Die Katheterablation mit Hochfrequenzstrom (HF) ist der Goldstandard für die Therapie vieler kardi-aler Tachyarrhythmien. Bei der HF-Ablation entstehen Temperaturen zwischen 50 °C und 70 °C, wo-durch bestimmte Strukturen im Herzgewebe gezielt zerstört werden können. Ziel der Studie ist, die HF-Ablation und deren Wärmeausbreitung in Bezug auf die zugeführte Leistung mit unterschiedli-chem Elektrodenmaterial und Elektrodengröße bei supraventrikülären Tachykardien zu simulieren.
Background: The application of high-frequency ablation is used for the treatment of tachycardia arrhythmias and is a respected method. Ablation with high frequency current leads to the targeted heat destruction of myocardial tissue at specific sites and thus prevents the pathological propagation of excitation through these structures.
Purpose: The aim of this study was to simulate heat propagation during RF ablation with modeled electrodes in different sizes and materials. The simulation was performed on atrioventricular node re-entry tachycardia (AVNRT), atrioventricular re-entry tachycardia (AVRT) and atrial flutter (AFL).
Methods: Using the modeling and simulation software CST, ablation catheters with 4 mm and 8 mm tip electrodes were modeled from gold and platinum for each. The designed catheters correspond to the manufacturer"s specifications of Medtronic, Biotronik and Osypka. The catheters were integrated into the Offenburg heart rhythm model to simulate and compare the heat propagation during an ablation application, which also takes into account the blood flow in the four heart chambers. A power of 5 W - 40 W was simulated for the 4 mm electrodes and a power of 50 W - 80 W for the 8 mm electrodes.
Results: During the simulated HF ablation application, the temperature at the ablation electrode was measured at different powers. This is 40.67°C at 5 W, 44.34°C at 10 W, 51.76°C at 20 W, 59.0°C at 30 W, and 66.33°C at 40 W. The measured temperature during 40 W application is 39.5°C at 0,5 mm depth in the myocardium and 37.5°C at 2 mm depth.
In the simulation, the 8 mm platinum electrode reached an ablation temperature of 72.85°C at its tip during an applied power of 60 W. In contrast, the 8 mm platinum electrode reached a depth of 5 mm at 39.5 C° and at a depth of 2 mm at 37.5 °C. In contrast, the 8 mm gold electrode reached a temperature of 64.66°C with the same performance. This is due to the thermal properties of gold, which has a better thermal conductivity than platinum.
Conclusions: CST offers the possibility to carry out a static and dynamic simulation of a heart model and the ablation electrodes integrated in it during an HF ablation. In variation with different electrode sizes and materials, therapy methods for the treatment of AVNRT, AVRT and AFL can be optimized
The high frequency (HF) catheter ablation is the gold standard for the therapy of many cardiac tachyarrhythmias, such as atrioventricular node re-entry tachycardia (AVNRT), atrioventricular re-entry tachycardia (AVRT) or atrial flutter (AFL). The aim of the study was to simulate the HF ablation of AVNRT, AVRT, AFL and its heat propagation in reference to the supplied power with different electrode material and electrode size. The modeling and simulation were performed with the thermal and electromagnetic simulation software CST® (Computer Simulation Technology, Darmstadt). The modeling and simulation were carried out using ablation catheters with 4 mm tip electrode and 8 mm tip electrode with different electrode materials. Both electrode types were made of platinum and gold respectively. For the measurement of the heat propagation in the heart tissue, the catheters were integrated in the Offenburg heart rhythm model. The HF ablation procedures were performed with the 4 mm platinum tip electrode, with an application duration of 45 seconds and a power output of 40 watts. The HF ablation of the atrioventricular node slow pathway produced a maximum temperature of 66.33 °C. The Kent bundle HF ablation in the left atrium achieved a maximum temperature of 67.14 °C. The HF ablation of the right atrial isthmus resulted 65.96 °C. The 8 mm distal platinum tip electrode and a power output of 60 watts reached 72.85 °C. The 8 mm distal gold tip electrode and a power output of 60 watt reached 64.66 °C, due to the improved thermal conductivity of gold. Virtual heart and ablation electrode models allow the static and dynamic simulation of HF ablation with different electrode material and electrode size. The 3D simulation of the temperature profile may be used to optimize the AVNRT, AVRT and AFL HF ablation.