Refine
Year of publication
Document Type
- Conference Proceeding (71)
- Article (reviewed) (15)
- Patent (11)
- Contribution to a Periodical (8)
- Article (unreviewed) (2)
Conference Type
- Konferenz-Abstract (55)
- Konferenzartikel (12)
- Konferenz-Poster (4)
Language
- English (80)
- German (25)
- Other language (1)
- Multiple languages (1)
Is part of the Bibliography
- yes (107)
Keywords
- CST (7)
- HF-Ablation (7)
- CRT (6)
- Herzrhythmusmodell (6)
- Heart rhythm model (5)
- Herzkrankheit (5)
- Modeling and simulation (5)
- Kardiale Resynchronisationstherapie (4)
- Synchronisierung (4)
- heart rhythm model (4)
- Cardiac resynchronization therapy (3)
- Herzschrittmacher (3)
- Vorhofflimmern (3)
- pulmonary vein isolation (3)
- Ablation (2)
- Atrial fibrillation (2)
- Bioimpedance measurement (2)
- Cardiac Resynchronization Therapy (2)
- Cryoballoon catheter ablation (2)
- Elektrokardiogramm (2)
- IVD (2)
- Katheterablation (2)
- Kryoballonkatheter (2)
- Modellierung und Simulation (2)
- Pulmonalvenenisolation (2)
- Signaltechnik (2)
- Simulation (2)
- Supraventricular tachycardia (2)
- Triggerstabilität (2)
- atrial fibrillation (2)
- cardiac ablation (2)
- cardiac resynchronization therapy (2)
- esophageal catheter (2)
- heart rhythm simulation (2)
- 3D modeling (1)
- 3D print (1)
- 3D-Druck (1)
- 3D-Modelling (1)
- 3d Heart Simulation (1)
- 3d print (1)
- AV nodal reentry tachycardia (1)
- AV reentrant tachycardia (1)
- Atrioventricular delay (1)
- Bioimpedanzmessung (1)
- Biventricular pacing (1)
- Cardiac modelling (1)
- Catheter Ablation (1)
- EKG (1)
- EPU (1)
- Electrode Model (1)
- Electromagnetic and thermal simulation (1)
- Elektrokardiografie (1)
- Elektrophysiologie (1)
- Elektrophysiologische Untersuchung (1)
- Esophageal catheter (1)
- Esophageal electrode (1)
- Eöelktrokardiogramm (1)
- Fractionated atrial signals (1)
- Fractionated ventricular signals (1)
- HF Ablation (1)
- Heart Rhythm Simulation (1)
- Heart rhythm simulation (1)
- Heilkunde (1)
- Hemodynamic monitoring (1)
- Herz (1)
- Herzfrequenz (1)
- Herzinsuffizienz (1)
- Herzkatheterablation (1)
- Herzmuskelkrankheit (1)
- Herzrhythmusstörung (1)
- High frequency ablation (1)
- His-Bundle Pacing (1)
- His-bundle pacing (1)
- Hochfrequenzablation (1)
- Kardiologie (1)
- Left atrial ECG (1)
- Left atrial delay (1)
- Left cardiac ECG (1)
- Multiside heart stimulation (1)
- Neurostimulation (1)
- Physiological Pacing (1)
- Physiological cardiac pacing (1)
- Pulmonary vein isolation (1)
- Schrittmacher (1)
- Signal averaging ECG (1)
- Simulation und Modellierung (1)
- Spectral-temporal mapping (1)
- Spinal cord stimulation (1)
- Steuerung (1)
- Supraventricular Tachycardia (1)
- Supraventrikuläre Tachyarrhythmien (1)
- Synchronisationstherapie (1)
- Telemetrie (1)
- Transesophageal electrocardiography (1)
- Transesophageal left atrial pacing (1)
- Transesophageal neurostimulation (1)
- Transösophageale Neurostimulation (1)
- Ventricular tachycardia (1)
- Verzögerung (1)
- atrial flutter (1)
- biomechanical stimulation (1)
- biventricular pacing (1)
- biventricular stimulation (1)
- cardiac contractility modulation (1)
- cardiac resynchronisation therapy (1)
- catheter ablation (1)
- chronic heart failure (1)
- computer simulation technology (1)
- cryo ablation (1)
- cryoballoon (1)
- electrophysiological investigation (1)
- elektrische Felder (1)
- heart failure (1)
- impedance cardiography (1)
- kardiale Resynchronisationstherapie (1)
- left ventricular ECG (1)
- nuclear imaging (1)
- radiofrequency ablation (1)
- semi-invasive electrocardiography (1)
- signal averaging (1)
- spinal cord stimulation (1)
- thermische Felder (1)
- transesophageal ECG (1)
- transesophageal atrial pacing (1)
- transesophageal electrical pacing (1)
- transesophageal neurostimulation (1)
- ventricular tachycardia (1)
- virtual simulations (1)
- wavelet transformation (1)
- Ösophaguselektrodensonde (1)
- Ösophaguskatheter (1)
- Überwachung (1)
Institute
- Fakultät Elektrotechnik und Informationstechnik (E+I) (bis 03/2019) (76)
- Fakultät Elektrotechnik, Medizintechnik und Informatik (EMI) (ab 04/2019) (28)
- POIM - Peter Osypka Institute of Medical Engineering (12)
- CRT - Campus Research & Transfer (2)
- Zentrale Einrichtungen (2)
- Fakultät Maschinenbau und Verfahrenstechnik (M+V) (1)
Open Access
- Open Access (70)
- Closed Access (29)
- Bronze (4)
- Closed (2)
Cardiac resynchronization therapy is an established therapy for heart failure patients with sinus rhythm, reduced left ventricular ejection fraction and prolongation of QRS duration. The aim of the study was to evaluate ventricular desynchronization with electrical interventricular delay (IVD) to left ventricular delay (LVD) ratio in atrial fibrillation heart failure patients. IVD and LVD were measured by transesophageal posterior left ventricular ECG recording. In atrial fibrillation heart failure patients with prolonged QRS duration, the mean IVD-to-LVD-ratio was 0.84 +/- 0.42 with a range from 0.17 to 2.2 IVD-to-LVD-ratio. IVD-to-LVD-ratio correlated with QRS duration. IVD-to-LVD-ratio may be a useful parameter to evaluate electrical ventricular desynchronization in atrial fibrillation heart failure patients.
Cardiac resynchronization therapy with atrioventricular and interventricular pacing delay optimized biventricular pacing is an established therapy for heart failure patients with sinus rhythm and reduced left ventricular ejection fraction. The aim of the study was to evaluate atrioventricular and interventricular pacing delay optimization in cardiac resynchroniza-tion therapy by transthoracic impedance cardiography in biventricular pacing with different left ventricular electrode po-sition. In biventricular pacing heart failure patients with lateral, posterolateral and anterolateral left ventricular electrode position, the mean optimal atrioventricular sening delay was 108.6 ± 20.3 ms and the mean optimal interventricular pac-ing delay -12.3 ± 25.9 ms. Transthoracic impedance cardiography may be a useful technique to optimize atrioventricular and interventricular pacing delay in biventricular pacing with different left ventricular electrode position.
Das Ausmaß der elektrischen ventrikulären Desynchronisation bei reduzierter linksventrikulärer Funktion ist von Bedeutung für den Erfolg der Resynchronisationstherapie der Herzinsuffizienz mit biventrikulärer Stimulation. Das Ziel der Untersuchung besteht in der nichtinvasiven Messung der elektrischen inter-ventrikulären Desynchronisation mit und ohne ischämische Herzerkrankung bei kardialen Resynchronisationstherapie Respondern. Bei Patienten mit 25,3 ± 7,3 % reduzierter linksventrikulärer Ejektionsfraktion und 166,9 ± 38,5 ms QRS-Dauer wurde das transösophageale linksventrikuläre EKG abgeleitet. Die QRS-Dauer korrelierte mit dem interventrikulären und links-ventrikulären Delay bei Resynchronisationstherapie Respondern mit nicht-ischämischer Herzerkrankung.
Cardiac resynchronization therapy with atrioventricular and interventricular delay optimized biventricular pacing is an established therapy for symptomatic heart failure patients with prolongation of QRS duration, left bundle branch block and reduced left ventricular ejection fraction. The aim of the investigation was to evaluate right atrial, right ventricular and left ventricular electrical signals of implantable electronic cardiac devices with and without signal averaging technique with novel LabVIEW software. Electrical interatrial conduction delay and inter-ventricular conduction delay may be useful parameters to evaluate electrical atrial and ventricular desynchronization in heart failure patients.
Transcatheter aortic valve implantation is a therapy for patients with reduced left ventricular ejection fraction and symptomatic aortic stenosis. The aim of the study was to compare the pre-and post- transcatheter aortic valve implantation procedures to determine the QRS and QT ventricular conduction times as a potential predictor of permanent pacemaker therapy requirement after transcatheter aortic valve implantation. QRS and QT ventricular conduction times were prolonged after transcatheter aortic valve implantation in heart failure patients with permanent dual chamber pacemaker therapy after transcatheter aortic valve implantation. QRS and QT ventricular conduction times may be useful parameters to evaluate the risk of post-procedural ventricular conduction block and permanent pacemaker therapy in transcatheter aortic valve implantation.
Cardiac resynchronization therapy (CRT) is an established therapy for heart failure patients and improves quality of life in patients with sinus rhythm, reduced left ventricular ejection fraction (LVEF), left bundle branch block and wide QRS duration. Since approximately sixty percent of heart failure patients have a normal QRS duration they do not benefit or respond to the CRT. Cardiac contractility modulation (CCM) releases nonexcitatoy impulses during the absolute refractory period in order to enhance the strength of the left ventricular contraction. The aim of the investigation was to evaluate differences in cardiac index between optimized and nonoptimized CRT and CCM devices versus standard values. Impedance cardiography, a noninvasive method was used to measure cardiac index (CI), a useful parameter which describes the blood volume during one minutes heart pumps related to the body surface. CRT patients indicate an increase of 39.74 percent and CCM patients an improvement of 21.89 percent more cardiac index with an optimized device.
Cardiac resynchronization therapy is an established therapy for heart failure patients. The aim of the study was to evaluate electrical left cardiac atrioventricular delay and interventricular desynchronization in sinus rhythm cardiac resynchronization therapy responder and non-responder. Cardiac electrical desynchronization were measured by surface ECG and focused transesophageal bipolar left atrial and left ventricular ECG before implantation of cardiac resynchronization therapy defibrillators. Preoperative electrical cardiac desynchronization was 195.7 ± 46.7 ms left cardiac atrioventricular delay and 74.8 ± 24.5 ms interventricular delay in cardiac resynchronization therapy responder. Cardiac resynchronization therapy responder New York Heart Association class improved during long term biventricular pacing. Transesophageal left cardiac atrioventricular delay and interventricular delay may be additional useful parameters to improve patient selection for cardiac resynchronization therapy.
Background: Cardiac resynchronization therapy (CRT) is an established therapy for heart failure (HF) patients (P) with reduced left ventricular (LV) ejection fraction and electrical interventricular desynchronization, but not all P improved clinically. The aim of the study was to evaluate electrical interventricular delay (IVD) to LV delay (LVD) ratio in atrial fibrillation (AF) CRT responder (R) and non-responder (NR).
Methods: AF P (n = 18, age 60.6 ± 11.4 years, 1 female, 17 males) with HF New York Heart Association (NYHA) class 3.0 ± 0.2, 25.3 ± 5.9 % LV ejection fraction and 157.8 ± 24.4 ms QRS duration (QRSD) were measured by surface ECG and focused transesophageal bipolar LV ECG before implantation of CRT pacemaker (n = 2) or CRT defibrillator (n = 16). IVD was measured between onset of QRS in the surface ECG and onset of LV signal in the LV ECG. LVD was measured between onset and offset of LV signal in the LV ECG.
Results: Electrical ventricular desynchronization in AF CRT P were 61.9 ± 26.9ms IVD, 80.6 ± 24.3ms LVD, 0.85 ± 0.41 IVD-LVD-ratio (Figure), 3.12 ± 1.89 QRSD-IVD-ratio and 2.07 ± 0.47 QRSD-LVD-ratio. There were 72.2 % AF CRT R (n = 13) with 64.2 ± 24.6ms IVD and 77.8 ± 21.6ms LVD with Pearson correlation to 0.89 ± 0.39 IVD-LVD-ratio (r = 0.87, P < 0.01; r = -0.69, P < 0.01), 2.82 ± 1.32 QRSD-IVD-ratio (r = -0.76, P < 0.01; r = 0.67, P = 0.011) and 2.13 ± 0.46 QRSD-LVD-ratio (r = 0.57, P = 0.041; r = -0.85, P < 0.01). There were 27.8% AF CRT NR (n = 5) with 56.0 ± 34.5ms IVD and 87.8 ± 31.9ms LVD without correlation to 0.74 ± 0.48 IVD-LVD-ratio, 3.88 ± 2.98 QRSD-IVD-ratio and 1.90 ± 0.48 QRSD-LVD-ratio. During 15.3 ± 13.1 month CRT follow-up, the AF CRT R NYHA class improved from 3.0 ± 0.2 to 2.2 ± 0.3 (P < 0.001). During 18.8 ± 20.7 month CRT follow-up, the AF CRT NR NYHA class not improved from 3 to 3.3 ± 0.97.
Die Erfindung betrifft eine Ösophaguselektrodensonde bzw. einen Ösophaguskatheter 10 zur Bioimpedanzmessung und/oder zur Neurostimulation, eine Vorrichtung 100 zur transösophagealen kardiologischen Behandlung und/oder kardiologischen Diagnose und ein Verfahren zum Steuern oder Regeln einer Ablationseinrichtung zum Durchführen einer Herzablation. Die Ösophaguselektrodensonde 10 umfasst eine Bioimpedanzmesseinrichtung zur Messung der Bioimpedanz von zumindest einem Teil des die Ösophaguselektrodensonde 10 umgebenden Gewebes. Die Bioimpedanzmesseinrichtung umfasst mindestens eine erste Elektrode 12A und mindestens eine zweite Elektrode 12B, wobei die mindestens eine erste Elektrode 12A auf einer dem Herzen zugewandten Seite 14 der Ösophaguselektrodensonde 10 angeordnet ist, und die mindestens eine zweite Elektrode 12B auf einer vom Herzen abgewandten Seite 16 der Ösophaguselektrodensonde 10 angeordnet ist.Die Vorrichtung 100 umfasst die Ösophaguselektrodensonde 10 und eine Steuer- und/oder Auswerteinrichtung 30. Die Steuer- und/oder Auswerteinrichtung 30 ist eingerichtet, ein erstes Bioimpedanzmesssignal von der mindestens einen ersten Elektrode 12A und ein zweites Bioimpedanzmesssignal von der mindestens einen zweiten Elektrode 12B zu empfangen und zu vergleichen, und ein Kontrollsignal auf Basis des Vergleichs zu generieren. Das Kontrollsignal kann ein Signal zum Steuern oder Regeln einer Ablationseinrichtung zum Durchführen einer Herzablation sein.
Background: The electrical field (E-field) of the biventricular (BV) stimulation is important for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex. The 3D modeling allows the simulation of CRT and high frequency (HF) ablation.
Purpose: The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of atrial and BV stimulation and high frequency (HF) ablation in atrial fibrillation (AF).
Methods: The modeling and simulation was carried out using the electromagnetic simulation software CST (CST Darmstadt). Five multipolar left ventricular (LV) electrodes, one epicardial LV electrode, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modeled. Selected electrodes were integrated into the Offenburg heart rhythm model for the electrical field simulation. The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
Results: The right atrial stimulation was performed with an amplitude of 1.5 V with a pulse width of 0.5. The far-field potentials generated by the atrial stimulation were perceived by the right and left ventricular electrode. The far-field potential at a distance of 1 mm from the right ventricular electrode tip was 36.1 mV. The far-field potential at a distance of 1 mm from the left ventricular electrode tip was measured with 37.1 mV. The RV and LV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode with a pulse width of 0.5 ms each. The far-field potentials generated by the BV stimulations could be perceived by the RA electrode. The far-field potential at the RA electrode tip was 32.86 mV. AV node ablation was simulated with an applied power of 5 W at 420 kHz and 10 W at 500 kHz at the distal 8 mm ablation electrode.
Conclusions: Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF. The 3D simulation of the electrical field and temperature profile may be used to optimize the CRT and AF ablation.
The electrical field (E-field) of the biventricular (BV) stimulation is important for the success of cardiac resynchronization therapy (CRT) in patients with cardiac insufficiency and widened QRS complex.
The aim of the study was to model different pacing and ablation electrodes and to integrate them into a heart model for the static and dynamic simulation of BV stimulation and HF ablation in atrial fibrillation (AF).
The modeling and simulation was carried out using the electromagnetic simulation software CST. Five multipolar left ventricular (LV) electrodes, four bipolar right atrial (RA) electrodes, two right ventricular (RV) electrodes and one HF ablation catheter were modelled. A selection were integrated into the heart rhythm model (Schalk, Offenburg) for the electrical field simulation. The simulation of an AV node ablation at CRT was performed with RA, RV and LV electrodes and integrated ablation catheter with an 8 mm gold tip.
The BV stimulation were performed simultaneously at amplitude of 3 V at the LV electrode and 1 V at the RV electrode with a pulse width of 0.5 ms each. The far-field potential at the RA electrode tip was 32.86 mV and 185.97 mV at a distance of 1 mm from the RA electrode tip. AV node ablation was simulated with an applied power of 5 W at 420 kHz at the distal ablation electrode. The temperature at the catheter tip was 103.87 °C after 5 s ablation time and 37.61 °C at a distance of 2 mm inside the myocardium. After 15 s, the temperature was 118.42 °C and 42.13 °C.
Virtual heart and electrode models as well as the simulations of electrical fields and temperature profiles allow the static and dynamic simulation of atrial synchronous BV stimulation and HF ablation at AF and could be used to optimize the CRT and AF ablation.
Spectral analysis of signal averaging electrocardiography in atrial and ventricular tachyarrhythmias
(2017)
Background: Targeting complex fractionated atrial electrograms detected by automated algorithms during ablation of persistent atrial fibrillation has produced conflicting outcomes in previous electrophysiological studies. The aim of the investigation was to evaluate atrial and ventricular high frequency fractionated electrical signals with signal averaging technique.
Methods: Signal averaging electrocardiography (ECG) allows high resolution ECG technique to eliminate interference noise signals in the recorded ECG. The algorithm uses automatic ECG trigger function for signal averaged transthoracic, transesophageal and intracardiac ECG signals with novel LabVIEW software (National Instruments, Austin, Texas, USA). For spectral analysis we used fast fourier transformation in combination with spectro-temporal mapping and wavelet transformation for evaluation of detailed information about the frequency and intensity of high frequency atrial and ventricular signals.
Results: Spectral-temporal mapping and wavelet transformation of the signal averaged ECG allowed the evaluation of high frequency fractionated atrial signals in patients with atrial fibrillation and high frequency ventricular signals in patients with ventricular tachycardia. The analysis in the time domain evaluated fractionated atrial signals at the end of the signal averaged P-wave and fractionated ventricular signals at the end of the QRS complex. The analysis in the frequency domain evaluated high frequency fractionated atrial signals during the P-wave and high frequency fractionated ventricular signals during QRS complex. The combination of analysis in the time and frequency domain allowed the evaluation of fractionated signals during atrial and ventricular conduction.
Conclusions: Spectral analysis of signal averaging electrocardiography with novel LabVIEW software can utilized to evaluate atrial and ventricular conduction delays in patients with atrial fibrillation and ventricular tachycardia. Complex fractionated atrial electrograms may be useful parameters to evaluate electrical cardiac arrhythmogenic signals in atrial fibrillation ablation.
Heart rhythm model and simulation of electrophysiological studies and high-frequency ablations
(2017)
Background: Target of the study was to create an accurate anatomic CAD heart rhythm model, and to show its usefulness for cardiac electrophysiological studies and high-frequency ablations. The method is more careful for the patients’ health and has the potential to replace clinical studies due to its high efficiency regarding time and costs.
Methods: All natural heart components of the new HRM were based on MRI records, which guaranteed electronic functionality. The software CST was used for the construction, while CST’s material library assured genuine tissue properties. It should be applicable to simulate different heart rhythm diseases as well as various diffusions of electromagnetic fields, caused by electrophysiological conduction, inside the heart tissue.
Results: It was achievable to simulate sinus rhythm and fourteen different heart rhythm disturbance with different atrial and ventricular conduction delays. The simulated biological excitation of healthy and sick HRM were plotted by simulated electrodes of four polar right atrial catheter, six polar His bundle catheter, ten polar coronary sinus catheter, four polar ablation catheter and eight polar transesophageal left cardiac catheter. Accordingly, six variables were rebuilt and inserted into the anatomic HRM in order to establish heart catheters for ECG monitoring and HF ablation. The HF ablation catheters made it possible to simulate various types of heart rhythm disturbance ablations with different HF ablation catheters and also showed a functional visualisation of tissue heating. The use of tetrahedral meshing HRM made it attainable to store the results faster accompanied by a higher degree of space saving. The smart meshing function reduced unnecessary high resolutions for coarse structures.
Conclusions: The new HRM for EPS simulation may be additional useful for simulation of heart rhythm disturbance, cardiac pacing, HF ablation and for locating and identification of complex fractioned signals within the atrium during atrial fibrillation HF ablation.
Die Entwicklung von neuartigen Elektrodentypen und die Weiterentwicklung bestehender Produkten machen einen großen Teil der entstehenden Kosten für ein Unternehmen aus. Mithilfe geeigneter Software können Änderungen der Konstruktionen erfasst und bestimmte Simulationen, bspw. das Auftreten von Wechselwirkungen im elektrischen Feld, vor der eigentlichen Prototypenerstellung durchgeführt werden. Das Ziel der Studie besteht in der Modellierung unterschiedlicher Schrittmacher- und Ablationselektroden und deren Integration in das Offenburger Herzrhythmusmodell (HRM) zur statischen und dynamischen Simulation der biventrikulären Stimulation und HF Ablation bei Vorhofflimmern (AF).
Heart rhythm model and simulation of electrophysiological studies and high-frequency ablations
(2017)
Background: The simulation of complex cardiologic structures has the potential to replace clinical studies due to its high efficiency regarding time and costs. Furthermore, the method is more careful for the patients’ health than the conventional ways. The aim of the study was to create an anatomic CAD heart rhythm model (HRM) as accurate as possible, and to show its usefulness for cardiac electrophysiological studies (EPS) and high-frequency (HF) ablations.
Methods: All natural heart components of the new HRM were based on MRI records, which guaranteed electronic functionality. The software CST (Computer Simulation Technology, Darmstadt) was used for the construction, while CST’s material library assured genuine tissue properties. It should be applicable to simulate different heart rhythm diseases as well as various diffusions of electromagnetic fields, caused by electrophysiological conduction, inside the heart tissue.
Results: It was achievable to simulate normal sinus rhythm and fourteen different heart rhythm disturbance with different atrial and ventricular conduction delays. The simulated biological excitation of healthy and sick HRM were plotted by simulated electrodes of four polar right atrial catheter, six polar His bundle catheter, ten polar coronary sinus catheter, four polar ablation catheter and eight polar transesophageal left cardiac catheter (Fig.). Accordingly, six variables were rebuilt and inserted into the anatomic HRM in order to establish heart catheters for ECG monitoring and HF ablation. The HF ablation catheters made it possible to simulate various types of heart rhythm disturbance ablations with different HF ablation catheters and also showed a functional visualisation of tissue heating. The use of tetrahedral meshing HRM made it attainable to store the results faster accompanied by a higher degree of space saving. The smart meshing function reduced unnecessary high resolutions for coarse structures.
Conclusions: The new HRM for EPS simulation may be additional useful for simulation of heart rhythm disturbance, cardiac pacing, HF ablation and for locating and identification of complex fractioned signals within the atrium during atrial fibrillation HF ablation.
Hintergrund: Das elektrische interventrikuläre Delay (IVD) ist bei Patienten (P) mit Herzinsuffizienz (HF), reduzierter linksventrikulärer (LV) Funktion und verbreitertem QRS Komplex von Bedeutung für den Erfolg der kardialen Resynchronisationstherapie (CRT). Die transösophageale LV Elektrokardiographie (EKG) ermöglicht die Bestimmung des elektrischen IVD und linksventrikulären Delays (LVD). Das Ziel der Studie besteht in der Untersuchung des transösophagealen elektrischen IVD, LVD und deren Verhältnis zur QRS Dauer bei rechtsventrikulärer (RV) Stimulation vor Aufrüstung auf eine biventrikuläre (BV) Stimulation.
Methoden: Bei 11 HF P (Alter 69,0 ± 7,9 Jahre; 10 Männer und 1 Frau) mit DDD Schrittmacher (n=10), DDD Defibrillator (n=1) und RV Stimulation, New York Heart Association (NYHA) Klasse 3,0 ± 0,2, LV Ejektionsfraktion 24,5 ± 4,9 % und QRS-Dauer 228,2 ± 44,8 ms wurden das elektrische IVD als Intervall zwischen Beginn des QRS-Komplexes im Oberflächen EKG und Beginn des LV Signals im transösophagealen LV EKG und das elektrische LVD als Intervall zwischen Beginn und Ende des LV Signals im transösophagealen LV EKG präoperativ vor Aufrüstung auf CRT Defibrillator (n=8) und CRT Schrittmacher (n=3) bestimmt. Der Anstieg des arteriellen Pulse Pressure (PP) wurde zwischen RV Stimulation und transösophagealer LV Stimulation mit unterschiedlichem AV-Delay (n=5) vor Aufrüstung von RV auf BV Stimulation getestet.
Ergebnisse: Bei RV Stimulation betrugen IVD 86,54 ± 32,80 ms, LVD 94,45 ± 23,80 ms, QRS-IVD-Verhältnis 2,63 ± 0,81 mit negativer Korrelation zwischen IVD und QRS-IVD-Verhältnis (r=-0,668 P=0,0248) (Fig.) und QRS-LVD-Verhältnis 2,33 ± 0,73. Vorhofsynchrone ventrikuläre Stimulation führte zu 63,6 ± 27,7 mmHg PP bei RV Stimulation und 80,6 ± 38,5 mmHg PP bei LV Stimulation und der PP erhöhte sich bei LV Stimulation mit optimalem AV Delay um 17 ± 11,2 mmHg gegenüber RV Stimulation (P<0,001). Nach Aufrüstung von RV Stimulation auf BV Stimulation verbesserten sich die NYHA Klasse von 3,1 ± 0,2 auf 2,2 ± 0,3 während 30,4 ± 29,6 Monaten CRT.
Schlussfolgerungen: Das transösophageale LV EKG ermöglicht die Bestimmung des elektrischen IVD und LVD bei RV Stimulation zur Evaluierung der interventrikulären und linksventrikulären elektrischen Desynchronisation. IVD, LVD und deren Verhältnis zur QRS Dauer können möglicherweise zur Vorhersage einer CRT Response vor Aufrüstung von RV auf BV Stimulation genutzt werden.
Heart rhythm model and simulation of electrophysiological studies and high-frequency ablations
(2017)
Background: The simulation of complex cardiologic structures has the potential to replace clinical studies due to its high efficiency regarding time and costs. Furthermore, the method is more careful for the patients’ health than the conventional ways. The aim of the study was to create an anatomic CAD heart rhythm model (HRM) as accurate as possible, and to show its usefulness for cardiac electrophysiological studies (EPS) and high-frequency (HF) ablations.
Methods: All natural heart components of the new HRM were based on MRI records, which guaranteed electronic functionality. The software CST (Computer Simulation Technology, Darmstadt) was used for the construction, while CST’s material library assured genuine tissue properties. It should be applicable to simulate different heart rhythm diseases as well as various diffusions of electromagnetic fields, caused by electrophysiological conduction, inside the heart tissue.
Results: It was achievable to simulate normal sinus rhythm and fourteen different heart rhythm disturbance with different atrial and ventricular conduction delays. The simulated biological excitation of healthy and sick HRM were plotted by simulated electrodes of four polar right atrial catheter, six polar His bundle catheter, ten polar coronary sinus catheter, four polar ablation catheter and eight polar transesophageal left cardiac catheter (Fig.). Accordingly, six variables were rebuilt and inserted into the anatomic HRM in order to establish heart catheters for ECG monitoring and HF ablation. The HF ablation catheters made it possible to simulate various types of heart rhythm disturbance ablations with different HF ablation catheters and also showed a functional visualisation of tissue heating. The use of tetrahedral meshing HRM made it attainable to store the results faster accompanied by a higher degree of space saving. The smart meshing function reduced unnecessary high resolutions for coarse structures.
Conclusions: The new HRM for EPS simulation may be additional useful for simulation of heart rhythm disturbance, cardiac pacing, HF ablation and for locating and identification of complex fractioned signals within the atrium during atrial fibrillation HF ablation.
Die Simulation komplexer kardialer Strukturen und kardialer Elektroden ist von Bedeutung für die Optimierung langatmiger und kostspieliger klinischer Studien. Das Risiko der Patientengefährdung wird durch diese Methode auf ein Minimum reduziert. Das Ziel der Studie besteht im Entwurf eines anatomisch korrekten 3D CAD Herzrhythmusmodells (HRM) zur Simulation von elektrophysiologischen Untersuchungen (EPU) und Hochfrequenz-(HF-)Ablationen.