Cardiac resynchronisation therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients with interventricular conduction delay (IVCD). The aim of the study was to evaluate transesophageal IVCD and left ventricular (LV) pacing with directed electrical pacing field (EPF) in HF patients.
Methods: HF patients were analysed with bipolar transesophageal LV electrocardiogram recording and LV pacing with constant voltage stimulus output, 4 ms stimulus duration, distal cylindrical electrode (CE) and seven 6 mm hemispherical electrodes (HE) with 15 mm electrode distance (TO, Dr. Osypka, Rheinfelden, Germany).
Results: LV electrocardiogram recording with HE-HE and CE-HE evaluated a mean IVCD of 79.9 ± 36.7 ms. Directed EPF with CE-HE and HE-HE allowed LV VAT (n=12) and LV D00 pacing (n=5) with a mean effective capture output of 97.35 ± 6.64 V. In 15 responders with IVCD of 87 ± 33 ms arterial pulse pressure (PP) increased from 65 ± 24 mmHg to 79 ± 27 mmHg (p < 0.001). EPF was simulated with finite element method.
Conclusions: Transesophageal LV electrocardiography and directed EPF pacing with CE and HE allowed the evaluation of IVCD and PP to select patients for BV pacing.
Termination of atrial flutter (AFL) is not possible in all AFL patients (P) with transesophageal left atrial pacing (TLAP) with undirected electrical pacing field (EPF) and high atrial pacing threshold. Purpose of the study was to evaluate bipo-lar transesophageal left atrial electrocardiography (TLAE) and TLAP with directed EPF for evaluation and termination of AFL with and without simultaneous transesophageal echocardiography (TEE).
Methods: AFL P were analysed using either a TO electrode with one cylindrical (CE) and three or seven hemispherical electrodes (HE) or TEE electrode with four HE (Osypka, Rheinfelden, Germany). Burst TLAP cycle length was between 200msand 50ms.
Results: AFL cycle length was 233±30 ms with mean ventricular cycle length of 540±149 ms. AFL could be terminated by rapid bipolar TLAP with directed EPF using HE-HE and CE-HE with induction of atrial fibrillation (AF), induction of AF and spontaneous conversion to sinus rhythm and direct conversion to sinus rhythm. Directed EPF was simulated with finite element method.
Conclusions: AFL can be evaluated by bipolar TLAE. AFL can be terminated with rapid TLAP with directed EPF with and without simultaneous TEE. Bipolar TLAE with rapid TLAP is a safe, simple and useful method for evaluation and termination of AFL.