In bimodal cochlear implant (CI) / hearing aid (HA) users a constant interaural time delay in the order of several milliseconds occurs due to differences in signal processing of the devices. For MED-EL CI systems in combination with different HA types, we have quantified the respective device delay mismatch (Zirn et al. 2015). In the current study, we investigate the effect of the device delay mismatch in simulated and actual bimodal listeners on sound localization accuracy.
To deal with the device delay mismatch in actual bimodal listeners we delayed the CI stimulation according to the measured HA processing delay and two other values. With all delay values highly significant improvements of the rms error in the localization task were observed compared to the test without the delay. The results help to narrow down the optimal patient-specific delay value.