Termination of atrial flutter (AFL) is not possible in all AFL patients (P) with transesophageal left atrial pacing (TLAP) with undirected electrical pacing field (EPF) and high atrial pacing threshold. Purpose of the study was to evaluate bipo-lar transesophageal left atrial electrocardiography (TLAE) and TLAP with directed EPF for evaluation and termination of AFL with and without simultaneous transesophageal echocardiography (TEE).
Methods: AFL P were analysed using either a TO electrode with one cylindrical (CE) and three or seven hemispherical electrodes (HE) or TEE electrode with four HE (Osypka, Rheinfelden, Germany). Burst TLAP cycle length was between 200msand 50ms.
Results: AFL cycle length was 233±30 ms with mean ventricular cycle length of 540±149 ms. AFL could be terminated by rapid bipolar TLAP with directed EPF using HE-HE and CE-HE with induction of atrial fibrillation (AF), induction of AF and spontaneous conversion to sinus rhythm and direct conversion to sinus rhythm. Directed EPF was simulated with finite element method.
Conclusions: AFL can be evaluated by bipolar TLAE. AFL can be terminated with rapid TLAP with directed EPF with and without simultaneous TEE. Bipolar TLAE with rapid TLAP is a safe, simple and useful method for evaluation and termination of AFL.
Cardiac resynchronisation therapy (CRT) with biventricular pacing (BV) is an established therapy for heart failure (HF) patients with interventricular conduction delay (IVCD). The aim of the study was to evaluate transesophageal IVCD and left ventricular (LV) pacing with directed electrical pacing field (EPF) in HF patients.
Methods: HF patients were analysed with bipolar transesophageal LV electrocardiogram recording and LV pacing with constant voltage stimulus output, 4 ms stimulus duration, distal cylindrical electrode (CE) and seven 6 mm hemispherical electrodes (HE) with 15 mm electrode distance (TO, Dr. Osypka, Rheinfelden, Germany).
Results: LV electrocardiogram recording with HE-HE and CE-HE evaluated a mean IVCD of 79.9 ± 36.7 ms. Directed EPF with CE-HE and HE-HE allowed LV VAT (n=12) and LV D00 pacing (n=5) with a mean effective capture output of 97.35 ± 6.64 V. In 15 responders with IVCD of 87 ± 33 ms arterial pulse pressure (PP) increased from 65 ± 24 mmHg to 79 ± 27 mmHg (p < 0.001). EPF was simulated with finite element method.
Conclusions: Transesophageal LV electrocardiography and directed EPF pacing with CE and HE allowed the evaluation of IVCD and PP to select patients for BV pacing.
Capture threshold (CT) for transesophageal left atrial (LA) pacing (TLAP) and transesophageal left ventricular (LV) pacing (TLVP) with conventional cylindrical electrodes (CE) are higher than TLAP feeling threshold (FT). Purpose of the study was to evaluate focused TLAP CT and FT for supraventricular tachycardia (SVT) initiation and focused TLVP CT for cardiac resynchronisation therapy (CRT) simulation.
Methods: SVT initiation in patients (P) with palpitations (n=49, age 47 ± 17 years) was analysed during spontaneous rhythm and during focused bipolar TLAP with atrial constant current stimulus output, distal CE and three or seven 6 mm hemispherical electrodes (HE) (TO, Osypka AG, Rheinfelden, Germany). CRT simulation in heart failure P (n=75, age 62 ± 11 years) was evaluated by focused bipolar TLAP and/or TLVP with ventricular constant voltage stimulus output and different pacing mode.
Results: Focused electrical pacing field between CE and HE (n=28) allowed low threshold TLAP with 8.0 ± 2.6 mA CT at 9.9 ms stimulus duration (SD) which was lower than 9.2 ± 4.5 mA FT at 9.9 ms SD. Focused electrical pacing field between HE and HE (n=21) allowed low threshold TLAP with 8.1 ± 2.2 mA CT at 9.9 ms SD which was lower than 9.8 ± 5.0 mA FT at 9.9 ms SD. SVT initiation by programmed AAI TLAP was possible in 23 P and not possible in 26 P. CRT simulation was evaluated with TLAP and TLVP with VAT, D00 and V00 pacing mode and 95.5 ± 10.9 V TLVP CT at 4.0 ms SD.
Conclusions: Programmed focused AAI TLAP allowed initiation of SVT with very low CT and high FT and focused electrical pacing field between CE-HE and HE-HE.CRT simulation with focused TLAP and/or TLVP with VAT, D00 and V00 pacing mode may be a useful technique to detect responders to CRT.
Cardiac resynchronization therapy (CRT) with biventricular (BV) pacing is an established therapy in approximately two-thirds of symptomatic heart failure (HF) patients (P) with left bundle branch block (LBBB). The aim of this study was to evaluate left atrial (LA) conduction delay (LACD) and left ventricular (LV) conduction delay (LVCD) using pre-implantational transesophageal electrocardiography (ECG) in sinus rhythm (SR) CRT responder (R) and non-responder (NR).
Methods: SR HF P (n=52, age 63.6±10.4 years; 6 females, 46 males) with New York Heart Association (NYHA) class 3.0±0.2, 24.4±7.1 % LV ejection fraction and 171.2±37.6 ms QRS duration (QRSD) were measured by bipolar filtered transesophageal LA and LV ECG recording with hemispherical electrodes (HE) TO catheter (Osypka AG, Rheinfelden, Germany). LACD was measured between onset of P-wave in the surface ECG and onset of LA deflection in the LA ECG. LVCD was measured between onset of QRS in the surface ECG and onset of LV deflection in the LV ECG.
Results: There were 78.8 % SR CRT R (n=41) with 171.2±36.9 ms QRSD, 73.3±25.7 ms LACD, 80.0±24.0 ms LVCD and 2.3±0.5 QRSD-LVCD-ratio. SR CRT R QRSD correlated with LACD (r=0.688, P<0.001) and LVCD (r=0.699, P<0.001). There were 21.2 % SR CRT NR (n=11) with 153.4±22.4 ms QRSD (P=0.133), 69.8±24.8 ms LACD (n=6, P=0.767), 54.2±31.0 ms LVCD (P<0.0046) and 3.9±2.5 QRSD-LVCD-ratio (P<0.001). SR CRT NR QRSD not corre-lated with IACD (r=-0.218, P=0.678) and IVCD (r=0.042, P=0.903). During a 22.8±21.3 month CRT follow-up, the CRT R NYHA class improved from 3.1±0.3 to 1.9±0.3 (P<0.001). In CRT NR, NYHA class not improved (2.9±0.4 to 2.9±0.2, P=1) during 11.2±9.8 months BV pacing.
Conclusions: Transesophageal LA and LV ECG with HE can be utilized to analyse LACD and LVCD in HF P. Pre-implantational LVCD and QRSD-LVCD-ratio may be additional useful parameters to improve P selection for SR CRT.
Hintergrund: Das elektrische interventrikuläre Delay (IVD) und die Lage der linksventrikulären (LV) Elektrode zum Ort der spätesten LV Erregung sind bei Patienten (P) mit Herzinsuffizienz (HF), reduzierter LV Funktion und breiter QRS Dauer (QRSD) von Bedeutung für den Erfolg der kardialen Resynchronisationstherapie (CRT). Die LV Elektrokardiographie ermöglicht eine Abschätzung des elektrischen IVD. Ziel der Studie besteht in der nicht-invasiven Evaluierung des elektrischen IVD bei Patienten (P) mit Vorhofflimmern (AFib) mit und ohne CRT mit biventrikulärer (BV) Stimulation.
Methoden: Bei 49 HF P mit AFib (Alter 63,9 ± 10,8 Jahre; 43 Männer und 6 Frauen) mit New York Heart Association (NYHA) Klasse 2,9 ± 0,4, LV Ejektionsfraktion 26,03 ± 7,99 % und QRS-Dauer (QRSD) 143,69 ± 35,62 ms wurde das elektrische IVD als Intervall zwischen Beginn des QRS-Komplexes im Oberflächen EKG und Beginn des LV Signals im transösophagealen LV EKG bei 31 HF P mit AFib und bei 18 HF P mit AFib und CRT präoperativ bestimmt. Das fokussierte bipolare LV EKG wurde mittels Osypka TO Sonde mit halbkugelförmigen Elektroden in Höhe des maximalen LV Signals registriert.
Ergebnisse: Bei 31 HF P mit AFib betrugen QRSD 135,48 ± 38,78 ms, IVD 49,55 ± 26,38 ms, QRSD-IVD-Verhältnis 3,12 ± 1,11 und das IVD korrelierte mit der QRSD (r=0,75, P<0,001) und dem QRSD-IVD-Verhältnis (r=-0,67, P<0,001) (Fig.). Bei 18 HF P mit AFib und CRT Defibrillator betrugen QRSD 157,83 ± 24,38 ms, IVD 61,94 ± 26,88 ms, QRSD-IVD-Verhältnis 3,12 ± 1,89 und das IVD korrelierte mit der QRSD (r=0,47, P=0,049) und dem QRSD-IVD-Verhältnis (r=-0,73, P<0,001). Bei 72,2 % CRT Responder (R) (n=13) betrugen QRSD 158,15 ± 22,4 ms, IVD 64,23 ± 24,62 ms, QRSD-IVD-Verhältnis 2,82 ± 1,32 und das IVD korrelierte mit der QRSD (r=0,57, P=0,043) und dem QRSD-IVD-Verhältnis (r=-0,76, P=0,0024). Bei 27,8 % CRT Non-Responder (NR) (n=5) betrugen QRSD 157 ± 31,94 ms, IVD 56 ± 34,52 ms, QRSD-IVD-Verhältnis 3,88 ± 2,98 und das IVD korrelierte nicht mit der QRSD (r=0,33, P=0,591) und dem QRSD-IVD-Verhältnis (r=-0,732, P=0,159). Die CRT R verbesserten sich in der NYHA Klasse von 3 ± 0,2 auf 2,2 ± 0,3 (P<0,001) während 15,3 ± 13,1 Monaten BV Stimulation. Bei 15 CRT NR kam es zu keiner Verbesserung der NYHA Klasse von 3 auf 3,3 ± 0,97 (P=0,529) während 18,8 ± 20,7 Monaten BV Stimulation.
Schlussfolgerungen: Das transösophageale LV EKG ermöglicht bei HF-P mit AFib die nichtinvasive Messung des elektrischen IVD präoperativ vor CRT. IVD und QRSD-IVD-Verhältnis sind möglicherweise einfach anwendbare Parameter zur Vorhersage von CRT R und CRT NR bei P mit AFib.
Cardiac resynchronization therapy (CRT) is an established therapy for heart failure patients and improves quality of life in patients with sinus rhythm, reduced left ventricular ejection fraction (LVEF), left bundle branch block and wide QRS duration. Since approximately sixty percent of heart failure patients have a normal QRS duration they do not benefit or respond to the CRT. Cardiac contractility modulation (CCM) releases nonexcitatoy impulses during the absolute refractory period in order to enhance the strength of the left ventricular contraction. The aim of the investigation was to evaluate differences in cardiac index between optimized and nonoptimized CRT and CCM devices versus standard values. Impedance cardiography, a noninvasive method was used to measure cardiac index (CI), a useful parameter which describes the blood volume during one minutes heart pumps related to the body surface. CRT patients indicate an increase of 39.74 percent and CCM patients an improvement of 21.89 percent more cardiac index with an optimized device.
Cardiac resynchronization therapy with atrioventricular and interventricular pacing delay optimized biventricular pacing is an established therapy for heart failure patients with sinus rhythm and reduced left ventricular ejection fraction. The aim of the study was to evaluate atrioventricular and interventricular pacing delay optimization in cardiac resynchroniza-tion therapy by transthoracic impedance cardiography in biventricular pacing with different left ventricular electrode po-sition. In biventricular pacing heart failure patients with lateral, posterolateral and anterolateral left ventricular electrode position, the mean optimal atrioventricular sening delay was 108.6 ± 20.3 ms and the mean optimal interventricular pac-ing delay -12.3 ± 25.9 ms. Transthoracic impedance cardiography may be a useful technique to optimize atrioventricular and interventricular pacing delay in biventricular pacing with different left ventricular electrode position.
Cardiac resynchronization therapy is an established therapy for heart failure patients with sinus rhythm, reduced left ventricular ejection fraction and prolongation of QRS duration. The aim of the study was to evaluate ventricular desynchronization with electrical interventricular delay (IVD) to left ventricular delay (LVD) ratio in atrial fibrillation heart failure patients. IVD and LVD were measured by transesophageal posterior left ventricular ECG recording. In atrial fibrillation heart failure patients with prolonged QRS duration, the mean IVD-to-LVD-ratio was 0.84 +/- 0.42 with a range from 0.17 to 2.2 IVD-to-LVD-ratio. IVD-to-LVD-ratio correlated with QRS duration. IVD-to-LVD-ratio may be a useful parameter to evaluate electrical ventricular desynchronization in atrial fibrillation heart failure patients.