Quantification of astaxanthin in salmons by chemiluminescence and absorption after TLC separation
(2018)
Astaxanthin is a keto-carotenoid, belongs to the chemical class of terpenes and is a yellow lipid soluble compound. The compound is present in marine animals like salmons and crustacean. Its colour is due to conjugated double bonds and these double bonds are responsible for its antioxidant effect. Its antioxidant activity is ten times stronger than other carotenoids and nearly 500 fold stronger than vitamin-E. We present a new thin layer chromatography (TLC) method to measure astaxanthin on TLC-plates (Merck, 1.05554) in the visible absorption range as well as by using chemiluminescence. For separation a solvent mixture of cyclohexane and acetone (10 + 2.4, v/v) was used. The RF-value of astaxanthin is 0.14.The limit of detection in vis-absorption is 64 ng / band and the limit of quantification is 92 ng/band. In chemiluminescence the values are 90 ng / band and 115 ng/band. The method offers two independently working measurement modes on a single plate which increase the accuracy of the quantification.
We present a planar chromatographic separation method for the compounds caffeine, artemisinin, and equol, separated on high-performance thin-layer chromatography (HPTLC) silica gel plates. As solvents for separation, methyl t-butyl ether and cyclohexane (1:1, V/V) have been used for equol, cyclohexane and ethyl acetate (7:3, V/V) for artemisinin, and ethyl acetate and acetone (7:3, V/V) for caffeine. After separation, the plate was scanned with a very specific time of flight-direct analysis in real time-mass spectrometry (TOF-DART-MS) system using the (M + 1)+ signals of equol, artemisinin, and caffeine. The (M + 1) peak of artemisinin at 283.13 m/z is clearly detectable, which is the proof that DART-MS is applicable for the quantitative determination of rather instable molecules. The planar set-up of DART source, HPTLC plate and detector inlet in a line showed higher sensitivities compared to desorption at an angle. The optimal detector voltage increases with the molar mass of the analyte, thus an individual determination of optimal detector voltage setting for the different analyte is recommended to achieve the best possible measurement conditions. In conclusion, DART-MS detection in combination with an HPTLC separation allows very specific quantification of all three compounds.
Improved separation of highly toxic contact herbicides paraquat (1,1′-dimethyl-4-4′-bipyridinium), diquat (6,7-dihydrodipyridol[ 1,2-a:2′,1′-c]pyrazine-5,8-di-ium), difenzoquat (1,2-dimethyl-3,5-diphenyl-1H-pyrazolium-methyl sulfate), mepiquat (1,1-dimethyl-piperidinium), and chloromequat (2-chloroethyltrimethylammonium) were presented by high-performance thin-layer chromatography (HPTLC). The quantification is based on a derivatization reaction, using sodium tetraphenylborate. Measurements were made in the wavelength range from 500 to 535 nm, using a light-emitting diode (LED) for excitation purposes, which emits very dense light at 365 nm. For calculations, a new theory of standard addition method was used, thus leading to a minimal error if exactly the same amount of sample content is added as a standard. The method provides a fast and inexpensive approach to quantification of the five most important quats used for plant protection purposes. The method works reliably because it takes into account losses during pre-treatment procedure. The method meets the European legislation limits for paraquat and diquat in drinking water according to United States Environmental Protection Agency (US EPA) method 549.2 which are 680 ng L−1 for paraquat and 720 ng L−1 for diquat. The method of standard addition in planar chromatography can be beneficially used to reduce systematic errors. Although recovery rates of 33.7% to 65.2% are observed, calculated contents according to the method of standard addition lie between 69% and 127% of the theoretical amounts.
Limits of quantification of some neonicotinoid insecticides measured by thin-layer chromatography
(2012)
A simple method to quantify the neonicotinoid insecticides nitenpyram, thiamethoxam, acetamiprid, imidacloprid, thiacloprid and clothianidin directly on an HPTLC-plate is presented. As stationary phase silica gel 60 RP-18WF254 s plates were used and a mixture of methyl-t-butyl ether, 2-butanone, NH3 (25%) (5 + 2+0.1, v/v) was used as solvent. All neonicotinoid insecticides show light absorptions below 300 nm. The calculated limits of quantification (LOQ) by UV-detection are in the range from 12 ng to 26 ng on plate depending on the different insecticides.Nitenpyram can be stained using fast blue salt B, forming red zones. The observed LOQ is 25 ng on plate. Acetamiprid can be specifically stained using phenylglyoxylic acid forming a yellow/green fluorescent compound. The LOQ is 52 ng per spot.The compounds thiamethoxam, acetamiprid, thiacloprid and clothianidin can be transformed into blue fluorescing zones, using a relatively new staining solution. This consists of tetraphenylborate and HCl. This is the first publication mentioning that neonicotinoids undergo this reaction. The calculated limits of quantification are in the range from 10 ng to 27 ng on plate.A simple pre-treatment procedure using an acetonitrile extraction and a Chromabond SiOH clean up procedure leads to overall LOQs for bee samples of 48 to 108 µg/Kg. The method can be used to measure neonicotinoid contaminations of bees.
We present a video-densitometric quantification method for the pain killer known as diclofenac and ibuprofen. These non-steroidal anti-inflammatory drugs were separated on cyanopropyl bonded plates using CH2Cl2, methanol, cyclohexane (95 + 5 + 40, v/v) as mobile phase. The quantification is based on a bio-effective-linked analysis using Vibrio fisheri bacteria. Within 10 min a CCD-camera registered the white light of the light-emitting bacteria. Diclofenac and ibuprofen effectively suppressed the bacterial light emission which can be used for quantification within a linear range of 10 to 2000 ng. The detection limit for ibuprofen is 20 ng and the limit of quantification 26 ng per zone. Measurements were carried out using a 16-bit ST-1603ME CCD camera with 1.56 megapixels (from Santa Barbara Instrument Group, Inc., Santa Barbara, USA). The range of linearity covers more than two magnitudes because the extended Kubelka-Munk expression is used for data transformation. The separation method is inexpensive, fast, and reliable.
A diode array HPTLC method for dequalinium chloride in pharmaceutical preparations is presented. For separation a Nano TLC silica gel plate (Merck) is used with the mobile phase methanol-7.8% aqueous NH(4)(+)CH(3)COO(-) (17:3, v/v) over a distance of 6 cm. Dequalinium chloride shows an R(F) value of 0.58. Pure dequalinium chloride is measured in the wavelength range from 200 to 500 nm and shows several by-products, contour plot visualized in absorption, fluorescence and using the Kubelka-Munk transformation. Scanning of a single track in absorption and fluorescence measuring 600 spectra in the range from 200 to 1100 nm takes 30s. As a sample pre-treatment of an ointment it is simply dissolved in methanol and can be quantified in absorption from 315 to 340 nm. The same separation can also be quantified using fluorescence spectrometry in the range from 355 to 370 nm. A new staining method for dequalinium chloride, using sodium tetraphenyl borate/HCl in water allows a fluorescence quantification in the range from 445 to 485 nm. The linearity range of absorption and fluorescence measurements is from 10 to 2000 ng. Sugar-containing preparations like liquids or lozenges with a reduced sample pre-treatment can be reliably quantified only in fluorescence. The total for the quantification of an ointment sample (measuring four standards and five samples), including all sample pre-treatment steps takes less than 45 min!
We report improved separation of the highly toxic contact herbicides paraquat, diquat, difenzoquat, mepiquat, and chloromequat by HPTLC. Quantification was based on a new derivatization reaction using sodium tetraphenylborate. Measurements were in the wavelength range from 440 to 480 nm or from 440 to 590 nm. An LED emitting very intense light at 365 nm was used for excitation. The quantification limits of paraquat and diquat in water, using improved solid-phase extraction, was in the low ng L −1 range. The linear range covered more than two orders of magnitude. Recovery was investigated for all the compounds, and was insufficient, ranging from 11 to 92%, but the method is inexpensive, rapid, and works reliably.
We present an improved quantification method for urethane found in spirits. The quantification is based on a derivatization reaction using cinnamaldehyde in combination with phosphoric acid. Measurements were carried out in the wavelength range from 445 to 460 nm using a diode-TLC device. An LED was used for illumination purposes. It emits very dense light at 365 nm. The quantification range of urethane is in the lower ng range. By applying 20 µL of sprits, the urethane quantification range is from 320 µg/L to 8.1 mg urethane per litre of spirit. The range of linearity covers nearly two magnitudes. The method is cheap, fast and reliable, and is able to monitor all European legislation limits without time-consuming sample pre-treatments.