Cardiac resynchronization therapy (CRT) with biventricular pacing is an established therapy for heart failure (HF) patients (P) with ventricular desynchronization and reduced left ventricular (LV) ejection fraction. The aim of this study was to evaluate electrical right atrial (RA), left atrial (LA), right ventricular (RV) and LV conduction delay with novel telemetric signal averaging electrocardiography (SAECG) in implantable cardioverter defibrillator (ICD) P to better select P for CRT and to improve hemodynamics in cardiac pacing.
Methods: ICD-P (n=8, age 70.8 ± 9.0 years; 2 females, 6 males) with VVI-ICD (n=4), DDD-ICD (n=3) and CRT-ICD (n=1) (Medtronic, Inc., Minneapolis, MN, USA) were analysed with telemetric ECG recording by Medronic programmer 2090, ECG cable 2090AB, PCSU1000 oscilloscope with Pc-Lab2000 software (Velleman®) and novel National Intruments LabView SAECG software.
Results: Electrical RA conduction delay (RACD) was measured between onset and offset of RA deflection in the RAECG. Interatrial conduction delay (IACD) was measured between onset of RA deflection and onset of far-field LA deflection in the RAECG. Interventricular conduction delay (IVCD) was measured between onset of RV deflection in the RVECG and onset of LV deflection in the LVECG. Telemetric SAECG recording was possible in all ICD-P with a mean of 11.7 ± 4.4 SAECG heart beats, 97.6 ± 33.7 ms QRS duration, 81.5 ± 44.6 ms RACD, 62.8 ± 28.4 ms RV conduction delay, 143.7 ± 71.4 ms right cardiac AV delay, 41.5 ms LA conduction delay, 101.6 ms LV conduction delay, 176.8 ms left cardiac AV delay, 53.6 ms IACD and 93 ms IVCD.
Conclusions: Determination of RA, LA, RV and LV conduction delay, IACD, IVCD, right and left cardiac AV delay by telemetric SAECG recording using LabView SAECG technique may be useful parameters of atrial and ventricular desynchronization to improve P selection for CRT and hemodynamics in cardiac pacing.
Currently, QRS width and bundle branch block morphology are used as electrocardiographic guideline criterias to selectheart failure (HF) patients with interventricular desynchronization in sinus rhythm (SR) for cardiac resynchronisationtherapy (CRT). Nevertheless, up to 30% of these patients do not benefit from implantation of CRT systems. Esophagealleft ventricular electrogram (LVE) enables semi-invasive measurement of interventricular conduction delays (IVCD)even in patients with atrial fibrillation (AF). To routinely apply this method, a programmer based semi-invasiveautomatic quantification of IVCD should to be developed. Our aims were todefine interventricular conduction delaysby analyzing fractionated left ventricular (LV) deflections in the esophageal left ventricular electrogram of HF patientsin SR or AF.
In 66 HF patients (49 male,17 female, age 65 ± 10 years) a 5F TOslim electrode (Osypka AG, Germany) was perorallyapplied. Using BARD EP Lab, cardiac desynchronization was quantified as interval IVCD between onset of QRS insurface ECG and the investigator-determined onset of the left ventricular deflection in LVE. IVCD was compared withthe intervals between QRS onset and the first maximum (IVCDm1) and between QRS onset and the second maximum(IVCDm2) of the LV complex.
QRS of 173 ± 26 ms was linked with empirical IVCD of 75 ± 25 ms, at mean. First and second LV maximum could beascertained beyond doubt in all patients. Significant correlations of the p<0,01 level were found between IVCD and theIVCDm1 of 96 ± 28 ms as well as between IVCD and the IVCDm2 of 147 ± 31 ms, at mean. To standardize automatic measurement of interventricular conduction delays with respect to patients with fractionatedLV complexes, the first maximum of the LV deflection should be utilized to qualify the IVCD of HF patients with sinusrhythm and atrial fibrillation.
Cardiac resynchronization therapy with biventricular pacing is an established therapy for heart failure patients with electrical left ventricular desynchronization. The aim of this study was to evaluate left atrial conduction delay, intra left atrial conduction delay, left ventricular conduction delay and intra left ventricular conduction delay in heart failure patients using novel signal averaging transesophageal left heart ECG software.
Methods: 8 heart failure patients with dilated cardiomyopathy (DCM), age 68 ± 9 years, New York Heart Association (NYHA) class 2.9 ± 0.2, 24.8 ± 6.7 % left ventricular ejection fraction, 188.8 ± 15.5 ms QRS duration and 8 heart failure patients with ischaemic cardiomyopathy (ICM), age 67 ± 8 years, NYHA class 2.9 ± 0.3, 32.5 ± 7.4 % left ventricular ejection fraction and 167.6 ± 19.4 ms QRS duration were analysed with transesophageal and transthoracic ECG by Bard LabDuo EP system and novel National Intruments LabView signal averaging ECG software.
Results: The electrical left atrial conduction delay was 71.3 ± 17.6 ms in ICM versus 72.3 ± 12.4 ms in DCM, intra left atrial conduction delay 66.8 ± 8.6 ms in ICM versus 63.4 ± 10.9 ms in DCM and left cardiac AV delay 180.5 ± 32.6 ms in ICM versus 152.4 ± 30.4 ms in DCM. The electrical left ventricular conduction delay was 40.9 ± 7.5 ms in ICM versus 42.6 ± 17 ms in DCM and intra left ventricular conduction delay 105.6 ± 19.3 ms in ICM versus 128.3 ± 24.1 ms in DCM.
Conclusions: Left heart signal averaging ECG can be utilized to analyse left atrial conduction delay, intra left atrial conduction delay, left ventricular conduction delay and intra left ventricular conduction delay to improve patient selection for cardiac resynchronization therapy.
New frontiers of supraventricular tachycardia and atrial flutter evaluation and catheter ablation
(2012)
Radiofrequency catheter ablation (RFCA) has revolutionized treatment for tachyarrhythmias and has become first-line therapy for some tachycardias. Although developed in the 1980s and widely applied in the 1990s, the technique is still in development. Transesophageal atrial pacing (TAP) can used for initiation and termination of supraventricular tachycardia (SVT).
Methods: The paroxysmal SVT include a wide spectrum of disorders including, in descending order of frequency, atrial flutter, atrioventricular (AV) nodal reentry, Wolff-Parkinson-White syndrome, and atrial tachycardia. While not life-threatening in most cases, they may cause important symptoms, such as palpitations, chest discomfort, breathlessness, anxiety, and syncope, which significantly impair quality of life. Medical therapy has variable efficacy, and most patients are not rendered free of symptoms. Research over the past several decades has revealed fundamental mechanisms involved in the initiation and maintenance of all of these arrhythmias. Knowledge of mechanisms has in turn led to highly effective surgical and catheter-based treatments. The supraventricular arrhythmias and their treatment are described in this report. SVT initiation was analysed with programmed TAP in 49 patients with palpitations (age 47 ± 17 years, 24 females, 25 males).
Results: In comparison to antiarrhythmic drug therapy the radiofrequency catheter ablation in patients suffering from atrial flutter, atrioventricular nodal reentry, atrioventricular reentry and atrial tachycardia is the better choice in most cases. TAP SVT initiation was possible in 23 patients before RFCA. Atrial cycle length of SVT was 320 ± 59 ms. We initiated AV nodal reentrant tachycardia (AVNRT, n=15), atrial tachycardia (AT, n=6) and AV reentrant tachycardia with Kent pathway conduction (AVRT, n=2) before RFCA.
Conclusions: Radiofrequency catheter ablation is a successful and safe method to cure most patients with paroxysmal supraventricular tachycardias. TAP allowed initiation and termination of SVT especially in outpatients.
ECG simulators, available on the market, imitate the electric activity of the heart in a simplified manner. Thus, they are suitable for education purposes but not really for testing algorithms to recognize complex arrhythmias needed for pacemakers and implantable defibrillators. Especially certain discrimination between various morphologies of atrial and ventricular fibrillation needs simulators providing native electrograms of different patients’ heart rhythm events. This explains the necessity to develop an ECG simulator providing high-resolution native intracardiac and surface electrograms of in-vivo rhythm events. In this paper we demonstrate an approach for an ECG simulator based on a consumer multichannel soundcard and a corresponding software application for a laptop computer. This Live-ECG Simulator is able to handle invasive electrogram recordings from electrophysiological studies and send the data to a modified external soundcard for subsequent digital to analog conversion. The hardware is completed with an electronic circuit providing level adjustment to adapt the output amplitude to the input conditions of several cardiac implants.
About 20% of those heart failure patients receiving cardiac resynchronization therapy (CRT) are in atrial fibrillation (AF). Current guidelines apply for patients in sinus rhythm only. Recent studies have shown again, that successful resynchronization is closely linked to a pre-existent ventricular desynchronization. In those studies, the interventricular conduction delay (IVCD) was determined prior to device implantation by ultrasound in patients with sinus rhythm (SR)only. In patients with AF this method ́s use is limited.
To implement left-heart electrogram (LHE) into standard programmers and to simplify IVCD measurement in heart failure patients with AF, LHE was recorded in 11 AF patients with heart failure by Biotronik ICS3000 programmer via a15Hz Butterworth high-pass filter. Therefore, TOslim esophageal electrode (Dr. Osypka GmbH, Rheinfelden, Germany) was perorally applied and fixed in position of maximal left ventricular defection. IVCD was measured between onset of QRS in surface ECG and left ventricular defection (LV) in LHE. In addition, intra-left ventricular conduction delay (ILVCD) was measured as duration of LV in LHE.
In all of the 11 AF patients, desynchronization was quantifiable by LHE. Mean QRS of 162 ± 27ms (120-206ms) was linked with IVCD of 62ms ± 27ms (37-98ms) and ILVCD of 110 ± 20ms (80-144ms), at mean. Correlation between IVCD and QRS was 0.39 (n. s.) with IVCD/QRS ratio of 0.38 ± 0.11 (0.22-0.81).
A 15Hz high-pass filtered LHE feature of the Biotronik ICS3000 programmer is feasible to quantify ventricular dyssynchrony in heart failure patients with AF in order to clearly indicate implantation of CRT systems. As relations between QRS duration, IVCD and ILVCD considerably differ interindividually, the predictive values of IVCD, ILVCD and IVCD/QRS ratio for individual CRT response or non-response shall be identified in follow-up studies.
Non-fluoroscopic Imaging with MRT/CT Image Integration - Catheter Positioning with Double Precision
(2014)
Introduction: When antiarrhythmic drug therapy has failed, different approaches of pulmonary vein isolation are considered a reasonable option in the treatment of atrial fibrillation. It will be performed predominantly by radiofrequency catheter ablation. As the individual anatomy of left atrium and the pulmonary veins differs considerably, accurate visualization of these structures is essential during catheter positioning. Using non-fluoroscopic electroanatomic mapping system with image integration, electroanatomic mapping can be combined with highly detailed anatomical MRT or CT information on complex left atrial structures. This may facilitate catheter navigation during ablation for atrial fibrillation.
Methods: The CARTO XP electroanatomic system was used in a project during biomedical engineering study to practice image integration of anonymized real patients that underwent pulmonary vein isolation by CARTO XP and a MRT/CT procedure. Using the image integration software, MRT or CT images were imported into the CARTO XP system. The next process was segmentation of the acquired images. It involves dividing the images into different regions in order to select the structures of interest. In clinical routine, this segmentation has to be performed before catheter ablation. Then, the segmented images were aligned with the reconstructed electroanatomic maps. This consists of several steps, including selection of the left atrium, scaling of the reconstructed geometry, fusion of the structures using landmarks, and optimization of the integration by adjusting the reconstructed geometry of the left atrium.
Results: In the 3 months lasting period of the project, image integration was trained in 13 patients undergoing catheter ablation for atrial fibrillation. Within this period, time consumption for the process decreased from about 90 minutes at the beginning to about 35 minutes at the end for one patient.
Conclusion: Image integration into non-fluoroscopic electroanatomic map is a sophisticated tool in cardiac radiofrequency catheter ablation. Intensive training is necessary to control the procedure.