The high frequency (HF) catheter ablation is the gold standard for the therapy of many cardiac tachyarrhythmias, such as atrioventricular node re-entry tachycardia (AVNRT), atrioventricular re-entry tachycardia (AVRT) or atrial flutter (AFL). The aim of the study was to simulate the HF ablation of AVNRT, AVRT, AFL and its heat propagation in reference to the supplied power with different electrode material and electrode size. The modeling and simulation were performed with the thermal and electromagnetic simulation software CST® (Computer Simulation Technology, Darmstadt). The modeling and simulation were carried out using ablation catheters with 4 mm tip electrode and 8 mm tip electrode with different electrode materials. Both electrode types were made of platinum and gold respectively. For the measurement of the heat propagation in the heart tissue, the catheters were integrated in the Offenburg heart rhythm model. The HF ablation procedures were performed with the 4 mm platinum tip electrode, with an application duration of 45 seconds and a power output of 40 watts. The HF ablation of the atrioventricular node slow pathway produced a maximum temperature of 66.33 °C. The Kent bundle HF ablation in the left atrium achieved a maximum temperature of 67.14 °C. The HF ablation of the right atrial isthmus resulted 65.96 °C. The 8 mm distal platinum tip electrode and a power output of 60 watts reached 72.85 °C. The 8 mm distal gold tip electrode and a power output of 60 watt reached 64.66 °C, due to the improved thermal conductivity of gold. Virtual heart and ablation electrode models allow the static and dynamic simulation of HF ablation with different electrode material and electrode size. The 3D simulation of the temperature profile may be used to optimize the AVNRT, AVRT and AFL HF ablation.
Abstract: Electric field of biventricular (BV) pacing, left ventricular (LV) electrode position and electrical interventricular desynchronization are important parameters for successful cardiac resynchronization therapy (CRT) in patients with heart failure, sinus rhythm and reduced LV ejection fraction. The aim of the study was to evaluate electric pacing field of transesophageal left atrial (LA) pacing and BV pacing with 3D heart rhythm simulation. Bipolar right atrial (RA), right ventricular (RV), LV electrodes and multipolar hemispherical esophageal LA electrodes were modeled with CST (Computer Simulation Technology, Darmstadt). Electric pacing field were simulated with bipolar RA and RV pacing with Solid S (Biotronik) electrode, bipolar LV pacing with Attain 4194 (Medtronic) electrode and bipolar LA pacing with TO8 (Osypka) esophageal electrode. 3D heart rhythm model with esophagus allowed electric pacing field simulation of 4-chamber pacing with bipolar intracardiac RA, RV, LV pacing and bipolar transesophageal LA pacing. The pacing amplitudes were 3V RA pacing amplitude, 50V LA pacing amplitude, 1.5V RV pacing amplitude and 3V LV pacing amplitude with 0.5ms pacing pulse duration. The atrioventricular delay between RA pacing and BV pacing was 140ms atrioventricular pacing delay and simultaneous RV and LV pacing. Electric pacing fields were simulated during the different pacing modes AAI, VVI, DDD and DDD0V. The intracardiac far-field pacing potentials were evaluated with intracardiac electrodes and a distance of 1mm from the electrodes with RA electrode 1.104V, RV electrode 0.703V and LV electrode 1.32V. The transesophageal far-field pacing potential was evaluated with transesophageal electrode and a distance of 10mm from the elelctrode with LA electrode 6.076V. Heart rhythm model simulation with esophagus allows evaluation of electric pacing fields in AAI, VVI, DDD, DDD0V and DDD0D pacing modes. Electric pacing field of RA, RV and LV pacing in combination with LA pacing may additional useful pacing mode in CRT non-responders.
Cardiac resynchronization therapy with atrioventricular and interventricular delay optimized biventricular pacing is an established therapy for symptomatic heart failure patients with prolongation of QRS duration, left bundle branch block and reduced left ventricular ejection fraction. The aim of the investigation was to evaluate right atrial, right ventricular and left ventricular electrical signals of implantable electronic cardiac devices with and without signal averaging technique with novel LabVIEW software. Electrical interatrial conduction delay and inter-ventricular conduction delay may be useful parameters to evaluate electrical atrial and ventricular desynchronization in heart failure patients.
Background: The electrical field (E-field) of the biventricular (BV) stimulation is important for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex. The 3D modeling allows the simulation of CRT and high frequency (HF) ablation.
Purpose: The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of atrial and BV stimulation and high frequency (HF) ablation in atrial fibrillation (AF).
Methods: The modeling and simulation was carried out using the electromagnetic simulation software CST (CST Darmstadt). Five multipolar left ventricular (LV) electrodes, one epicardial LV electrode, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modeled. Selected electrodes were integrated into the Offenburg heart rhythm model for the electrical field simulation. The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
Results: The right atrial stimulation was performed with an amplitude of 1.5 V with a pulse width of 0.5. The far-field potentials generated by the atrial stimulation were perceived by the right and left ventricular electrode. The far-field potential at a distance of 1 mm from the right ventricular electrode tip was 36.1 mV. The far-field potential at a distance of 1 mm from the left ventricular electrode tip was measured with 37.1 mV. The RV and LV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode with a pulse width of 0.5 ms each. The far-field potentials generated by the BV stimulations could be perceived by the RA electrode. The far-field potential at the RA electrode tip was 32.86 mV. AV node ablation was simulated with an applied power of 5 W at 420 kHz and 10 W at 500 kHz at the distal 8 mm ablation electrode.
Conclusions: Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF. The 3D simulation of the electrical field and temperature profile may be used to optimize the CRT and AF ablation.
Cardiac resynchronization therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients with inter- and intraventricular conduction delay. The aim of this pilot study was to test the feasibility of both transesophageal measurement of left ventricular (LV) electrical delay and transesophageal LV pacing prior to implantation, to better select patients for CRT.
Cardiac resynchronization therapy is an established therapy for heart failure patients. The aim of the study was to evaluate electrical left cardiac atrioventricular delay and interventricular desynchronization in sinus rhythm cardiac resynchronization therapy responder and non-responder. Cardiac electrical desynchronization were measured by surface ECG and focused transesophageal bipolar left atrial and left ventricular ECG before implantation of cardiac resynchronization therapy defibrillators. Preoperative electrical cardiac desynchronization was 195.7 ± 46.7 ms left cardiac atrioventricular delay and 74.8 ± 24.5 ms interventricular delay in cardiac resynchronization therapy responder. Cardiac resynchronization therapy responder New York Heart Association class improved during long term biventricular pacing. Transesophageal left cardiac atrioventricular delay and interventricular delay may be additional useful parameters to improve patient selection for cardiac resynchronization therapy.
Pulmonary vein isolation (PVI) is a common therapy in atrial fibrillation (AF). The cryoballoon was invented to isolate the pulmonary vein in one step and in a shorter time than a point-by-point radiofrequency (RF) ablation. The aim of the study was to model two cryoballoon catheters, one RF catheter and to integrate them into a heart rhythm model for the static and dynamic simulation of PVI by cryoablation and RF ablation in AF. The modeling and simulation were carried out using the electromagnetic and thermal simulation software CST (CST, Darmstadt). Two cryoballons and one RF ablation catheter were modeled based on the technical manuals of the manufacturers Medtronic and Osypka. The PVI especially the isolation of the left inferior pulmonary vein using a cryoballoon catheter was performed with a -50 °C heatsource and an exponential signal. The temperature at the balloon surface was -50 °C after 20 s ablation time, -24 °C from the balloon 0,5 mm in the myocardium, at a distance of 1 mm -3 °C, at 2 mm 18 °C and at a distance of 3mm 29 °C. PVI with RF energy was simulated with an applied power of 5 W at 420 kHz at the distal 8 mm ablation electrode. The temperature at the tip electrode was 110 °C after 15 s ablation time, 75 °C from the balloon at 0,5 mm in the myocardium, at a distance of 1 mm 58 °C, at 2 mm 45 °C and at a distance of 3 mm 38 °C. Virtual heart rhythm and catheter models as well as the simulation of the temperature allow the simulation of PVI in AF by cryo ablation and RF ablation. The 3D simulation of the temperature profile may be used to optimize RF and cryo ablation.
A disturbed synchronization of the ventricular contraction can cause a highly developed systolic heart failure in affected patients with reduction of the left ventricular ejection fraction, which can often be explained by a diseased left bundle branch block (LBBB). If medication remains unresponsive, the concerned patients will be treated with a cardiac resynchronization therapy (CRT) system. The aim of this study was to integrate His-bundle pacing into the Offenburg heart rhythm model in order to visualize the electrical pacing field generated by His-Bundle-Pacing. Modelling and electrical field simulation activities were performed with the software CST (Computer Simulation Technology) from Dessault Systèms. CRT with biventricular pacing is to be achieved by an apical right ventricular electrode and an additional left ventricular electrode, which is floated into the coronary vein sinus. The non-responder rate of the CRT therapy is about one third of the CRT patients. His- Bundle-Pacing represents a physiological alternative to conventional cardiac pacing and cardiac resynchronization. An electrode implanted in the His-bundle emits a stronger electrical pacing field than the electrical pacing field of conventional cardiac pacemakers. The pacing of the Hisbundle was performed by the Medtronic Select Secure 3830 electrode with pacing voltage amplitudes of 3 V, 2 V and 1,5 V in combination with a pacing pulse duration of 1 ms. Compared to conventional pacemaker pacing, His-bundle pacing is capable of bridging LBBB conduction disorders in the left ventricle. The His-bundle pacing electrical field is able to spread via the physiological pathway in the right and left ventricles for CRT with a narrow QRS-complex in the surface ECG.
Patients with focal ventricular tachycardia are at risk of hemodynamic failure and if no treatment is provided the mortality rate can exceed 30%. Therefore, medical professionals must be adequately trained in the management of these conditions. To achieve the best treatment, the origin of the abnormality should be known, as well as the course of the disease. This study provides an opportunity to visualize various focal ventricular tachycardias using the Offenburg heart rhythm model. Modeling and simulation of focal ventricular tachycardias in the Offenburg heart rhythm model was performed using CST (Computer Simulation Technology) software from Dessault Systèms. A bundle of nerve tissue in different regions in the left and right ventricle was defined as the focus in the already existing heart rhythm model. This ultimately served as the origin of the focal excitation sites. For the simulations, the heart rhythm model was divided into a mesh consisting of 5354516 tetrahedra, which is required to calculate the electric field lines. The simulations in the Offenburg heart rhythm model were able to successfully represent the progression of focal ventricular tachycardia in the heart using measured electrical field lines. The simulation results were realized as an animated sequence of images running in real time at a frame rate of 20 frames per second. By changing the frame rate, these simulations can additionally be produced at different speeds. The Offenburg heart rhythm model allows visualization of focal ventricular arrhythmias using computer simulations.