Refine
Year of publication
Document Type
- Conference Proceeding (29) (remove)
Conference Type
- Konferenz-Abstract (22)
- Konferenzartikel (4)
- Konferenz-Poster (3)
Language
- English (29) (remove)
Is part of the Bibliography
- yes (29) (remove)
Keywords
- Herzkrankheit (3)
- CST (2)
- HF-Ablation (2)
- Heart rhythm model (2)
- Herzschrittmacher (2)
- Modeling and simulation (2)
- Signaltechnik (2)
- cardiac ablation (2)
- heart rhythm model (2)
- 3D print (1)
- 3D-Modelling (1)
- AV nodal reentry tachycardia (1)
- AV reentrant tachycardia (1)
- Bioimpedance measurement (1)
- CRT (1)
- Cardiac Resynchronization Therapy (1)
- Cryoballoon catheter ablation (1)
- Elektrokardiogramm (1)
- Esophageal catheter (1)
- Heart Rhythm Simulation (1)
- Hemodynamic monitoring (1)
- Herzmuskelkrankheit (1)
- His-Bundle Pacing (1)
- IVD (1)
- Physiological Pacing (1)
- Spinal cord stimulation (1)
- Supraventricular tachycardia (1)
- Telemetrie (1)
- atrial fibrillation (1)
- atrial flutter (1)
- biomechanical stimulation (1)
- heart rhythm simulation (1)
- impedance cardiography (1)
- pulmonary vein isolation (1)
- signal averaging (1)
- ventricular tachycardia (1)
- wavelet transformation (1)
- Überwachung (1)
Institute
Open Access
- Open Access (29) (remove)
Termination of atrial flutter (AFL) is not possible in all AFL patients (P) with transesophageal left atrial pacing (TLAP) with undirected electrical pacing field (EPF) and high atrial pacing threshold. Purpose of the study was to evaluate bipo-lar transesophageal left atrial electrocardiography (TLAE) and TLAP with directed EPF for evaluation and termination of AFL with and without simultaneous transesophageal echocardiography (TEE).
Methods: AFL P were analysed using either a TO electrode with one cylindrical (CE) and three or seven hemispherical electrodes (HE) or TEE electrode with four HE (Osypka, Rheinfelden, Germany). Burst TLAP cycle length was between 200msand 50ms.
Results: AFL cycle length was 233±30 ms with mean ventricular cycle length of 540±149 ms. AFL could be terminated by rapid bipolar TLAP with directed EPF using HE-HE and CE-HE with induction of atrial fibrillation (AF), induction of AF and spontaneous conversion to sinus rhythm and direct conversion to sinus rhythm. Directed EPF was simulated with finite element method.
Conclusions: AFL can be evaluated by bipolar TLAE. AFL can be terminated with rapid TLAP with directed EPF with and without simultaneous TEE. Bipolar TLAE with rapid TLAP is a safe, simple and useful method for evaluation and termination of AFL.
Cardiac resynchronisation therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients with interventricular conduction delay (IVCD). The aim of the study was to evaluate transesophageal IVCD and left ventricular (LV) pacing with directed electrical pacing field (EPF) in HF patients.
Methods: HF patients were analysed with bipolar transesophageal LV electrocardiogram recording and LV pacing with constant voltage stimulus output, 4 ms stimulus duration, distal cylindrical electrode (CE) and seven 6 mm hemispherical electrodes (HE) with 15 mm electrode distance (TO, Dr. Osypka, Rheinfelden, Germany).
Results: LV electrocardiogram recording with HE-HE and CE-HE evaluated a mean IVCD of 79.9 ± 36.7 ms. Directed EPF with CE-HE and HE-HE allowed LV VAT (n=12) and LV D00 pacing (n=5) with a mean effective capture output of 97.35 ± 6.64 V. In 15 responders with IVCD of 87 ± 33 ms arterial pulse pressure (PP) increased from 65 ± 24 mmHg to 79 ± 27 mmHg (p < 0.001). EPF was simulated with finite element method.
Conclusions: Transesophageal LV electrocardiography and directed EPF pacing with CE and HE allowed the evaluation of IVCD and PP to select patients for BV pacing.
In cardiac resynchronization therapy (CRT) for heart failure, individualization of the AV delay is essential to improve hemodynamics and to minimize non-responder rate. In patients in sinus rhythm having additional disposition to bradycardia, optimization is necessary for both situations, atrial sensing and pacing. Therefore, echo-optimization is the goldstandard but time consuming. Unfortunately, it depends on the particular CRT systems parameter set if the resulting individually optimal AV delays can be programmed or not. Some CRT systems provide a set of AV delays for DDD operation combined with a set of the pace-sense-compensation to optimize the AV delay in DDD and VDD operation. The pace-sense-compensation (PSC) can be defined by the difference of implant-related interatrial conduction intervals in DDD and VDD operation measured in the esophageal left atrial electrogram. In a cohort of 96 CRT patients we found mean PSC of 59-35ms ranging between 0-143ms. As a consequence, allowing 10ms tolerance, AVD optimization is completely impossible in one of the two modes, VDD or DDD operation, in 34 (35%) or 5 (5%) patients with implants restricting the PSC range to 60ms or 100ms, respectively. Thus, we propose companies to provide CRT systems with programmable pace-sense- compensation between 0ms and 150ms.
Responder-rate in cardiac resynchronization therapy (CRT) of patients in sinus rhythm (SR) or atrial fibrillation (AF) mainly depends on accurat selection, optimal position of the left ventricular electrode and individualization of hemodynamical parameters of the implanted biventricular pacing system during follow-up. High resolution esophageal left heart electrocardiography offers a quick and semi-invasive approach to the electrical activity of left atrium and left ventricle. It was used in 62 heart failure patients in sinus rhythm and 11 in atrial fibrillation after implantation of CRT systems to compare the semi-invasive interventricular conduction delay (IVCDE) with QRS width. In all of the patients, guideline decision for CRT was linked with IVCDE of about 40ms and up. From logical point of view, IVCDE provides the minimal target interval for the left ventricular electrode placement in order to exclude non-responders. Esophageal measurement of interatrial conduction intervals in VDD and DDD pacing was utilized to individualize the AV delay and to exclude adverse hemodynamic effects.
Capture threshold (CT) for transesophageal left atrial (LA) pacing (TLAP) and transesophageal left ventricular (LV) pacing (TLVP) with conventional cylindrical electrodes (CE) are higher than TLAP feeling threshold (FT). Purpose of the study was to evaluate focused TLAP CT and FT for supraventricular tachycardia (SVT) initiation and focused TLVP CT for cardiac resynchronisation therapy (CRT) simulation.
Methods: SVT initiation in patients (P) with palpitations (n=49, age 47 ± 17 years) was analysed during spontaneous rhythm and during focused bipolar TLAP with atrial constant current stimulus output, distal CE and three or seven 6 mm hemispherical electrodes (HE) (TO, Osypka AG, Rheinfelden, Germany). CRT simulation in heart failure P (n=75, age 62 ± 11 years) was evaluated by focused bipolar TLAP and/or TLVP with ventricular constant voltage stimulus output and different pacing mode.
Results: Focused electrical pacing field between CE and HE (n=28) allowed low threshold TLAP with 8.0 ± 2.6 mA CT at 9.9 ms stimulus duration (SD) which was lower than 9.2 ± 4.5 mA FT at 9.9 ms SD. Focused electrical pacing field between HE and HE (n=21) allowed low threshold TLAP with 8.1 ± 2.2 mA CT at 9.9 ms SD which was lower than 9.8 ± 5.0 mA FT at 9.9 ms SD. SVT initiation by programmed AAI TLAP was possible in 23 P and not possible in 26 P. CRT simulation was evaluated with TLAP and TLVP with VAT, D00 and V00 pacing mode and 95.5 ± 10.9 V TLVP CT at 4.0 ms SD.
Conclusions: Programmed focused AAI TLAP allowed initiation of SVT with very low CT and high FT and focused electrical pacing field between CE-HE and HE-HE.CRT simulation with focused TLAP and/or TLVP with VAT, D00 and V00 pacing mode may be a useful technique to detect responders to CRT.
Cardiac resynchronization therapy (CRT) with biventricular (BV) pacing is an established therapy in approximately two-thirds of symptomatic heart failure (HF) patients (P) with left bundle branch block (LBBB). The aim of this study was to evaluate left atrial (LA) conduction delay (LACD) and left ventricular (LV) conduction delay (LVCD) using pre-implantational transesophageal electrocardiography (ECG) in sinus rhythm (SR) CRT responder (R) and non-responder (NR).
Methods: SR HF P (n=52, age 63.6±10.4 years; 6 females, 46 males) with New York Heart Association (NYHA) class 3.0±0.2, 24.4±7.1 % LV ejection fraction and 171.2±37.6 ms QRS duration (QRSD) were measured by bipolar filtered transesophageal LA and LV ECG recording with hemispherical electrodes (HE) TO catheter (Osypka AG, Rheinfelden, Germany). LACD was measured between onset of P-wave in the surface ECG and onset of LA deflection in the LA ECG. LVCD was measured between onset of QRS in the surface ECG and onset of LV deflection in the LV ECG.
Results: There were 78.8 % SR CRT R (n=41) with 171.2±36.9 ms QRSD, 73.3±25.7 ms LACD, 80.0±24.0 ms LVCD and 2.3±0.5 QRSD-LVCD-ratio. SR CRT R QRSD correlated with LACD (r=0.688, P<0.001) and LVCD (r=0.699, P<0.001). There were 21.2 % SR CRT NR (n=11) with 153.4±22.4 ms QRSD (P=0.133), 69.8±24.8 ms LACD (n=6, P=0.767), 54.2±31.0 ms LVCD (P<0.0046) and 3.9±2.5 QRSD-LVCD-ratio (P<0.001). SR CRT NR QRSD not corre-lated with IACD (r=-0.218, P=0.678) and IVCD (r=0.042, P=0.903). During a 22.8±21.3 month CRT follow-up, the CRT R NYHA class improved from 3.1±0.3 to 1.9±0.3 (P<0.001). In CRT NR, NYHA class not improved (2.9±0.4 to 2.9±0.2, P=1) during 11.2±9.8 months BV pacing.
Conclusions: Transesophageal LA and LV ECG with HE can be utilized to analyse LACD and LVCD in HF P. Pre-implantational LVCD and QRSD-LVCD-ratio may be additional useful parameters to improve P selection for SR CRT.
ECG simulators, available on the market, imitate the electric activity of the heart in a simplified manner. Thus, they are suitable for education purposes but not really for testing algorithms to recognize complex arrhythmias needed for pacemakers and implantable defibrillators. Especially certain discrimination between various morphologies of atrial and ventricular fibrillation needs simulators providing native electrograms of different patients’ heart rhythm events. This explains the necessity to develop an ECG simulator providing high-resolution native intracardiac and surface electrograms of in-vivo rhythm events. In this paper we demonstrate an approach for an ECG simulator based on a consumer multichannel soundcard and a corresponding software application for a laptop computer. This Live-ECG Simulator is able to handle invasive electrogram recordings from electrophysiological studies and send the data to a modified external soundcard for subsequent digital to analog conversion. The hardware is completed with an electronic circuit providing level adjustment to adapt the output amplitude to the input conditions of several cardiac implants.
Cardiac resynchronization therapy with biventricular pacing is an established therapy for heart failure patients with electrical left ventricular desynchronization. The aim of this study was to evaluate left atrial conduction delay, intra left atrial conduction delay, left ventricular conduction delay and intra left ventricular conduction delay in heart failure patients using novel signal averaging transesophageal left heart ECG software.
Methods: 8 heart failure patients with dilated cardiomyopathy (DCM), age 68 ± 9 years, New York Heart Association (NYHA) class 2.9 ± 0.2, 24.8 ± 6.7 % left ventricular ejection fraction, 188.8 ± 15.5 ms QRS duration and 8 heart failure patients with ischaemic cardiomyopathy (ICM), age 67 ± 8 years, NYHA class 2.9 ± 0.3, 32.5 ± 7.4 % left ventricular ejection fraction and 167.6 ± 19.4 ms QRS duration were analysed with transesophageal and transthoracic ECG by Bard LabDuo EP system and novel National Intruments LabView signal averaging ECG software.
Results: The electrical left atrial conduction delay was 71.3 ± 17.6 ms in ICM versus 72.3 ± 12.4 ms in DCM, intra left atrial conduction delay 66.8 ± 8.6 ms in ICM versus 63.4 ± 10.9 ms in DCM and left cardiac AV delay 180.5 ± 32.6 ms in ICM versus 152.4 ± 30.4 ms in DCM. The electrical left ventricular conduction delay was 40.9 ± 7.5 ms in ICM versus 42.6 ± 17 ms in DCM and intra left ventricular conduction delay 105.6 ± 19.3 ms in ICM versus 128.3 ± 24.1 ms in DCM.
Conclusions: Left heart signal averaging ECG can be utilized to analyse left atrial conduction delay, intra left atrial conduction delay, left ventricular conduction delay and intra left ventricular conduction delay to improve patient selection for cardiac resynchronization therapy.
Cardiac resynchronization therapy (CRT) with biventricular pacing is an established therapy for heart failure (HF) patients (P) with ventricular desynchronization and reduced left ventricular (LV) ejection fraction. The aim of this study was to evaluate electrical right atrial (RA), left atrial (LA), right ventricular (RV) and LV conduction delay with novel telemetric signal averaging electrocardiography (SAECG) in implantable cardioverter defibrillator (ICD) P to better select P for CRT and to improve hemodynamics in cardiac pacing.
Methods: ICD-P (n=8, age 70.8 ± 9.0 years; 2 females, 6 males) with VVI-ICD (n=4), DDD-ICD (n=3) and CRT-ICD (n=1) (Medtronic, Inc., Minneapolis, MN, USA) were analysed with telemetric ECG recording by Medronic programmer 2090, ECG cable 2090AB, PCSU1000 oscilloscope with Pc-Lab2000 software (Velleman®) and novel National Intruments LabView SAECG software.
Results: Electrical RA conduction delay (RACD) was measured between onset and offset of RA deflection in the RAECG. Interatrial conduction delay (IACD) was measured between onset of RA deflection and onset of far-field LA deflection in the RAECG. Interventricular conduction delay (IVCD) was measured between onset of RV deflection in the RVECG and onset of LV deflection in the LVECG. Telemetric SAECG recording was possible in all ICD-P with a mean of 11.7 ± 4.4 SAECG heart beats, 97.6 ± 33.7 ms QRS duration, 81.5 ± 44.6 ms RACD, 62.8 ± 28.4 ms RV conduction delay, 143.7 ± 71.4 ms right cardiac AV delay, 41.5 ms LA conduction delay, 101.6 ms LV conduction delay, 176.8 ms left cardiac AV delay, 53.6 ms IACD and 93 ms IVCD.
Conclusions: Determination of RA, LA, RV and LV conduction delay, IACD, IVCD, right and left cardiac AV delay by telemetric SAECG recording using LabView SAECG technique may be useful parameters of atrial and ventricular desynchronization to improve P selection for CRT and hemodynamics in cardiac pacing.
New frontiers of supraventricular tachycardia and atrial flutter evaluation and catheter ablation
(2012)
Radiofrequency catheter ablation (RFCA) has revolutionized treatment for tachyarrhythmias and has become first-line therapy for some tachycardias. Although developed in the 1980s and widely applied in the 1990s, the technique is still in development. Transesophageal atrial pacing (TAP) can used for initiation and termination of supraventricular tachycardia (SVT).
Methods: The paroxysmal SVT include a wide spectrum of disorders including, in descending order of frequency, atrial flutter, atrioventricular (AV) nodal reentry, Wolff-Parkinson-White syndrome, and atrial tachycardia. While not life-threatening in most cases, they may cause important symptoms, such as palpitations, chest discomfort, breathlessness, anxiety, and syncope, which significantly impair quality of life. Medical therapy has variable efficacy, and most patients are not rendered free of symptoms. Research over the past several decades has revealed fundamental mechanisms involved in the initiation and maintenance of all of these arrhythmias. Knowledge of mechanisms has in turn led to highly effective surgical and catheter-based treatments. The supraventricular arrhythmias and their treatment are described in this report. SVT initiation was analysed with programmed TAP in 49 patients with palpitations (age 47 ± 17 years, 24 females, 25 males).
Results: In comparison to antiarrhythmic drug therapy the radiofrequency catheter ablation in patients suffering from atrial flutter, atrioventricular nodal reentry, atrioventricular reentry and atrial tachycardia is the better choice in most cases. TAP SVT initiation was possible in 23 patients before RFCA. Atrial cycle length of SVT was 320 ± 59 ms. We initiated AV nodal reentrant tachycardia (AVNRT, n=15), atrial tachycardia (AT, n=6) and AV reentrant tachycardia with Kent pathway conduction (AVRT, n=2) before RFCA.
Conclusions: Radiofrequency catheter ablation is a successful and safe method to cure most patients with paroxysmal supraventricular tachycardias. TAP allowed initiation and termination of SVT especially in outpatients.
Cardiac resynchronization therapy with atrioventricular and interventricular pacing delay optimized biventricular pacing is an established therapy for heart failure patients with sinus rhythm and reduced left ventricular ejection fraction. The aim of the study was to evaluate atrioventricular and interventricular pacing delay optimization in cardiac resynchroniza-tion therapy by transthoracic impedance cardiography in biventricular pacing with different left ventricular electrode po-sition. In biventricular pacing heart failure patients with lateral, posterolateral and anterolateral left ventricular electrode position, the mean optimal atrioventricular sening delay was 108.6 ± 20.3 ms and the mean optimal interventricular pac-ing delay -12.3 ± 25.9 ms. Transthoracic impedance cardiography may be a useful technique to optimize atrioventricular and interventricular pacing delay in biventricular pacing with different left ventricular electrode position.
Cardiac resynchronization therapy is an established therapy for heart failure patients with sinus rhythm, reduced left ventricular ejection fraction and prolongation of QRS duration. The aim of the study was to evaluate ventricular desynchronization with electrical interventricular delay (IVD) to left ventricular delay (LVD) ratio in atrial fibrillation heart failure patients. IVD and LVD were measured by transesophageal posterior left ventricular ECG recording. In atrial fibrillation heart failure patients with prolonged QRS duration, the mean IVD-to-LVD-ratio was 0.84 +/- 0.42 with a range from 0.17 to 2.2 IVD-to-LVD-ratio. IVD-to-LVD-ratio correlated with QRS duration. IVD-to-LVD-ratio may be a useful parameter to evaluate electrical ventricular desynchronization in atrial fibrillation heart failure patients.
Transcatheter aortiv valve implantation is a new safe strategy treatment for patients with symptomatic severe aortic stenosis and high operative risk. The aim of the study was to compare the pre-and post- muiscatheter aortiv valve implantation procedures to determine the atrioventricuktr conduction time as a potential predictor of permanent pacemaker therapy requirement after transcatheter aortiv valve implantation. The transcatheter aortiv valve implantation patients were divided into groups without pacemaker and with dual or single chamber pacemEtker with diffent atrioventrieular conduction time disturbance before and after transcatheter aortiv valve implantation. In heart failure, patients without permanent pacemaker therapy after transcatheter aortiv valve implantation, atrioventricular conduction time was prolonged after transcatheter aortiv valve implantation. In patients with permanent dual chamber pacemaker therapy after transcatheter aortiv valve implantation, atrioventricular conduction time was normalised with dual chaniber atrioventrieuku pacing mode. Atrioventricular conduction time may be a useful parameter to evaluate the risk of post-procedural atrioventricular conduction block and permanent pacemaker therapy in transcatheter north, valve implantation patients.
In contrast to conventional aortic valve replacement, the Transcatheter Aortic Valve Implantation (TAVI) is a new highly specialist alternative to surgical valve replacement for patients with symptomatic severe aortic stenosis and high operative risk. The procedure was performed in a minimally invasive way and was introduced at the university heart centre, Freiburg – Bad Krozingen in 2008. The results have been getting better and better over the years. The aim of the investigation is the analysis of electrocardiogram conduction time and the electrocardiography changes recorded hours and days after the procedure depending on artificial heart valve models, which may lead to pacemaker implantation, even the analysis of the effectiveness of treatment.
Transthoracic impedance cardiography (ICG) is a non-invasive method for determination of hemodynamic parameters. The basic principle of transthoracic ICG is the measurement of electrical conductivity of the thorax over the time. The aim of the study was the analysis of hemodynamic parameters from healthy individuals and the evaluation of various hemodynamic monitoring devices. Fourteen men (mean age 25 ± 4.59 years) and twelve women (mean age 24 ± 3.5 years) were measured during the cardiovascular engineering laboratory at Offenburg University of Applied Sciences, Offenburg, Germany. The ICG recordings were measured with the devices CardioScreen 1000, CardioScreen 2000 and TensoScreen with the corresponding Software Cardiovascular Lab 2.5 (Medis Medizinische Messtechnik GmbH, Illmenau, Germany). In order to create identical frame conditions, all measurements were recorded in the same position and for the same duration. Various positions were simulated from horizontal lying position to vertical standing position. Altogether, more than 30 hemodynamic parameters were measured.
The electrical field (E-field) of the biventricular (BV) stimulation is important for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex.
The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of BV stimulation and HF ablation in atrial fibrillation (AF).
The modeling and simulation was carried out using the electromagnetic simulation software CST. Five multipolar left ventricular (LV) electrodes, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modelled. A selection were integrated into the heart rhythm model (Schalk, Offenburg) for the electrical field simulation. The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
The BV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode with a pulse width of 0.5 ms each. The far-field potential at the RA electrode tip was 32.86 mV and 185.97 mV at a distance of 1 mm from the RA electrode tip. AV node ablation was simulated with an applied power of 5 W at 420 kHz at the distal ablation electrode. The temperature at the catheter tip was 103.87 °C after 5 s ablation time and 37.61 °C at a distance of 2 mm inside the myocardium. After 15 s, the temperature was 118.42 °C and 42.13 °C.
Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF and could be used to optimize the CRT and AF ablation.
Spectral analysis of signal averaging electrocardiography in atrial and ventricular tachyarrhythmias
(2017)
Background: Targeting complex fractionated atrial electrograms detected by automated algorithms during ablation of persistent atrial fibrillation has produced conflicting outcomes in previous electrophysiological studies. The aim of the investigation was to evaluate atrial and ventricular high frequency fractionated electrical signals with signal averaging technique.
Methods: Signal averaging electrocardiography (ECG) allows high resolution ECG technique to eliminate interference noise signals in the recorded ECG. The algorithm uses automatic ECG trigger function for signal averaged transthoracic, transesophageal and intracardiac ECG signals with novel LabVIEW software (National Instruments, Austin, Texas, USA). For spectral analysis we used fast fourier transformation in combination with spectro-temporal mapping and wavelet transformation for evaluation of detailed information about the frequency and intensity of high frequency atrial and ventricular signals.
Results: Spectral-temporal mapping and wavelet transformation of the signal averaged ECG allowed the evaluation of high frequency fractionated atrial signals in patients with atrial fibrillation and high frequency ventricular signals in patients with ventricular tachycardia. The analysis in the time domain evaluated fractionated atrial signals at the end of the signal averaged P-wave and fractionated ventricular signals at the end of the QRS complex. The analysis in the frequency domain evaluated high frequency fractionated atrial signals during the P-wave and high frequency fractionated ventricular signals during QRS complex. The combination of analysis in the time and frequency domain allowed the evaluation of fractionated signals during atrial and ventricular conduction.
Conclusions: Spectral analysis of signal averaging electrocardiography with novel LabVIEW software can utilized to evaluate atrial and ventricular conduction delays in patients with atrial fibrillation and ventricular tachycardia. Complex fractionated atrial electrograms may be useful parameters to evaluate electrical cardiac arrhythmogenic signals in atrial fibrillation ablation.
Heart rhythm model and simulation of electrophysiological studies and high-frequency ablations
(2017)
Background: Target of the study was to create an accurate anatomic CAD heart rhythm model, and to show its usefulness for cardiac electrophysiological studies and high-frequency ablations. The method is more careful for the patients’ health and has the potential to replace clinical studies due to its high efficiency regarding time and costs.
Methods: All natural heart components of the new HRM were based on MRI records, which guaranteed electronic functionality. The software CST was used for the construction, while CST’s material library assured genuine tissue properties. It should be applicable to simulate different heart rhythm diseases as well as various diffusions of electromagnetic fields, caused by electrophysiological conduction, inside the heart tissue.
Results: It was achievable to simulate sinus rhythm and fourteen different heart rhythm disturbance with different atrial and ventricular conduction delays. The simulated biological excitation of healthy and sick HRM were plotted by simulated electrodes of four polar right atrial catheter, six polar His bundle catheter, ten polar coronary sinus catheter, four polar ablation catheter and eight polar transesophageal left cardiac catheter. Accordingly, six variables were rebuilt and inserted into the anatomic HRM in order to establish heart catheters for ECG monitoring and HF ablation. The HF ablation catheters made it possible to simulate various types of heart rhythm disturbance ablations with different HF ablation catheters and also showed a functional visualisation of tissue heating. The use of tetrahedral meshing HRM made it attainable to store the results faster accompanied by a higher degree of space saving. The smart meshing function reduced unnecessary high resolutions for coarse structures.
Conclusions: The new HRM for EPS simulation may be additional useful for simulation of heart rhythm disturbance, cardiac pacing, HF ablation and for locating and identification of complex fractioned signals within the atrium during atrial fibrillation HF ablation.