at 2 mm depth 18°C and at 3 mm depth 29°C. Under the 15 second application of a RF catheter with a 8 mm electrode and a power of 5 W at 420 kHz, the temperature at the tip of the electrode was 110°C. At a depth of 0.5 mm in the myocardium, the temperature was 75°C, at a depth of 1 mm 58°C, at 2 mm depth 45°C and at 3 mm depth 38°C.
Conclusions: The simulation of temperature profiles during the virtual application of several catheter models in the heart rhythm model allows the static and dynamic simulation of PVI by cryoballoon ablation and RF ablation. The three-dimensional simulation can be used to improve ablation applications by creating a model in personalized cardiac rhythm therapy from MRI or CT data of a heart and finding a favourable position for ablation of AF.
Hintergrund: Die Pulmonalvenenisolation (PVI) mit Hilfe von Kryoballonkathetern ist eine anerkannte Methode zur Behandlung von Vorhofflimmern (AF). Diese Methode bietet eine kürzere Behandlungsdauer als die klassische Therapie durch die Hochfrequenzablation (HF). Ziel dieser Studie war es, verschiedene Kryoballonkatheter, HF-Katheter und Ösophaguskatheter in ein Herzrhythmusmodell zu integrieren und mittels statischer und dynamischer Simulation elektrische und thermische Felder bei PVI unter Vorhofflimmern zu untersuchen.
Methodik: Die Modellierung und Simulation erfolgte mit der elektromagnetischen und thermischen Simulationssoftware CST (CST Darmstadt). Zwei Kryoballons, ein HF-Ablationskatheter und ein Ösophaguskatheter wurden auf der Grundlage der technischen Handbücher der Hersteller Medtronic und Osypka modelliert. Der 23 mm Kryoballon und ein kreisförmiger Mappingkatheter wurden in das Offenburger Herzrhythmusmodell integriert, insbesondere die left inferior pulmonary vein (LIPV) zur Simulation der thermischen Feldausbreitung während einer PVI. Die Simulation einer PVI mit HF-Energie wurde mit dem integrierten HF-Ablationskatheter in der Nähe der LIPV durchgeführt. Der im Herzrhythmusmodell platzierte TO8 Ösophaguskatheter ermöglichte die Ableitung linksatrialer elektrischer Felder bei AF und die Analyse thermischer Felder während PVI.
Ergebnisse: Elektrische Felder konnten bei Sinusrhythmus und AF mit einem AF-Fokus in der LIVP statisch und dynamisch im Herzen und Ösophagus simuliert werden. Bei einer simulierten 20 Sekunden Applikation eines Kryoballon-Katheters bei -50°C wurde eine Temperatur von -24°C in einer Tiefe von 0,5 mm im Myokard gemessen. In einer Tiefe von 1 mm betrug die Temperatur -3°C, bei 2 mm Tiefe 18°C und bei 3 mm Tiefe 29°C. Unter der 15 sekündigen Anwendung eines HF-Katheters mit einer 8-mm-Elektrode und einer Leistung von 5 W bei 420 kHz betrug die Temperatur an der Spitze der Elektrode 110°C. In einer Tiefe von 0,5 mm im Myokard betrug die Temperatur 75°C, in einer Tiefe von 1 mm 58°C, in einer Tiefe von 2 mm 45°C und in einer Tiefe von 3 mm 38°C. Im Ösophagus konnte bei den meisten Simulationen eine konstante Temperatur von 37°C gemessen und die Gefahr einer Ösophagus-Fistel ausgeschlossen werden. Bei Kryoablation der LIPV wurde eine Abkühlung des Ösophagus auf 30°C gemessen.
Schlussfolgerungen: Die Herzrhythmussimulation elektrischer und thermaler Felder ermöglichen mit Anwendung unterschiedlicher Herzkatheter eine statische und dynamische Simulation von PVI durch Kryoablation, HF-Ablation und Temperaturanalyse im Ösophagus. Unter Einbeziehung von MRT- oder CT-Daten können elektrische und thermale Simulationen möglicherweise zur Optimierung von PVIs genutzt werden.
Cardiac resynchronization therapy with biventricular pacing is an established therapy for heart failure patients with electrical left ventricular desynchronization. The aim of this study was to evaluate left atrial conduction delay, intra left atrial conduction delay, left ventricular conduction delay and intra left ventricular conduction delay in heart failure patients using novel signal averaging transesophageal left heart ECG software.
Methods: 8 heart failure patients with dilated cardiomyopathy (DCM), age 68 ± 9 years, New York Heart Association (NYHA) class 2.9 ± 0.2, 24.8 ± 6.7 % left ventricular ejection fraction, 188.8 ± 15.5 ms QRS duration and 8 heart failure patients with ischaemic cardiomyopathy (ICM), age 67 ± 8 years, NYHA class 2.9 ± 0.3, 32.5 ± 7.4 % left ventricular ejection fraction and 167.6 ± 19.4 ms QRS duration were analysed with transesophageal and transthoracic ECG by Bard LabDuo EP system and novel National Intruments LabView signal averaging ECG software.
Results: The electrical left atrial conduction delay was 71.3 ± 17.6 ms in ICM versus 72.3 ± 12.4 ms in DCM, intra left atrial conduction delay 66.8 ± 8.6 ms in ICM versus 63.4 ± 10.9 ms in DCM and left cardiac AV delay 180.5 ± 32.6 ms in ICM versus 152.4 ± 30.4 ms in DCM. The electrical left ventricular conduction delay was 40.9 ± 7.5 ms in ICM versus 42.6 ± 17 ms in DCM and intra left ventricular conduction delay 105.6 ± 19.3 ms in ICM versus 128.3 ± 24.1 ms in DCM.
Conclusions: Left heart signal averaging ECG can be utilized to analyse left atrial conduction delay, intra left atrial conduction delay, left ventricular conduction delay and intra left ventricular conduction delay to improve patient selection for cardiac resynchronization therapy.
Disturbances of the cardiac conduction system causing reentry mechanisms above the atrioventricular (AV) node are induced by at least one accessory pathway with different conducting properties and refractory periods. This work aims to further develop the already existing and continuously expanding Offenburg heart rhythm model to visualise the most common supraventricular reentry tachycardias to provide a better understanding of the cause of the respective reentry mechanism.