Currently, QRS width and bundle branch block morphology are used as electrocardiographic guideline criterias to selectheart failure (HF) patients with interventricular desynchronization in sinus rhythm (SR) for cardiac resynchronisationtherapy (CRT). Nevertheless, up to 30% of these patients do not benefit from implantation of CRT systems. Esophagealleft ventricular electrogram (LVE) enables semi-invasive measurement of interventricular conduction delays (IVCD)even in patients with atrial fibrillation (AF). To routinely apply this method, a programmer based semi-invasiveautomatic quantification of IVCD should to be developed. Our aims were todefine interventricular conduction delaysby analyzing fractionated left ventricular (LV) deflections in the esophageal left ventricular electrogram of HF patientsin SR or AF.
In 66 HF patients (49 male,17 female, age 65 ± 10 years) a 5F TOslim electrode (Osypka AG, Germany) was perorallyapplied. Using BARD EP Lab, cardiac desynchronization was quantified as interval IVCD between onset of QRS insurface ECG and the investigator-determined onset of the left ventricular deflection in LVE. IVCD was compared withthe intervals between QRS onset and the first maximum (IVCDm1) and between QRS onset and the second maximum(IVCDm2) of the LV complex.
QRS of 173 ± 26 ms was linked with empirical IVCD of 75 ± 25 ms, at mean. First and second LV maximum could beascertained beyond doubt in all patients. Significant correlations of the p<0,01 level were found between IVCD and theIVCDm1 of 96 ± 28 ms as well as between IVCD and the IVCDm2 of 147 ± 31 ms, at mean. To standardize automatic measurement of interventricular conduction delays with respect to patients with fractionatedLV complexes, the first maximum of the LV deflection should be utilized to qualify the IVCD of HF patients with sinusrhythm and atrial fibrillation.
Using guideline parameters for indication of cardiac resynchronization therapy (CRT), only about two thirds of the patients improve clinically. Unfortunately both, surface ECG and echo are uncertain to predict CRT response. To better characterize cardiac desynchronization in heart failure, interventricular (IVCD) and intra-leftventricular conduction delays (ILVCD) were measured by esophageal left ventricular electrogram (LVE). Recordings in 43 CRT patients (34m, 9f, age: 64.7 ± 9.5yrs) evidenced only weak correlation between IVCD and QRS of 0.53 and between ILVCD and QRS of 0.33. This demonstrated that QRS duration is not a reliable indicator of desynchronization. Therefore, the study resulted into development of LVE feature for a programmer with implant support device. It can be used interoperatively to guide the left ventricular electrode location in order to increase responder rate in CRT.
Semi-invasive electromechanical target interval to guide left ventricular electrode placement
(2011)
AV delay (AVD) optimization can improve hemodynamics and avoid nonresponding to cardiac resynchronization therapy (CRT). AVD can be approximated by the sum of the individual implant-related interatrial conduction interval and a mean electromechanical interval of about 50ms. We searched for methods to facilitate automatic, implant-based AV delay optimization. In 25 patients (19m, 6f, age: 65±8yrs.) with Medtronic Insync III Marquis CRT-D series systems and left ventricular electrode at lateral or posterolateral wall, we determined interatrial conduction intervals by telemetric left ventricular tip versus superior vena cava coil electrogram (LVCE). Compared with esophageal measurements, the duration of optimal AV delay by LVCE showed good correlation (k=0.98, p=0.01) with a difference of 1.5±4.9ms, only. Therefore, LVCE is feasible to determine interatrial conduction intervals in order to automate AV delay optimization in CRT-D pacing promising increased accuracy compared to other algorithms.