Cardiac resynchronization therapy is an established therapy for heart failure patients with sinus rhythm, reduced left ventricular ejection fraction and prolongation of QRS duration. The aim of the study was to evaluate ventricular desynchronization with electrical interventricular delay (IVD) to left ventricular delay (LVD) ratio in atrial fibrillation heart failure patients. IVD and LVD were measured by transesophageal posterior left ventricular ECG recording. In atrial fibrillation heart failure patients with prolonged QRS duration, the mean IVD-to-LVD-ratio was 0.84 +/- 0.42 with a range from 0.17 to 2.2 IVD-to-LVD-ratio. IVD-to-LVD-ratio correlated with QRS duration. IVD-to-LVD-ratio may be a useful parameter to evaluate electrical ventricular desynchronization in atrial fibrillation heart failure patients.
Introduction: Cardiac resynchronization therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients (P) with ventricular desynchronisation, but not all patients improved clinically. Aim of this study was to evaluate electrical intra-left ventricular conduction delay (LVCD) and interventricular conduction delay (IVCD), to better select patients for CRT.
Methods: 65 HF patients (age 63.4 ± 10.6 years; 7 females, 58 males) with New York Heart Association (NYHA) class 3 ± 0.2, 24.4 ± 6.7 % left ventricular (LV) ejection fraction and 167.4 ± 35.6 ms QRSD were included. Esophageal TO Osypka focused hemispherical electrodes catheter was perorally applied in position of maximum LV deflection to measure LVCD between onset and offset of LV deflection and IVCD between earliest onset of QRS in the 12-channel surface ECG and onset of LV deflection in the focused bipolar transesophageal LV electrogram.
Results: There were 50 responders with LVCD of 76.5 ± 20.4 ms, IVCD of 80.5 ± 26.1 ms (P=0.34) and QRSD of 171 ± 37.7 ms. 15 non-responders had longer LVCD of 90 ± 28.5 ms (P = 0.045), shorter IVCD of 50.1 ± 29.1 ms (P < 0.001) and QRSD of 155.3 ± 25 ms (P=0.14). During 21.3 ± 20.3 month BV pacing follow-up, the responder`s NYHA classes improved from 3 ± 0.2 to 2. ± 0.3 (P < 0.001) whereas the non-responders NYHA classes did not improve from 3 ± 0.2 to 2.9 ± 0.3 (P = 0.43) during 15.7 ± 13.9 month BV pacing follow-up (53 Boston, 10 Medtronic and 2 St. Jude CRT devices).
Conclusion: Determination of electrical LVCD and IVCD by focused bipolar transesophageal LV electrogram recording may be an additional useful technique to improve patient selection for CRT.
Cardiac resynchronization therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients with inter- and intraventricular conduction delay. The aim of this pilot study was to test the feasibility of both transesophageal measurement of left ventricular (LV) electrical delay and transesophageal LV pacing prior to implantation, to better select patients for CRT.
Introduction: Cardiac resynchronization therapy (CRT) with biventricular pacing is an established therapy for heart failure (HF) patients with sinus rhythm and ventricular desynchronisation. The aim of this study was to evaluate interventricular conduction delay (IVCD) and interatrial conduction delay (IACD) before and after premature ventricular contractions (PVC) in HF patients.
Methods: 13 HF patients (age 68 ± 10 years; 2 females, 11 males) with New York Heart Association functional class 2,8 ± 0.5, left ventricular (LV) ejection fraction 28,6 ± 12,6 %, 154 ± 25 ms QRS duration and PVC were analysed with bipolar transesophageal LV and left atrial electrogram recording and National Instruments LabView 2009 software. The level of significance of the t-test is 0,005.
Results: QRS duration increases during PVC (188 ± 32 ms) in comparison to the beat before (154 ± 25 ms, P = ) and after PVC (152 ± 25 ms,). IVCD increases during PVC up to 65 ± 33 ms (51 ± 19 ms in the beat before PVC, P=0.18, 49 ± 19 ms after PVC, P = 0.12). Intra-LV delay of 90 ± 16 ms is not different in the beat before PVC, 90 ± 14 ms during PVC (P = 0.99) and 94 ± 16 ms in the beat after PVC (P = 0.38). IACD is not significantly PVC influenced (67 ± 12 ms before PVC and 65 ± 13 ms after PVC, P = 0.71). Intra-left atrial conduction delay is not significant longer during PVC (57 ± 28 ms) than in the beat before PVC (54 ± 13 ms, P = 0.51) or after PVC (54 ± 8 ms, P = 0.45). PQ duration increases significantly after PVC (224 ± 95 ms) in comparison to the beat before PVC (176± 29 ms, P =...).
Conclusion: Transesophageal left cardiac electrocardiography with LabView 2009 software can improve evaluation of IVCD and IACD before, during and after PVC in HF patient selection for CRT.
Introduction: Cardiac resynchronization therapy (CRT) with left ventricular (LV) pacing is an established therapy for heart failure (HF) patients (P) with ventricular desynchronisation and reduced LV ejection fraction (EF). The aim of this study was to test the utilization of the transesophageal approach to measure arterial pulse pressure (PP) during LV pacing and electrical interventricular conduction delay (IVCD), to better select patients for CRT.
Methods: 32 HF patients (age 64 ± 10 years; 5 females, 27 males) with New York Heart Association (NYHA) class 2.8 ± 0.6, 27 ± 11 % LV EF and 155 ± 35 ms QRS duration were analysed with semi-invasive left cardiac pacing and electrocardiography. Esophageal TO8 Osypka catheter of 10.5 F diameter was perorally applied to the esophagus and placed in the position of maximum left atrial (LA) deflection and maximum LV deflection to measure PP with VAT or D00 pacing modes.
Results: Temporary transesophageal LV pacing was possible with VAT mode (n=16) and D00 mode (n=16) in all patients. In 15 Δ-PP-responders, PP was higher during LV pacing on than LV pacing off (78.3 ± 26.6 versus 65.9 ± 23.7 mmHg, P < 0.001) and NYHA class improved from 3.1 ± 0.35 to 2.1 ± 0.35 (P < 0.001) during 29 ± 26 month biventricular (BV) pacing follow-up (6 Medtronic and 9 Boston BV pacing devices). In 17 Δ-PP-non-responders, PP was not higher during LV pacing on than LV pacing off (61.5 ± 23.9 versus 60.9 ± 23.5 mmHg, P = 0.066). IVCD was significant longer in Δ-PP-responders than in Δ-PP-non-responders (87 ± 33 ms versus 37± 29 ms, P < 0.001).
Conclusion: Semi-invasive transesophageale LA and LV pacing with D00 and VAT mode and LV electrogram recording may be useful techniques to predict CRT improvement.
Cardiac resynchronization therapy with atrioventricular and interventricular pacing delay optimized biventricular pacing is an established therapy for heart failure patients with sinus rhythm and reduced left ventricular ejection fraction. The aim of the study was to evaluate atrioventricular and interventricular pacing delay optimization in cardiac resynchroniza-tion therapy by transthoracic impedance cardiography in biventricular pacing with different left ventricular electrode po-sition. In biventricular pacing heart failure patients with lateral, posterolateral and anterolateral left ventricular electrode position, the mean optimal atrioventricular sening delay was 108.6 ± 20.3 ms and the mean optimal interventricular pac-ing delay -12.3 ± 25.9 ms. Transthoracic impedance cardiography may be a useful technique to optimize atrioventricular and interventricular pacing delay in biventricular pacing with different left ventricular electrode position.
Cardiac resynchronization therapy (CRT) with biventricular pacing is an established therapy for heart failure (HF) patients (P) with ventricular desynchronization and reduced left ventricular (LV) ejection fraction. The aim of this study was to evaluate electrical right atrial (RA), left atrial (LA), right ventricular (RV) and LV conduction delay with novel telemetric signal averaging electrocardiography (SAECG) in implantable cardioverter defibrillator (ICD) P to better select P for CRT and to improve hemodynamics in cardiac pacing.
Methods: ICD-P (n=8, age 70.8 ± 9.0 years; 2 females, 6 males) with VVI-ICD (n=4), DDD-ICD (n=3) and CRT-ICD (n=1) (Medtronic, Inc., Minneapolis, MN, USA) were analysed with telemetric ECG recording by Medronic programmer 2090, ECG cable 2090AB, PCSU1000 oscilloscope with Pc-Lab2000 software (Velleman®) and novel National Intruments LabView SAECG software.
Results: Electrical RA conduction delay (RACD) was measured between onset and offset of RA deflection in the RAECG. Interatrial conduction delay (IACD) was measured between onset of RA deflection and onset of far-field LA deflection in the RAECG. Interventricular conduction delay (IVCD) was measured between onset of RV deflection in the RVECG and onset of LV deflection in the LVECG. Telemetric SAECG recording was possible in all ICD-P with a mean of 11.7 ± 4.4 SAECG heart beats, 97.6 ± 33.7 ms QRS duration, 81.5 ± 44.6 ms RACD, 62.8 ± 28.4 ms RV conduction delay, 143.7 ± 71.4 ms right cardiac AV delay, 41.5 ms LA conduction delay, 101.6 ms LV conduction delay, 176.8 ms left cardiac AV delay, 53.6 ms IACD and 93 ms IVCD.
Conclusions: Determination of RA, LA, RV and LV conduction delay, IACD, IVCD, right and left cardiac AV delay by telemetric SAECG recording using LabView SAECG technique may be useful parameters of atrial and ventricular desynchronization to improve P selection for CRT and hemodynamics in cardiac pacing.
New frontiers of supraventricular tachycardia and atrial flutter evaluation and catheter ablation
(2012)
Radiofrequency catheter ablation (RFCA) has revolutionized treatment for tachyarrhythmias and has become first-line therapy for some tachycardias. Although developed in the 1980s and widely applied in the 1990s, the technique is still in development. Transesophageal atrial pacing (TAP) can used for initiation and termination of supraventricular tachycardia (SVT).
Methods: The paroxysmal SVT include a wide spectrum of disorders including, in descending order of frequency, atrial flutter, atrioventricular (AV) nodal reentry, Wolff-Parkinson-White syndrome, and atrial tachycardia. While not life-threatening in most cases, they may cause important symptoms, such as palpitations, chest discomfort, breathlessness, anxiety, and syncope, which significantly impair quality of life. Medical therapy has variable efficacy, and most patients are not rendered free of symptoms. Research over the past several decades has revealed fundamental mechanisms involved in the initiation and maintenance of all of these arrhythmias. Knowledge of mechanisms has in turn led to highly effective surgical and catheter-based treatments. The supraventricular arrhythmias and their treatment are described in this report. SVT initiation was analysed with programmed TAP in 49 patients with palpitations (age 47 ± 17 years, 24 females, 25 males).
Results: In comparison to antiarrhythmic drug therapy the radiofrequency catheter ablation in patients suffering from atrial flutter, atrioventricular nodal reentry, atrioventricular reentry and atrial tachycardia is the better choice in most cases. TAP SVT initiation was possible in 23 patients before RFCA. Atrial cycle length of SVT was 320 ± 59 ms. We initiated AV nodal reentrant tachycardia (AVNRT, n=15), atrial tachycardia (AT, n=6) and AV reentrant tachycardia with Kent pathway conduction (AVRT, n=2) before RFCA.
Conclusions: Radiofrequency catheter ablation is a successful and safe method to cure most patients with paroxysmal supraventricular tachycardias. TAP allowed initiation and termination of SVT especially in outpatients.
Cardiac resynchronization therapy is an established therapy for heart failure patients. The aim of the study was to evaluate electrical left cardiac atrioventricular delay and interventricular desynchronization in sinus rhythm cardiac resynchronization therapy responder and non-responder. Cardiac electrical desynchronization were measured by surface ECG and focused transesophageal bipolar left atrial and left ventricular ECG before implantation of cardiac resynchronization therapy defibrillators. Preoperative electrical cardiac desynchronization was 195.7 ± 46.7 ms left cardiac atrioventricular delay and 74.8 ± 24.5 ms interventricular delay in cardiac resynchronization therapy responder. Cardiac resynchronization therapy responder New York Heart Association class improved during long term biventricular pacing. Transesophageal left cardiac atrioventricular delay and interventricular delay may be additional useful parameters to improve patient selection for cardiac resynchronization therapy.
Cardiac resynchronization therapy (CRT) with biventricular (BV) pacing is an established therapy in approximately two-thirds of symptomatic heart failure (HF) patients (P) with left bundle branch block (LBBB). The aim of this study was to evaluate left atrial (LA) conduction delay (LACD) and left ventricular (LV) conduction delay (LVCD) using pre-implantational transesophageal electrocardiography (ECG) in sinus rhythm (SR) CRT responder (R) and non-responder (NR).
Methods: SR HF P (n=52, age 63.6±10.4 years; 6 females, 46 males) with New York Heart Association (NYHA) class 3.0±0.2, 24.4±7.1 % LV ejection fraction and 171.2±37.6 ms QRS duration (QRSD) were measured by bipolar filtered transesophageal LA and LV ECG recording with hemispherical electrodes (HE) TO catheter (Osypka AG, Rheinfelden, Germany). LACD was measured between onset of P-wave in the surface ECG and onset of LA deflection in the LA ECG. LVCD was measured between onset of QRS in the surface ECG and onset of LV deflection in the LV ECG.
Results: There were 78.8 % SR CRT R (n=41) with 171.2±36.9 ms QRSD, 73.3±25.7 ms LACD, 80.0±24.0 ms LVCD and 2.3±0.5 QRSD-LVCD-ratio. SR CRT R QRSD correlated with LACD (r=0.688, P<0.001) and LVCD (r=0.699, P<0.001). There were 21.2 % SR CRT NR (n=11) with 153.4±22.4 ms QRSD (P=0.133), 69.8±24.8 ms LACD (n=6, P=0.767), 54.2±31.0 ms LVCD (P<0.0046) and 3.9±2.5 QRSD-LVCD-ratio (P<0.001). SR CRT NR QRSD not corre-lated with IACD (r=-0.218, P=0.678) and IVCD (r=0.042, P=0.903). During a 22.8±21.3 month CRT follow-up, the CRT R NYHA class improved from 3.1±0.3 to 1.9±0.3 (P<0.001). In CRT NR, NYHA class not improved (2.9±0.4 to 2.9±0.2, P=1) during 11.2±9.8 months BV pacing.
Conclusions: Transesophageal LA and LV ECG with HE can be utilized to analyse LACD and LVCD in HF P. Pre-implantational LVCD and QRSD-LVCD-ratio may be additional useful parameters to improve P selection for SR CRT.
Background: Cardiac resynchronization therapy (CRT) with biventricular (BV) pacing is an established therapy for heart failure (HF) patients (P) with sinus rhythm, reduced left ventricular (LV) ejection fraction (EF) and electrical ventricular desynchronization. The aim of the study was to evaluate electrical interventricular delay (IVD) and left ventricular delay (LVD) in right ventricular (RV) pacemaker pacing before upgrading to CRT BV pacing.
Methods: HF P (n=11, age 69.0 ± 7.9 years, 1 female, 10 males) with DDD pacemaker (n=10), DDD defibrillator (n=1), RV pacing, New York Heart Association (NYHA) class 3.0 ± 0.2 and 24.5 ± 4.9 % LVEF were measured by surface ECG and transesophageal bipolar LV ECG before upgrading to CRT defibrillator (n=8) and CRT pacemaker (n=3). IVD was measured between onset of QRS in the surface ECG and onset of LV signal in the transesophageal ECG. LVD was measured between onset and offset of LV signal in the transesophageal ECG. CRT atrioventricular (AV) and BV pacing delay were optimized by impedance cardiography.
Results: Interventricular and intraventricular desynchronization in RV pacemaker pacing were 228.2 ± 44.8 ms QRS duration, 86.5 ± 32.8ms IVD, 94.4 ± 23.8ms LVD, 2.6 ± 0.8 QRS-IVD-ratio with correlation between IVD and QRS-IVD-ratio (r=-0.668 P=0.0248) and 2.3 ± 0.7 QRS-LVD-ratio. The LVEF-IVD-ratio was 0.3 ± 0.1 with correlation between IVD and LVEF-IVD-ratio (r=-0.8063 P=0.00272) and with correlation between QRS duration and LVEF-IVD-ratio (r=-0.7251 P=0.01157). Optimal sensing and pacing AV delay were 128.3 ± 24.8 ms AV delay after atrial sensing (n=6) and 173.3 ± 40.4 ms AV delay after atrial pacing (n=3). Optimal BV pacing delay was -4.3 ± 11.3 ms between LV and RV pacing (n=7). During 30.4 ± 29.6 month CRT follow-up, the NYHA class improved from 3.1 ± 0.2 to 2.2 ± 0.3.
Conclusions: Transesophageal electrical IVD and LVD in RV pacemaker pacing may be additional useful ventricular desynchronization parameters to improve P selection for upgrading RV pacemaker pacing to CRT BV pacing.
Introduction: Cardiac resynchronisation therapy (CRT) with atrioventricular (AV) and interventricular (VV) optimized biventricular pacing (BV) is an established therapy for heart failure (HF) patients with electrical interventricular conduction delay (IVCD). The aim of the study was to compare AV and VV delay optimization with cardiac output (CO) and acceleration index (ACI) impedance cardiographic (ICG) methods.
Methods: HF patients with IVCD 86.8 ± 33 ms (n=15, age 66 ± 10 years; 2 females, 13 males), New York Heart Association (NYHA) functional class 3.1 ± 0.4, left ventricular (LV) ejection fraction 21.3 ± 7.8 % and QRS duration 176.1 ± 31.7 ms underwent AV and VV delay optimization with CO and ACI methods (Cardioscreen, Medis GmbH, Ilmenau, Germany). After evaluation of optimal AV delay, we evaluated optimal VV delay during simultaneous LV and right ventricular (RV) pacing (LV=RV), LV before RV pacing (LV-RV) and RV before LV pacing (RV-LV).
Results: Optimal VV delay was -12.3 ± 25.9 ms LV-RV pacing with VV delay range from -80 ms LV-RV pacing to +20 ms RV-LV pacing and RV=LV pacing. Optimal AV delay after atrial sensing was 108.6 ± 20.3 ms (n=14) and optimal AV delay after atrial pacing 190 ± 14.1 ms (n=2) with AV delay range from 80 ms to 200 ms. RV versus BV pacing mode resulted in improvement of CO from 3.4 ± 1.2 l/min to 4.4 ± 1.4 l/min (p<0.001) and ACI from 0.667 ± 0.227 1/s² to 0.834 ± 0.282 1/s² (p<0.002). During 34 ± 26 month BV pacing, the NYHA class improved from 3.1 ± 0.4 to 2.1 ± 0.4 (p<0.001).
Conclusion: AV and VV delay optimized BV pacing acutely improve ICG CO and ACI and their NYHA class during long-term follow-up. ICG may be a simple and useful technique to optimize AV and VV delay in CRT.
Capture threshold (CT) for transesophageal left atrial (LA) pacing (TLAP) and transesophageal left ventricular (LV) pacing (TLVP) with conventional cylindrical electrodes (CE) are higher than TLAP feeling threshold (FT). Purpose of the study was to evaluate focused TLAP CT and FT for supraventricular tachycardia (SVT) initiation and focused TLVP CT for cardiac resynchronisation therapy (CRT) simulation.
Methods: SVT initiation in patients (P) with palpitations (n=49, age 47 ± 17 years) was analysed during spontaneous rhythm and during focused bipolar TLAP with atrial constant current stimulus output, distal CE and three or seven 6 mm hemispherical electrodes (HE) (TO, Osypka AG, Rheinfelden, Germany). CRT simulation in heart failure P (n=75, age 62 ± 11 years) was evaluated by focused bipolar TLAP and/or TLVP with ventricular constant voltage stimulus output and different pacing mode.
Results: Focused electrical pacing field between CE and HE (n=28) allowed low threshold TLAP with 8.0 ± 2.6 mA CT at 9.9 ms stimulus duration (SD) which was lower than 9.2 ± 4.5 mA FT at 9.9 ms SD. Focused electrical pacing field between HE and HE (n=21) allowed low threshold TLAP with 8.1 ± 2.2 mA CT at 9.9 ms SD which was lower than 9.8 ± 5.0 mA FT at 9.9 ms SD. SVT initiation by programmed AAI TLAP was possible in 23 P and not possible in 26 P. CRT simulation was evaluated with TLAP and TLVP with VAT, D00 and V00 pacing mode and 95.5 ± 10.9 V TLVP CT at 4.0 ms SD.
Conclusions: Programmed focused AAI TLAP allowed initiation of SVT with very low CT and high FT and focused electrical pacing field between CE-HE and HE-HE.CRT simulation with focused TLAP and/or TLVP with VAT, D00 and V00 pacing mode may be a useful technique to detect responders to CRT.
Cardiac resynchronization therapy (CRT) with hemodynamic optimized biventricular pacing is an established therapy for heart failure patients with sinus rhythm, reduced left ventricular ejection fraction and wide QRS complex. The aim of the study was to evaluate electrical right and left cardiac atrioventricular delay and left atrial delay in CRT responder and non-responder with sinus rhythm.
Methods: Heart failure patients with New York Heart Association class 3.0 ± 0.3, sinus rhythm and 27.7 ± 6.1% left ventricular ejection fraction were measured by surface ECG and transesophageal bipolar left atrial and left ventricular ECG before implantation of CRT devices. Electrical right cardiac atrioventricular delay was measured between onset of P wave and onset of QRS complex in the surface ECG, left cardiac atrioventricular delay between onset of left atrial signal and onset of left ventricular signal in the transesophageal ECG and left atrial delay between onset and offset of left atrial signal in the transesophageal ECG.
Results: Electrical atrioventricular and left atrial delay were 196.9 ± 38.7 ms right and 194.5 ± 44.9 ms left cardiac atrioventricular delay, and 47.7 ± 13.9 ms left atrial delay. There were positive correlation between right and left cardiac atrioventricular delay (r = 0.803 P < 0.001) and negative correlation between left atrial delay and left ventricular ejection fraction (r = −0.694 P = 0.026) with 67% CRT responder.
Conclusions: Transesophageal electrical left cardiac atrioventricular delay and left atrial delay may be useful preoperative atrial desynchronization parameters to improve CRT optimization.
Cardiac resynchronization therapy (CRT) is an established biventricular pacing therapy in heart failure patients with left bundle branch block and reduced left ventricular ejection fraction, but not all patients improved clinically as CRT responder. Purpose of the study was to evaluate electrical left atrial conduction delay (LACD) with focused transesophageal electrocardiography in CRT responder and CRT non-responder.
Methods: Twenty heart failure patients (age 66.6±8.2 years; 2 females, 18 males) with New York Heart Association functional class 3.0±0.3 and 174.2±40.2ms QRS duration were analysed using posterior left atrial transesophageal electrocardiography with hemispherical electrodes. Electrical LACD was measured between onset and offset of transesophageal left atrial signal before implantation of CRT devices.
Results: Electrical LACD could be evaluated by bipolar transesophageal left atrial electrocardiography using TO Osypka electrode in all heart failure patients with negative correlation between 54.7±18.1ms LACD and 24.9±6.4% left ventricular ejection fraction (r=-0.65, P=0.002). There were 16 CRT responders with reduction of New York Heart Association functional class from 3.0±0.29 to 2.1±0.2 (r=0.522, P=0.038) during 9.41±10.96 month biventricular pacing and negative correlation between 49.6±14.2ms LACD and 26.0±6.2% left ventricular ejection fraction (r=-0.533, P=0.034). There were 4 CRT non-responders with no reduction of New York Heart Association functional class from 3.0±0.4 to 2.8±0.5 (r=0.816, P=0.184) during with 13.88±16.39 month biventricular pacing and no correlation between 75.25±19.17ms LACD and 20.75±6.4% left ventricular ejection fraction (r=-0.831, P=0.169).
Conclusions: Focused transesophageal left atrial electrocardiography can be utilized to analyse electrical LACD in heart failure patients. LACD correlated negative with left ventricular ejection fraction in CRT responders. LACD may be a useful parameter to evaluate electrical left atrial desynchronization in heart failure patients.
Background: Cardiac resynchronization therapy (CRT) is an established therapy for heart failure (HF) patients (P) with reduced left ventricular (LV) ejection fraction and electrical interventricular desynchronization, but not all P improved clinically. The aim of the study was to evaluate electrical interventricular delay (IVD) to LV delay (LVD) ratio in atrial fibrillation (AF) CRT responder (R) and non-responder (NR).
Methods: AF P (n = 18, age 60.6 ± 11.4 years, 1 female, 17 males) with HF New York Heart Association (NYHA) class 3.0 ± 0.2, 25.3 ± 5.9 % LV ejection fraction and 157.8 ± 24.4 ms QRS duration (QRSD) were measured by surface ECG and focused transesophageal bipolar LV ECG before implantation of CRT pacemaker (n = 2) or CRT defibrillator (n = 16). IVD was measured between onset of QRS in the surface ECG and onset of LV signal in the LV ECG. LVD was measured between onset and offset of LV signal in the LV ECG.
Results: Electrical ventricular desynchronization in AF CRT P were 61.9 ± 26.9ms IVD, 80.6 ± 24.3ms LVD, 0.85 ± 0.41 IVD-LVD-ratio (Figure), 3.12 ± 1.89 QRSD-IVD-ratio and 2.07 ± 0.47 QRSD-LVD-ratio. There were 72.2 % AF CRT R (n = 13) with 64.2 ± 24.6ms IVD and 77.8 ± 21.6ms LVD with Pearson correlation to 0.89 ± 0.39 IVD-LVD-ratio (r = 0.87, P < 0.01; r = -0.69, P < 0.01), 2.82 ± 1.32 QRSD-IVD-ratio (r = -0.76, P < 0.01; r = 0.67, P = 0.011) and 2.13 ± 0.46 QRSD-LVD-ratio (r = 0.57, P = 0.041; r = -0.85, P < 0.01). There were 27.8% AF CRT NR (n = 5) with 56.0 ± 34.5ms IVD and 87.8 ± 31.9ms LVD without correlation to 0.74 ± 0.48 IVD-LVD-ratio, 3.88 ± 2.98 QRSD-IVD-ratio and 1.90 ± 0.48 QRSD-LVD-ratio. During 15.3 ± 13.1 month CRT follow-up, the AF CRT R NYHA class improved from 3.0 ± 0.2 to 2.2 ± 0.3 (P < 0.001). During 18.8 ± 20.7 month CRT follow-up, the AF CRT NR NYHA class not improved from 3 to 3.3 ± 0.97.
Cardiac resynchronization therapy (CRT) is an established class I level A biventricular pacing therapy in chronic heart failure patients with left bundle branch block and reduced left ventricular ejection fraction, but not all patients improved clinically. Purpose of the study was to evaluate electrical interatrial conduction delay (IACD) to interventricular conduction delay (IVCD) ratio with focused transesophageal left atrial and left ventricular electrocardiography.
Methods: Thirty eight chronic heart failure patients (age 63.4±10.2 years; 3 females, 35 males) with New York Heart Association (NYHA) functional class 3.0±0.2 and 171.71±36.17ms QRS duration were analysed using posterior left atrial and left ventricular transesophageal electrocardiography with hemispherical electrodes before CRT. Electrical IACD was measured between onset of P-wave in the surface ECG and onset of left atrial signal. Electrical IVCD was measured between onset of QRS complex in the surface ECG and onset of left ventricular signal.
Results: Electrical IACD and IVCD could be evaluated by transesophageal left atrial and left ventricular electrocardiography in all heart failure patients with correlation to 1.18±0.92 IACD-IVCD-ratio (r=-0.57, P<0.001; r=0.66, P<0.001). There were 32 CRT responder with reduction of NYHA class from 3.0±0.22 to 1.97±0.31 (P<0.001) during 16.5±18.9 month CRT with 75.19±33.49ms IACD, 78.91±24.73ms IVCD, 1.04±0.66 IACD-IVCD-ratio and correlation between IACD and IACDIVCD- ratio (r=0.84, P<0.001). There were 6 CRT nonresponder with no reduction of NYHA class from 3.0±0.3 to 2.9±0.5 during 14.3±13.7 month biventricular pacing, 50.0±28.26ms IVCD (P=0.014), 1.92±1.65 IACD-IVCD-ratio (P=0,029) and correlation between 67.0±24.9ms IACD and IACD-IVCD-ratio (r=0.85, P=0.031).
Conclusions: Focused transesophageal left atrial and left ventricular electrocardiography can be utilized to analyse electrical IACD and IVCD in heart failure patients. IACDIVDC- ratio may be a useful parameter to evaluate electrical left cardiac desynchronization in heart failure patients.
Introduction: Cardiac resynchronisation therapy (CRT) with atrioventricular (AV) and interventricular (VV) optimized biventricular pacing (BV) is an established therapy for heart failure (HF) patients. The aim of the study was to compare AV and VV delay optimization with cardiac output (CO), cardiac index (CI), contractility index (IC) and acceleration index (ACI) impedance cardiographic (ICG) methods in CRT.
Methods: 15 HF patients (age 66 ± 10 years; 2 females, 13 males) in New York Heart Association (NYHA) class 3.1 ± 0.4, left ventricular (LV) ejection fraction 21.3 ± 7.8 % and QRS duration 176.1 ± 31.7 ms underwent AV and VV delay optimization with CO, CI, IC and ACI (Cardioscreen ®, Medis GmbH, Ilmenau, Germany) at different AV and VV delay BV pacing settings versus right ventricular (RV) pacing one day after implantation of a CRT device.
Results: Optimal AV delay after atrial sensing was 108.6 ± 20.3 ms (n=14) and optimal AV delay after atrial pacing 190 ± 14.1 ms (n=2) with AV delay range from 80 ms to 200 ms. Optimal VV delay was -12.3 ± 25.9 ms left ventricular before RV pacing. RV versus BV pacing mode resulted in improvement of CO from 3.4 ± 1.2 l/min to 4.4 ± 1.4 l/min (p<0.001), CI from 1.8 ± 0.64 l/min/m² to 2.4 ± 0.78 l/min/m² (p<0.001), IC from 0.028 ± 0.011 1/s to 0.036 ± 0.013 1/s (p<0.001) and ACI from 0.667 ± 0.227 1/s² to 0.834 ± 0.282 1/s² (p<0.002). During 34 ± 26 month BV pacing, the NYHA class improved from 3.1 ± 0.4 to 2.1 ± 0.4 (p<0.001).
Conclusion: AV and VV delay optimized BV pacing acutely improve hemodynamic parameters of transthoracic ICG and their NYHA class during long-term follow-up. ICG may be a simple and useful technique to optimize AV and VV delay in CRT.
Cardiac resynchronization therapy (CRT) is an established therapy for heart failure patients and improves quality of life in patients with sinus rhythm, reduced left ventricular ejection fraction (LVEF), left bundle branch block and wide QRS duration. Since approximately sixty percent of heart failure patients have a normal QRS duration they do not benefit or respond to the CRT. Cardiac contractility modulation (CCM) releases nonexcitatoy impulses during the absolute refractory period in order to enhance the strength of the left ventricular contraction. The aim of the investigation was to evaluate differences in cardiac index between optimized and nonoptimized CRT and CCM devices versus standard values. Impedance cardiography, a noninvasive method was used to measure cardiac index (CI), a useful parameter which describes the blood volume during one minutes heart pumps related to the body surface. CRT patients indicate an increase of 39.74 percent and CCM patients an improvement of 21.89 percent more cardiac index with an optimized device.
Cardiac resynchronisation therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients with interventricular conduction delay (IVCD). The aim of the study was to evaluate transesophageal IVCD and left ventricular (LV) pacing with directed electrical pacing field (EPF) in HF patients.
Methods: HF patients were analysed with bipolar transesophageal LV electrocardiogram recording and LV pacing with constant voltage stimulus output, 4 ms stimulus duration, distal cylindrical electrode (CE) and seven 6 mm hemispherical electrodes (HE) with 15 mm electrode distance (TO, Dr. Osypka, Rheinfelden, Germany).
Results: LV electrocardiogram recording with HE-HE and CE-HE evaluated a mean IVCD of 79.9 ± 36.7 ms. Directed EPF with CE-HE and HE-HE allowed LV VAT (n=12) and LV D00 pacing (n=5) with a mean effective capture output of 97.35 ± 6.64 V. In 15 responders with IVCD of 87 ± 33 ms arterial pulse pressure (PP) increased from 65 ± 24 mmHg to 79 ± 27 mmHg (p < 0.001). EPF was simulated with finite element method.
Conclusions: Transesophageal LV electrocardiography and directed EPF pacing with CE and HE allowed the evaluation of IVCD and PP to select patients for BV pacing.
Termination of atrial flutter (AFL) is not possible in all AFL patients (P) with transesophageal left atrial pacing (TLAP) with undirected electrical pacing field (EPF) and high atrial pacing threshold. Purpose of the study was to evaluate bipo-lar transesophageal left atrial electrocardiography (TLAE) and TLAP with directed EPF for evaluation and termination of AFL with and without simultaneous transesophageal echocardiography (TEE).
Methods: AFL P were analysed using either a TO electrode with one cylindrical (CE) and three or seven hemispherical electrodes (HE) or TEE electrode with four HE (Osypka, Rheinfelden, Germany). Burst TLAP cycle length was between 200msand 50ms.
Results: AFL cycle length was 233±30 ms with mean ventricular cycle length of 540±149 ms. AFL could be terminated by rapid bipolar TLAP with directed EPF using HE-HE and CE-HE with induction of atrial fibrillation (AF), induction of AF and spontaneous conversion to sinus rhythm and direct conversion to sinus rhythm. Directed EPF was simulated with finite element method.
Conclusions: AFL can be evaluated by bipolar TLAE. AFL can be terminated with rapid TLAP with directed EPF with and without simultaneous TEE. Bipolar TLAE with rapid TLAP is a safe, simple and useful method for evaluation and termination of AFL.