Background: A disturbed synchronization of the ventricular contraction can cause a highly developed systolic heart failure in affected patients, which can often be explained by a diseased left bundle branch block (LBBB). If medication remains unresponsive, the concerned patients will be treated with a cardiac resynchronization therapy (CRT) system. The aim of this study was to integrate His bundle pacing into the Offenburg heart rhythm model in order to visualize the electrical pacing field generated by His bundle pacing.
Methods: Modelling and electrical field simulation activities were performed with the software CST (Computer Simulation Technology) from Dessault Systèms. CRT with biventricular pacing is to be achieved by an apical right ventricular electrode and an additional left ventricular electrode, which is floated into the coronary vein sinus. This conventional type of biventricular pacing leads to a reduction of the left ventricular ejection fraction. Furthermore, the non-responder rate of the CRT therapy is about one third of the CRT patients.
Results: His bundle pacing represents a physiological alternative to conventional cardiac pacing and cardiac resynchronization. An electrode implanted in the His bundle emits a stronger electrical pacing field than the electrical pacing field of conventional cardiac pacemakers. The pacing of the His bundle was performed by the Medtronic Select Secure 3830 electrode with pacing voltage amplitudes of 3 V, 2 V and 1.5 V in combination with a pacing pulse duration of 1 ms.
Conclusions: Compared to conventional cardiac pacemaker pacing, His bundle pacing is capable of bridging LBBB conduction disorders in the left ventricle. The His bundle pacing electrical field is able to spread via the physiological pathway in the right and left ventricles for CRT with a narrow QRS-complex in the surface ECG.
Heart rhythm model and simulation of electrophysiological studies and high-frequency ablations
(2017)
Background: Target of the study was to create an accurate anatomic CAD heart rhythm model, and to show its usefulness for cardiac electrophysiological studies and high-frequency ablations. The method is more careful for the patients’ health and has the potential to replace clinical studies due to its high efficiency regarding time and costs.
Methods: All natural heart components of the new HRM were based on MRI records, which guaranteed electronic functionality. The software CST was used for the construction, while CST’s material library assured genuine tissue properties. It should be applicable to simulate different heart rhythm diseases as well as various diffusions of electromagnetic fields, caused by electrophysiological conduction, inside the heart tissue.
Results: It was achievable to simulate sinus rhythm and fourteen different heart rhythm disturbance with different atrial and ventricular conduction delays. The simulated biological excitation of healthy and sick HRM were plotted by simulated electrodes of four polar right atrial catheter, six polar His bundle catheter, ten polar coronary sinus catheter, four polar ablation catheter and eight polar transesophageal left cardiac catheter. Accordingly, six variables were rebuilt and inserted into the anatomic HRM in order to establish heart catheters for ECG monitoring and HF ablation. The HF ablation catheters made it possible to simulate various types of heart rhythm disturbance ablations with different HF ablation catheters and also showed a functional visualisation of tissue heating. The use of tetrahedral meshing HRM made it attainable to store the results faster accompanied by a higher degree of space saving. The smart meshing function reduced unnecessary high resolutions for coarse structures.
Conclusions: The new HRM for EPS simulation may be additional useful for simulation of heart rhythm disturbance, cardiac pacing, HF ablation and for locating and identification of complex fractioned signals within the atrium during atrial fibrillation HF ablation.
Transcatheter aortiv valve implantation is a new safe strategy treatment for patients with symptomatic severe aortic stenosis and high operative risk. The aim of the study was to compare the pre-and post- muiscatheter aortiv valve implantation procedures to determine the atrioventricuktr conduction time as a potential predictor of permanent pacemaker therapy requirement after transcatheter aortiv valve implantation. The transcatheter aortiv valve implantation patients were divided into groups without pacemaker and with dual or single chamber pacemEtker with diffent atrioventrieular conduction time disturbance before and after transcatheter aortiv valve implantation. In heart failure, patients without permanent pacemaker therapy after transcatheter aortiv valve implantation, atrioventricular conduction time was prolonged after transcatheter aortiv valve implantation. In patients with permanent dual chamber pacemaker therapy after transcatheter aortiv valve implantation, atrioventricular conduction time was normalised with dual chaniber atrioventrieuku pacing mode. Atrioventricular conduction time may be a useful parameter to evaluate the risk of post-procedural atrioventricular conduction block and permanent pacemaker therapy in transcatheter north, valve implantation patients.