Refine
Year of publication
Document Type
- Conference Proceeding (58)
- Book (7)
- Article (unreviewed) (7)
- Article (reviewed) (5)
- Contribution to a Periodical (3)
- Part of a Book (1)
- Patent (1)
Conference Type
- Konferenz-Abstract (46)
- Konferenzartikel (10)
- Konferenz-Poster (1)
- Sonstiges (1)
Language
- English (63)
- German (18)
- Other language (1)
Is part of the Bibliography
- yes (82) (remove)
Keywords
- Abtragung (6)
- Defibrillator (5)
- Herz (5)
- Kardiale Resynchronisationstherapie (5)
- Elektrokardiogramm (4)
- Herzkrankheit (3)
- Herzschrittmacher (3)
- Hochfrequenztechnik (3)
- Katheter (3)
- Medizintechnik (3)
Institute
Open Access
- Open Access (36)
- Closed Access (27)
- Closed (4)
Transösophageales interventrikuläres Delay bei Vorhofflimmern und kardialer Resynchronisation
(2013)
Die transösophageale linksventrikuläre Elektrokardiographie ermöglicht die Evaluierung der elektrischen ventrikulären Desynchronisation im Rahmen der kardialen Resynchronisationstherapie der Herzinsuffizienz. Das Ziel der Untersuchung besteht in der präoperativen Abschätzung des transösophagealen interventrikulären Delays bei Vorhofflimmern und kardialer Resynchronisationstherapie. Bei Patienten mit Vorhofflimmern, Herzinsuffizienz New York Heart Association Klasse 3,0 ± 0,2 und QRS-Dauer 159,6 ± 23,9 ms wurde das fokusierte transösophageale linksventrikuläre EKG abgeleitet. Die kardiale Resynchronisationstherapie Responder QRS-Dauer korrelierte mit dem transösophagealen interventrikulären Delay bei Vorhofflimmern.
Das Ausmaß der elektrischen ventrikulären Desynchronisation bei reduzierter linksventrikulärer Funktion ist von Bedeutung für den Erfolg der Resynchronisationstherapie der Herzinsuffizienz mit biventrikulärer Stimulation. Das Ziel der Untersuchung besteht in der nichtinvasiven Messung der elektrischen inter-ventrikulären Desynchronisation mit und ohne ischämische Herzerkrankung bei kardialen Resynchronisationstherapie Respondern. Bei Patienten mit 25,3 ± 7,3 % reduzierter linksventrikulärer Ejektionsfraktion und 166,9 ± 38,5 ms QRS-Dauer wurde das transösophageale linksventrikuläre EKG abgeleitet. Die QRS-Dauer korrelierte mit dem interventrikulären und links-ventrikulären Delay bei Resynchronisationstherapie Respondern mit nicht-ischämischer Herzerkrankung.
Decrease of non-responder rate is the main chal-lenge in cardiac resynchronization therapy. The problem could be solved, partly, in the follow-up by consequent indi-vidualization of hemodynamic pacing parameters. The eso-phageal electrogram feature of the Biotronik ICS 3000 programmer was used in the follow-up of 20 heart failure patients carrying implants for cardiac resynchronization therapy. Adverse hemodynamic programming of the sensed and paced AV delay could be easily observed and replaced by the individual optimal duration in 3 patients (15%) VDD and DDD operation.This result proves the value of esophageal electrogram recording CRT follow-up.
Distribution of esophageal interventricular conduction delays in CRT patients and healthy subjects
(2015)
Background: R-wave synchronised atrial pacing is an effective temporary pacing
therapy in infants with postoperative junctional ectopic tachycardia. In the technique
currently used, adverse short or long intervals between atrial pacing and ventricular
sensing (AP–VS) may be observed during routine clinical practice.
Objectives: The aim of the study was to analyse outcomes of R-wave synchronised
atrial pacing and the relationship between maximum tracking rates and AP–VS
intervals.
Methods: Calculated AP–VS intervals were compared with those predicted by experienced
pediatric cardiologist.
Results: A maximum tracking rate (MTR) set 10 bpm higher than the heart rate (HR)
may result in undesirable short AP–VS intervals (minimum 83 ms). A MTR set 20 bpm
above the HR is the hemodynamically better choice (minimum 96 ms). Effects of either
setting on the AP–VS interval could not be predicted by experienced observers. In our
newly proposed technique the AP–VS interval approaches 95 ms for HR > 210 bpm
and 130 ms for HR < 130 bpm. The progression is linear and decreases strictly
(− 0.4 ms/bpm) between the two extreme levels.
Conclusions: Adjusting the AP–VS interval in the currently used technique is complex
and may imply unfavorable pacemaker settings. A new pacemaker design is advisable
to allow direct control of the AP–VS interval.