Background: The electrical field (E-field) of the biventricular (BV) stimulation is important for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex. The 3D modeling allows the simulation of CRT and high frequency (HF) ablation.
Purpose: The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of atrial and BV stimulation and high frequency (HF) ablation in atrial fibrillation (AF).
Methods: The modeling and simulation was carried out using the electromagnetic simulation software CST (CST Darmstadt). Five multipolar left ventricular (LV) electrodes, one epicardial LV electrode, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modeled. Selected electrodes were integrated into the Offenburg heart rhythm model for the electrical field simulation. The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
Results: The right atrial stimulation was performed with an amplitude of 1.5 V with a pulse width of 0.5. The far-field potentials generated by the atrial stimulation were perceived by the right and left ventricular electrode. The far-field potential at a distance of 1 mm from the right ventricular electrode tip was 36.1 mV. The far-field potential at a distance of 1 mm from the left ventricular electrode tip was measured with 37.1 mV. The RV and LV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode with a pulse width of 0.5 ms each. The far-field potentials generated by the BV stimulations could be perceived by the RA electrode. The far-field potential at the RA electrode tip was 32.86 mV. AV node ablation was simulated with an applied power of 5 W at 420 kHz and 10 W at 500 kHz at the distal 8 mm ablation electrode.
Conclusions: Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF. The 3D simulation of the electrical field and temperature profile may be used to optimize the CRT and AF ablation.
The electrical field (E-field) of the biventricular (BV) stimulation is important for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex.
The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of BV stimulation and HF ablation in atrial fibrillation (AF).
The modeling and simulation was carried out using the electromagnetic simulation software CST. Five multipolar left ventricular (LV) electrodes, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modelled. A selection were integrated into the heart rhythm model (Schalk, Offenburg) for the electrical field simulation. The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
The BV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode with a pulse width of 0.5 ms each. The far-field potential at the RA electrode tip was 32.86 mV and 185.97 mV at a distance of 1 mm from the RA electrode tip. AV node ablation was simulated with an applied power of 5 W at 420 kHz at the distal ablation electrode. The temperature at the catheter tip was 103.87 °C after 5 s ablation time and 37.61 °C at a distance of 2 mm inside the myocardium. After 15 s, the temperature was 118.42 °C and 42.13 °C.
Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF and could be used to optimize the CRT and AF ablation.
Die Entwicklung von neuartigen Elektrodentypen und die Weiterentwicklung bestehender Produkten machen einen großen Teil der entstehenden Kosten für ein Unternehmen aus. Mithilfe geeigneter Software können Änderungen der Konstruktionen erfasst und bestimmte Simulationen, bspw. das Auftreten von Wechselwirkungen im elektrischen Feld, vor der eigentlichen Prototypenerstellung durchgeführt werden. Das Ziel der Studie besteht in der Modellierung unterschiedlicher Schrittmacher- und Ablationselektroden und deren Integration in das Offenburger Herzrhythmusmodell (HRM) zur statischen und dynamischen Simulation der biventrikulären Stimulation und HF Ablation bei Vorhofflimmern (AF).
Background: The electrical field (E-field) of the biventricular (BV) stimulation is essential for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex. 3D modeling allows the simulation of CRT and high frequency (HF) ablation.
Purpose: The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of BV stimulation and HF ablation in atrial fibrillation (AF).
Methods: The modeling and simulation was carried out using the electromagnetic simulation software. Five multipolar left ventricular (LV) electrodes, one epicardial LV electrode, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modeled. Different models of electrodes were integrated into a heart rhythm model for the electrical field simulation (fig.1). The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
Results: The RV and LV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode, each with a pulse width of 0.5 ms. The far-field potentials generated by the BV stimulations were perceived by the RA electrode. The far-field potential at the RA electrode tip was 32.86 mV. A far-field potential of 185.97 mV resulted at a distance of 1 mm from the RA electrode tip. AV node ablation was simulated with an applied power of 5 W at 420 kHz at the distal 8 mm ablation electrode. The temperature at the catheter tip was 103.87 ° C after 5 s ablation time, 44.17 ° C from the catheter tip in the myocardium and 37.61 ° C at a distance of 2 mm. After 10 s, the temperature at the three measuring points described above was 107.33 ° C, 50.87 ° C, 40.05 ° C and after 15 seconds 118.42 ° C, 55.75 ° C and 42.13 ° C.
Conclusions: Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF. The 3D simulation of the electrical field and temperature profile may be used to optimize the CRT and AF ablation.