Subjects utilizing a cochlear implant (CI) in one ear and a hearing aid (HA) on the contralateral ear suffer from mismatches in stimulation timing due to different processing latencies of both devices. This device delay mismatch leads to a temporal mismatch in auditory nerve stimulation. Compensating for this auditory nerve stimulation mismatch by compensating for the device delay mismatch can significantly improve sound source localization accuracy. One CI manufacturer has already implemented the possibility of mismatch compensation in its current fitting software. This study investigated if this fitting parameter can be readily used in clinical settings and determined the effects of familiarization to a compensated device delay mismatch over a period of 3–4 weeks. Sound localization accuracy and speech understanding in noise were measured in eleven bimodal CI/HA users, with and without a compensation of the device delay mismatch. The results showed that sound localization bias improved to 0°, implying that the localization bias towards the CI was eliminated when the device delay mismatch was compensated. The RMS error was improved by 18% with this improvement not reaching statistical significance. The effects were acute and did not further improve after 3 weeks of familiarization. For the speech tests, spatial release from masking did not improve with a compensated mismatch. The results show that this fitting parameter can be readily used by clinicians to improve sound localization ability in bimodal users. Further, our findings suggest that subjects with poor sound localization ability benefit the most from the device delay mismatch compensation.
Users of a cochlear implant (CI) in one ear, who are provided with a hearing aid (HA) in the contralateral ear, so-called bimodal listeners, are typically affected by a constant and relatively large interaural time delay offset due to differences in signal processing and differences in stimulation. For HA stimulation, the cochlear travelling wave delay is added to the processing delay, while for CI stimulation, the auditory nerve fibers are stimulated directly. In case of MED-EL CI systems in combination with different HA types, the CI stimulation precedes the acoustic HA stimulation by 3 to 10 ms. A self-designed, battery-powered, portable, and programmable delay line was applied to the CI to reduce the device delay mismatch in nine bimodal listeners. We used an A-B-B-A test design and determined if sound source localization improves when the device delay mismatch is reduced by delaying the CI stimulation by the HA processing delay (τ HA ). Results revealed that every subject in our group of nine bimodal listeners benefited from the approach. The root-mean-square error of sound localization improved significantly from 52.6° to 37.9°. The signed bias also improved significantly from 25.2° to 10.5°, with positive values indicating a bias toward the CI. Furthermore, two other delay values (τ HA –1 ms and τ HA +1 ms) were applied, and with the latter value, the signed bias was further reduced in some test subjects. We conclude that sound source localization accuracy in bimodal listeners improves instantaneously and sustainably when the device delay mismatch is reduced.
In asymmetric treatment of hearing loss, processing latencies of the modalities typically differ. This often alters the reference interaural time difference (ITD) (i.e., the ITD at 0° azimuth) by several milliseconds. Such changes in reference ITD have shown to influence sound source localization in bimodal listeners provided with a hearing aid (HA) in one and a cochlear implant (CI) in the contralateral ear. In this study, the effect of changes in reference ITD on speech understanding, especially spatial release from masking (SRM) in normal-hearing subjects was explored. Speech reception thresholds (SRT) were measured in ten normal-hearing subjects for reference ITDs of 0, 1.75, 3.5, 5.25 and 7 ms with spatially collocated (S0N0) and spatially separated (S0N90) sound sources. Further, the cues for separation of target and masker were manipulated to measure the effect of a reference ITD on unmasking by A) ITDs and interaural level differences (ILDs), B) ITDs only and C) ILDs only. A blind equalization-cancellation (EC) model was applied to simulate all measured conditions. SRM decreased significantly in conditions A) and B) when the reference ITD was increased: In condition A) from 8.8 dB SNR on average at 0 ms reference ITD to 4.6 dB at 7 ms, in condition B) from 5.5 dB to 1.1 dB. In condition C) no significant effect was found. These results were accurately predicted by the applied EC-model. The outcomes show that interaural processing latency differences should be considered in asymmetric treatment of hearing loss.
Bei bimodaler Cochlea-Implantat-/Hörgerät-Versorgung kann es aufgrund seitenverschiedener Signalverarbeitung zu einer zeitlich versetzten Stimulation der beiden Modalitäten kommen. Jüngste Studien haben gezeigt, dass durch zeitlichen Abgleich der Modalitäten die Schalllokalisation bei bimodaler Versorgung verbessert werden kann. Um solch einen Abgleich vornehmen zu können, ist die messtechnische Bestimmung der Durchlaufzeit von Hörgeräten erforderlich. Kommerziell verfügbare Hörgerätemessboxen können diese Werte häufig liefern. Die dazu verwendete Signalverarbeitung wird dabei aber oft nicht vollständig offengelegt. In dieser Arbeit wird ein alternativer und nachvollziehbarer Ansatz zum Design eines simplen Messaufbaus basierend auf einem Arduino DUE Mikrocontroller-Board vorgestellt. Hierzu wurde ein Messtisch im 3D-Druck gefertigt, auf welchem Hörgeräte über einen 2-ccm-Kuppler an ein Messmikrofon angeschlossen werden können. Über einen Latenzvergleich mit dem simultan erfassten Signal eines Referenzmikrofons kann die Durchlaufzeit von Hörgeräten bestimmt werden. Frequenzspezifische Durchlaufzeiten werden mittels einer Kreuzkorrelation zwischen Ziel- und Referenzsignal errechnet. Aufnahme, Ausgabe und Speicherung der Signale erfolgt über einen ATMEL SAM3X8E Mikrocontroller, welcher auf dem Arduino DUE-Board verbaut ist. Über eigens entworfene elektronische Schaltungen werden die Mikrofone und der verwendete Lautsprecher angesteuert. Nach Abschluss einer Messung (Messdauer ca. 5 s) werden die Messdaten seriell an einen PC übertragen, auf dem die Datenauswertung mittels MATLAB erfolgt. Erste Validierungen zeigten eine hohe Stabilität der Messergebnisse mit sehr geringen Standardabweichungen im Bereich weniger Mikrosekunden für Pegel zwischen 50 und 75 dB (A). Der Messaufbau wird in laufenden Studien zur Quantifizierung der Durchlaufzeit von Hörgeräten verwendet.
Introduction: Subjects with mild to moderate hearing loss today often receive hearing aids (HA) with open-fitting (OF). In OF, direct sound reaches the eardrums with minimal damping. Due to the required processing delay in digital HA, the amplified HA sound follows some milliseconds later. This process occurs in both ears symmetrically in bilateral HA provision and is likely to have no or minor detrimental effect on binaural hearing. However, the delayed and amplified sound are only present in one ear in cases of unilateral hearing loss provided with one HA. This processing alters interaural timing differences in the resulting ear signals.
Methods: In the present study, an experiment with normal-hearing subjects to investigate speech intelligibility in noise with direct and delayed sound was performed to mimic unilateral and bilateral HA provision with OF.
Results: The outcomes reveal that these delays affect speech reception thresholds (SRT) in the unilateral OF simulation when presenting speech and noise from different spatial directions. A significant decrease in the median SRT from –18.1 to –14.7 dB SNR is observed when typical HA processing delays are applied. On the other hand, SRT was independent of the delay between direct and delayed sound in the bilateral OF simulation.
Discussion: The significant effect emphasizes the development of rapid processing algorithms for unilateral HA provision.
Im Rahmen der Cochleaimplantat (CI)-Versorgung werden sowohl intraoperativ als auch postoperativ verschiedene elektrische und elektrophysiologische Diagnostikverfahren eingesetzt, bei denen elektrische Messgrößen vom CI erfasst und elektrophysiologische Messungen bei CI-Patienten durchgeführt werden. Zu den elektrophysiologischen Diagnostikverfahren zählen die Messung der elektrisch evozierten Summenaktionspotenziale des Hörnervs, die Registrierung der elektrisch evozierten auditorischen Hirnstammpotenziale und die Erfassung der elektrisch evozierten auditorischen kortikalen Potenziale. Diese Potenziale widerspiegeln die Erregung des Hörnervs und die Reizverarbeitung in verschiedenen Stationen der aufsteigenden Hörbahn bei intracochleärer elektrischer Stimulation mittels eines CI. Bei den aktuellen CI sind die Beurteilung der Elektrodenlage sowie die Prüfung der Ankopplung des Implantats an den Hörnerv wichtige Anwendungsgebiete der elektrophysiologischen Diagnostikverfahren. Ein weiteres bedeutendes Einsatzfeld stellt die Prüfung der Reizverarbeitung in der Hörbahn dar. Das Hauptanwendungsgebiet dieser Verfahren bildet jedoch die Unterstützung der Anpassung der CI-Sprachprozessoren bei Säuglingen und Kleinkindern auf der Basis elektrophysiologischer Schwellen.
The interaural time difference (ITD) is an important cue for the localization of sounds. ITD changes as little as 10 μs can be detected by the human auditory system. By provision of one ear with a cochlear implant (CI) ITD are altered due to the partial replacement of the peripheral auditory system. A hearing aid (HA), in contrast, does not replace but adds a processing delay component to the peripheral auditory system extending ITD. The aim of the present study was to quantify interaural stimulation timing between these different modalities to estimate the need for central auditory temporal compensation in single sided deaf CI users or bimodal CI/HA users. For this purpose, wave V latencies of auditory brainstem responses evoked either acoustically (ABR) or electrically via the CI (EABR) have been measured. The sum of delays consisting of CI signal processing measured in the MED-EL OPUS2 audio processor and EABR wave V latencies evoked on different intracochlear sites allowed an estimation of the entire CI channel-specific delay for MED-EL MAESTRO CI systems. We compared these values with ABR wave V latencies measured in the contralateral normal hearing or HA provided ear in different frequency bands. The results showed that EABR wave V latencies were consistently shorter than those evoked acoustically in the unaided normal hearing ear. Thus, artificial delays within the audio processor can be implemented to adjust interaural stimulation timing. The currently implemented group delays in the MED-EL CI system turned out to be reasonably similar to those of the unaided ear. For adjustment of CI and contralateral HA, in contrast, an adjustable additional across-frequency delay in the range of 1–11 ms implemented in the CI would be required. Especially for bimodal CI/HA users the adjustment of interaural stimulation timing may induce improved binaural hearing, reduced need for central auditory temporal compensation and increased acceptance of the CI/HA provision.
Objectives: Speech recognition on the telephone poses a challenge for patients with cochlear implants (CIs) due to a reduced bandwidth of transmission. This trial evaluates a home-based auditory training with telephone-specific filtered speech material to improve sentence recognition. Design: Randomised controlled parallel double-blind. Setting: One tertiary referral centre. Participants: A total of 20 postlingually deafened patients with CIs. Main outcome measures: Primary outcome measure was sentence recognition assessed by a modified version of the Oldenburg Sentence Test filtered to the telephone bandwidth of 0.3-3.4 kHz. Additionally, pure tone thresholds, recognition of monosyllables and subjective hearing benefit were acquired at two separate visits before and after a home-based training period of 10-14 weeks. For training, patients received a CD with speech material, either unmodified for the unfiltered training group or filtered to the telephone bandwidth in the filtered group. Results: Patients in the unfiltered training group achieved an average sentence recognition score of 70.0%±13.6% (mean±SD) before and 73.6%±16.5% after training. Patients in the filtered training group achieved 70.7%±13.8% and 78.9%±7.0%, a statistically significant difference (P=.034, t10 =2.292; two-way RM ANOVA/Bonferroni). An increase in the recognition of monosyllabic words was noted in both groups. The subjective benefit was positive for filtered and negative for unfiltered training. Conclusions: Auditory training with specifically filtered speech material provided an improvement in sentence recognition on the telephone compared to training with unfiltered material.
In users of a cochlear implant (CI) together with a contralateral hearing aid (HA), so-called bimodal listeners, differences in processing latencies between digital HA and CI up to 9 ms constantly superimpose interaural time differences. In the present study, the effect of this device delay mismatch on sound localization accuracy was investigated. For this purpose, localization accuracy in the frontal horizontal plane was measured with the original and minimized device delay mismatch. The reduction was achieved by delaying the CI stimulation according to the delay of the individually worn HA. For this, a portable, programmable, battery-powered delay line based on a ring buffer running on a microcontroller was designed and assembled. After an acclimatization period to the delayed CI stimulation of 1 hr, the nine bimodal study participants showed a highly significant improvement in localization accuracy of 11.6% compared with the everyday situation without the delay line (p < .01). Concluding, delaying CI stimulation to minimize the device delay mismatch seems to be a promising method to increase sound localization accuracy in bimodal listeners.
The ability to detect a target signal masked by noise is improved in normal-hearing listeners when interaural phase differences (IPDs) between the ear signals exist either in the masker or in the signal. To improve binaural hearing in bilaterally implanted cochlear implant (BiCI) users, a coding strategy providing the best possible access to IPD is highly desirable. In this study, we compared two coding strategies in BiCI users provided with CI systems from MED-EL (Innsbruck, Austria). The CI systems were bilaterally programmed either with the fine structure processing strategy FS4 or with the constant rate strategy high definition continuous interleaved sampling (HDCIS). Familiarization periods between 6 and 12 weeks were considered. The effect of IPD was measured in two types of experiments: (a) IPD detection thresholds with tonal signals addressing mainly one apical interaural electrode pair and (b) with speech in noise in terms of binaural speech intelligibility level differences (BILD) addressing multiple electrodes bilaterally. The results in (a) showed improved IPD detection thresholds with FS4 compared with HDCIS in four out of the seven BiCI users. In contrast, 12 BiCI users in (b) showed similar BILD with FS4 (0.6 ± 1.9 dB) and HDCIS (0.5 ± 2.0 dB). However, no correlation between results in (a) and (b) both obtained with FS4 was found. In conclusion, the degree of IPD sensitivity determined on an apical interaural electrode pair was not an indicator for BILD based on bilateral multielectrode stimulation.